46 . Frontier Optica FT-IR User's Guide
Figure 16 Atmospheric suppression
NOTE:
Atmospheric suppression should be used with caution when measuring optical
components. For example, when measuring a 1600 cm
–1
CO
2
passband filter
suppressing the CO
2
absorption may produce errors in that region of the spectrum.