Using the Spectrometer . 45
Atmospheric (CO
2
/H
2
O) Suppression
Atmospheric suppression can be selected in the Setup Instrument Advanced tab (Figure 14).
Figure 14 Setup Instrument Advanced tab in Spectrum
What is atmospheric suppression?
This is an atmospheric correction routine. This routine is more powerful than simple
subtraction, overcoming the following issues:
•
Non-linearity due to resolution
•
The measured spectrum is temperature dependent
•
Lineshape and calibration are affected by J-stop and sample or accessory.
What does atmospheric suppression do?
When
CO2/H2O
is switched on, the software uses a single reference spectrum derived from
high resolution data and our understanding of the instrument to model the lineshape, then
finds the current real instrument parameters by least squares fitting to the measured
spectrum.
Figure 15 Correction of an MIR spectrum at 4 cm
−1
resolution
Summary of Contents for Frontier Optica FT-IR
Page 1: ...FRONTIER OPTICA FT IR User s Guide MOLECULAR SPECTROSCOPY ...
Page 5: ...Introduction ...
Page 10: ...10 Frontier Optica FT IR User s Guide ...
Page 11: ...Warnings and Safety Information ...
Page 23: ...An Overview of the Frontier Optica FT IR Spectrometer ...
Page 29: ...Unpacking and Installation ...
Page 40: ...40 Frontier Optica FT IR User s Guide ...
Page 41: ...Using the Spectrometer ...
Page 50: ...50 Frontier Optica FT IR User s Guide ...
Page 51: ...Measuring Optical Components ...
Page 58: ...58 Frontier Optica FT IR User s Guide ...
Page 59: ...Routine Maintenance ...
Page 71: ...Advanced Maintenance ...
Page 85: ...Appendices ...