Percon CYGNUS PRO User Manual Download Page 8

Calibración del Cygnus Pro

CALIBRATE: 

It indicates the calibration status, as well as the option to calibrate 

or clear the former calibration to make a new one. For the device recalibration, 
you  must  connect  the  Cygnus  Pro  Loop  at  the  Cygnus  Pro  Main  and  press   
ENTER (Picture 8).

To erase the actual calibration, just press DOWN (Picture 9). For recalibrating 
the Cygnus Pro, it´s necessary to erase the existing one and calibrate as usual.

The device cannot be calibrated if the attenuation is over 4 dB. 

The calibration is associated to the model and serial number of the SFP with the 
head. When we change the head with the device switched off or we make a hot 
swapped  and  the  SFP  serial  number  has  changed,  the  status  will  be  “not 
calibrated” if the SFP inserted is correct; otherwise it will inform the user to insert 
a correct SFP.

 

 

After the calibration, the display shows the attenuation of the system once 
deducted the attenuation of the calibration.

7

Cambio de Cabezal ( cambio conector)

The Cygnus Pro allows one main module and several headers with the most usual  
connectors. 

The assembling system follows the design of the rack cases with expansion cards. 
The system consists of assembling the head in the main module through a slot and 
fixing it with two captive ribbed screws. The screws can be comfortably tightened 
manually  for  a  standard  fixation  or  use  a  standard  screwdriver  for  a  stronger 
fixation.

The Cygnus Pro is composed by a  main module, head and loop.

If you require different connectors, you will receive a head and a loop with the 
connectors. You can use the Cygnus Pro Main with the connectors you require 
without buying the complete system. 

Remember to keep the headers protected in case that you have several Cygnus 
Pro Systems.

8

Cygnus Pro Calibration

Change of connector ( head and loop)

Picture 8

picture 9

Summary of Contents for CYGNUS PRO

Page 1: ...US PRO CYGNUS PRO Hybrid Fiber Optic Tester Permanyer Gri Connexions S L U Diputaci 203 bj 08011 Barcelona Spain Tel 34 93 451 78 30 Fax 34 93 323 75 88 mail central percon es http www percon es twitt...

Page 2: ...modo 1310nm Cygnus Pro Stratos HX1080 Neutrik OpticalCON Duo Monomodo 1310nm Neutrik OpticalCON Duo Monomodo 1550nm Neutrik OpticalCON Duo Multimodo 850nm Neutrik OpticalCON Duo Monomodo 1310nm SMPTE...

Page 3: ...out the measuring you must connect the Cygnus Pro Main to the point you want to test and the Cygnus Pro Loop in the other end to close the circuit Elementos del Cygnus Pro Thank you State indicators A...

Page 4: ...printed circuit that manages the operation of the tester Smaller size module that has one or several F O connections that link the optical ports of the connectors and one resistive shield between the...

Page 5: ...f there is a breakdown in the fiber line the device shows attenuation 18 flashing and increases the counter of fiber errors Shortcut in the copper lines If the device detects a shortcut in any of the...

Page 6: ...ST to CURRENT CURR If the tester does not detect any fault the counter is shown in green otherwise the number of errors will be shown in red The attenuation of the fiber is shown in green up to 5 99 d...

Page 7: ...suring configured in continuous mode CONT and history HIST It shows the cumulative errors in the copper lines the current status of the fiber and the number of fiber and copper errors detected until t...

Page 8: ...a correct SFP After the calibration the display shows the attenuation of the system once deducted the attenuation of the calibration 7 Cambio de Cabezal cambio conector The Cygnus Pro allows one main...

Reviews: