7. Operating the IVIS
®
Spectrum
38
CAUTION
!
CAUTION!
Caliper recommends using the XGI-8 Gas Anesthesia System
when imaging small animals. This system supplies a controlled
amount of isoflurane to the imaging chamber and continuously reduces the
build-up of isoflurane in the chamber. If you want to use a gas other than the
recommended isoflurane/oxygen gas mixture or pure air, contact IVIS
Imaging Systems Technology Support.
NOTICE
Be careful to use only tubing and other plumbing fixtures that do not fluoresce
or phosphoresce (glow) in the imaging chamber. Contact IVIS Imaging
Systems Technology Support for a list of acceptable materials.
The IVIS Spectrum is equipped with plumbing connections for supplying
isoflurane anesthetic gas to the imaging chamber and exhausting gases out. The
flow of gas into the imaging chamber is manually controlled by the gas valve
on the front panel of the console. The imaging chamber does not include a gas
scavenging system. This is provided by suitable auxiliary equipment such as the
XGI-8 Gas Anesthesia System.
Gas ports are located on the rear of the system console and are labeled GAS IN
and GAS OUT. GAS IN
means the direction of flow is into that port.
Similarly, the port labeled GAS OUT
means that flow can be exhausted out
of this port. The flow of gas from the rear panel GAS IN port is directed to the
main gas control valve on the front of the system and proceeds to a distribution
block located on the imaging platform where it exits through the port labeled
GAS.
Figure 7.3
XGI-8 Gas Anesthesia System
Summary of Contents for Caliper IVIS Spectrum
Page 6: ...Contents iv ...
Page 10: ...1 Welcome 4 This page intentionally blank ...
Page 22: ...3 Warnings 16 This page intentionally blank ...
Page 26: ...4 Legal Notices 20 This page intentionally left blank ...
Page 30: ...5 Specifications 24 This page intentionally blank ...
Page 40: ...6 IVIS Spectrum Components 34 This page intentionally blank ...
Page 50: ...7 Operating the IVIS Spectrum 44 This page intentionally blank ...
Page 68: ...10 Servicing Maintaining the ThermoCube Chiller Unit 62 This page intentionally blank ...
Page 70: ...A Spare Parts List 64 This page intentionally blank ...
Page 74: ...Index 68 ...