![PerkinElmer Frontier Optica FT-IR User Manual Download Page 47](http://html1.mh-extra.com/html/perkinelmer/frontier-optica-ft-ir/frontier-optica-ft-ir_user-manual_1530202047.webp)
Using the Spectrometer . 47
AVI Correction
AVI correction can be selected in the Setup Instrument Advanced tab in Spectrum
AVI correction can only be performed if an AVI Calibration has been set up for the current
configuration and resolution. Select
AVI
from the
Adjustments Toolbox
, available from the
Setup Instrument Advanced Tab. Follow the instructions on-screen. For further information,
refer to the Spectrum on-screen Help file.
What is AVI correction?
The objectives of Absolute Virtual Instrument (AVI) correction are:
•
Consistent performance over time and between instruments
•
Traceability for all measurements.
Although FT-IR spectrometers use a reference laser, the wavenumber calibration and
lineshape are affected by differences in beam divergence and uniformity. This is true for all
FT-IR spectrometers. Differences can occur between instruments, when using different
sampling accessories and when components are changed. AVI allows calibration and
lineshape to be maintained.
The Absolute Virtual Instrument is an instrument with theoretical performance, such that the
result of measuring a known sample on such an instrument can be predicted. So, if we
measure with a real instrument and calculate the software transform to match the theoretical
result, we can apply this transform to future measurements.
The Absolute Virtual Instrument is defined by wavenumber calibration, instrument lineshape
and ordinate accuracy.
What does AVI correction do?
When
AVI
is switched on, the software measures the current instrument profile relative to an
absolute standard (a methane cell) and an ideal lineshape function, and applies a correction.
If you have an automated, internal filter wheel, this can provide correction for any sampling
configuration.
Figure 17 Spectra of methane at 4 cm
−1
resolution as measured (top) and with
AVI (bottom)
Summary of Contents for Frontier Optica FT-IR
Page 1: ...FRONTIER OPTICA FT IR User s Guide MOLECULAR SPECTROSCOPY ...
Page 5: ...Introduction ...
Page 10: ...10 Frontier Optica FT IR User s Guide ...
Page 11: ...Warnings and Safety Information ...
Page 23: ...An Overview of the Frontier Optica FT IR Spectrometer ...
Page 29: ...Unpacking and Installation ...
Page 40: ...40 Frontier Optica FT IR User s Guide ...
Page 41: ...Using the Spectrometer ...
Page 50: ...50 Frontier Optica FT IR User s Guide ...
Page 51: ...Measuring Optical Components ...
Page 58: ...58 Frontier Optica FT IR User s Guide ...
Page 59: ...Routine Maintenance ...
Page 71: ...Advanced Maintenance ...
Page 85: ...Appendices ...