Advanced Maintenance . 79
Installing Filters in the Filter Wheel
The following procedure describes how to install a filter in the filter wheel. The same
procedure should be followed when replacing a filter or the methane cell.
The filter wheel assembly is removed from its mounting and placed on the standard sample
slide baseplate.
1.
Set the
Filter Wheel
to the Position 1 on the Setup Instrument BeamPath tab in the
Spectrum software.
Position 1 will be labelled as
None 15000 – 0 cm
-1
and be the first option in the drop-
down list.
WARNING
Switch off the mains power supply to the spectrometer, wait 60 seconds,
and disconnect the power cable before you open the cover of the
spectrometer. This makes sure that you are safe from electrical shock
and laser radiation.
2.
Open the main cover of the instrument as detailed on page 72 and identify the filter
wheel.
Figure 42 Filter wheel location
Filter wheel
Summary of Contents for Frontier Optica FT-IR
Page 1: ...FRONTIER OPTICA FT IR User s Guide MOLECULAR SPECTROSCOPY ...
Page 5: ...Introduction ...
Page 10: ...10 Frontier Optica FT IR User s Guide ...
Page 11: ...Warnings and Safety Information ...
Page 23: ...An Overview of the Frontier Optica FT IR Spectrometer ...
Page 29: ...Unpacking and Installation ...
Page 40: ...40 Frontier Optica FT IR User s Guide ...
Page 41: ...Using the Spectrometer ...
Page 50: ...50 Frontier Optica FT IR User s Guide ...
Page 51: ...Measuring Optical Components ...
Page 58: ...58 Frontier Optica FT IR User s Guide ...
Page 59: ...Routine Maintenance ...
Page 71: ...Advanced Maintenance ...
Page 85: ...Appendices ...