Level Probe LGC2
Electrical Connection
2017
-0
2
31
6.3
Cable Specifications
Fuchs recommends using shielded, twisted-pair two-wire cables.
6.3.1
Level probe and Pt100 (optional)
• Commercially available instrument cable
• Terminals, terminal box: 0.08 to 2.5 mm
2
(28 to 14 AWG)
6.3.2
HUT Temperature Head Transmitter (optional) for Device with
4 to 20 mA Analog
• Commercially available instrument cable
• Terminals, terminal box: 0.08 to 2.5 mm
2
(28 to 14 AWG)
• Transmitter connection: max. 1.75 mm
2
(15 AWG)
6.4
Power Consumption
6.4.1
Level probe and Pt100 (optional)
•
0.805 W at 35 V DC (non-hazardous areas)
•
0.690 W at 30 V DC (hazardous areas)
6.4.2
HUT Temperature Head Transmitter (optional) for Device with
4 to 20 mA Analog
0.875 W at 35 V DC
6.5
Current Consumption
6.5.1
Level probe and Pt100 (optional)
• Max.current consumption:
23 mA
• Min. current consumption:
3.6 mA
6.5.2
HUT Temperature Head Transmitter (optional) for Device with
4 to 20 mA Analog
• Max. current consumption:
25 mA
• Min. current consumption:
3.5 mA
Note!
The probe cables are shielded for device versions with outer diameters of 22 mm (0.87 inch)
and 42 mm (1.65 inch).