HAFFMANS RPU 352, RPU 353, RPC 80 RPC 50
2
RPU 352, RPU 353, RPC 80, RPC 50
Page 81 of 112
C3
There has been an error in the communications between the RCP-80 and an external computer
on the External link.
•
The lead that connects the external computer to the RPC-80 may have become
disconnected. Never unplug while the
"Data Transfer..."
message shows.
•
The data might be send in the wrong format or at the wrong data rate. Check if the
settings of the External link match those of the external computer.
•
It is impossible to recover from this type of error and continue the current data transfer.
The external computer will need to re-start the transfer.
ATTENTION!
It is possible that an error could be caused by very severe interference
from other nearby equipment or on the mains wiring. If this is suspected
then the cause should be investigated because the levels of interference
will have to be very high to cause errors.
26.3 M-MESSAGES
All messages starting with a “M” are messages concerning the memory of the PRC-80.
M1
The memory in the clock / calendar chip has become corrupted and the settings stored in it
have been lost. The memory has been tested and found to be operating correctly. The most
likely reason for the memory loss is that the back-up battery power was lost for a short time.
This may have been due to a fault in the battery or the connector. You should check that the
battery connector has not been disconnected by accident.
•
The settings will have all been returned to 00:00:00 entries
•
Templates will have been turned off.
•
The clock time will have returned to 1st January 1992.
M2
The memory that stores all the operational settings for the RPC-80 has become corrupted and
the settings in it have been lost. The memory has been tested and found to be operating
correctly. The following items will have been lost;
•
The optional channel names will have been replaced by blank entries
•
The clock time will have returned to 00:00:00 on 1st January 1992
•
All saved templates will have been deleted and templates turned off
The memory used has a minimum data retention time of over 10 years. If the fault occurs it may
be necessary to replace this memory, return the instrument for repair.
Summary of Contents for HAFFMANS RPU-352
Page 2: ......
Page 4: ......
Page 12: ...HAFFMANS RPU 352 RPU 353 RPC 80 RPC 50 2 Page 12 of 112 RPU 352 RPU 353 RPC 80 RPC 50...
Page 14: ...HAFFMANS RPU 352 RPU 353 RPC 80 RPC 50 2 Page 14 of 112 RPU 352 RPU 353 RPC 80 RPC 50...
Page 18: ...HAFFMANS RPU 352 RPU 353 RPC 80 RPC 50 2 Page 18 of 112 RPU 352 RPU 353 RPC 80 RPC 50...
Page 22: ...HAFFMANS RPU 352 RPU 353 RPC 80 RPC 50 2 Page 22 of 112 RPU 352 RPU 353 RPC 80 RPC 50...
Page 24: ...HAFFMANS RPU 352 RPU 353 RPC 80 RPC 50 2 Page 24 of 112 RPU 352 RPU 353 RPC 80 RPC 50...
Page 32: ...HAFFMANS RPU 352 RPU 353 RPC 80 RPC 50 2 Page 32 of 112 RPU 352 RPU 353 RPC 80 RPC 50...
Page 34: ...HAFFMANS RPU 352 RPU 353 RPC 80 RPC 50 2 Page 34 of 112 RPU 352 RPU 353 RPC 80 RPC 50...
Page 36: ...HAFFMANS RPU 352 RPU 353 RPC 80 RPC 50 2 Page 36 of 112 RPU 352 RPU 353 RPC 80 RPC 50...
Page 42: ...HAFFMANS RPU 352 RPU 353 RPC 80 RPC 50 2 Page 42 of 112 RPU 352 RPU 353 RPC 80 RPC 50...
Page 50: ...HAFFMANS RPU 352 RPU 353 RPC 80 RPC 50 2 Page 50 of 112 RPU 352 RPU 353 RPC 80 RPC 50...
Page 52: ...HAFFMANS RPU 352 RPU 353 RPC 80 RPC 50 2...
Page 54: ...HAFFMANS RPU 352 RPU 353 RPC 80 RPC 50 2 Page 54 of 112 RPU 352 RPU 353 RPC 80 RPC 50...
Page 56: ...HAFFMANS RPU 352 RPU 353 RPC 80 RPC 50 2 Page 56 of 112 RPU 352 RPU 353 RPC 80 RPC 50...
Page 58: ...HAFFMANS RPU 352 RPU 353 RPC 80 RPC 50 2 Page 58 of 112 RPU 352 RPU 353 RPC 80 RPC 50...
Page 60: ...HAFFMANS RPU 352 RPU 353 RPC 80 RPC 50 2 Page 60 of 112 RPU 352 RPU 353 RPC 80 RPC 50...
Page 64: ...HAFFMANS RPU 352 RPU 353 RPC 80 RPC 50 2 Page 64 of 112 RPU 352 RPU 353 RPC 80 RPC 50...
Page 74: ...HAFFMANS RPU 352 RPU 353 RPC 80 RPC 50 2 Page 74 of 112 RPU 352 RPU 353 RPC 80 RPC 50...
Page 76: ...HAFFMANS RPU 352 RPU 353 RPC 80 RPC 50 2 Page 76 of 112 RPU 352 RPU 353 RPC 80 RPC 50...
Page 78: ...HAFFMANS RPU 352 RPU 353 RPC 80 RPC 50 2 Page 78 of 112 RPU 352 RPU 353 RPC 80 RPC 50...
Page 86: ...HAFFMANS RPU 352 RPU 353 RPC 80 RPC 50 2 Page 86 of 112 RPU 352 RPU 353 RPC 80 RPC 50...
Page 88: ...HAFFMANS RPU 352 RPU 353 RPC 80 RPC 50 2...
Page 90: ...HAFFMANS RPU 352 RPU 353 RPC 80 RPC 50 2 Page 90 of 112 RPU 352 RPU 353 RPC 80 RPC 50...
Page 92: ...HAFFMANS RPU 352 RPU 353 RPC 80 RPC 50 2 Page 92 of 112 RPU 352 RPU 353 RPC 80 RPC 50...
Page 94: ...HAFFMANS RPU 352 RPU 353 RPC 80 RPC 50 2 Page 94 of 112 RPU 352 RPU 353 RPC 80 RPC 50...
Page 96: ...HAFFMANS RPU 352 RPU 353 RPC 80 RPC 50 2 Page 96 of 112 RPU 352 RPU 353 RPC 80 RPC 50...
Page 100: ...HAFFMANS RPU 352 RPU 353 RPC 80 RPC 50 2 Page 100 of 112 RPU 352 RPU 353 RPC 80 RPC 50...
Page 102: ...HAFFMANS RPU 352 RPU 353 RPC 80 RPC 50 2 Page 102 of 112 RPU 352 RPU 353 RPC 80 RPC 50...
Page 104: ...HAFFMANS RPU 352 RPU 353 RPC 80 RPC 50 2 Page 104 of 112 RPU 352 RPU 353 RPC 80 RPC 50...
Page 106: ...HAFFMANS RPU 352 RPU 353 RPC 80 RPC 50 2 Page 106 of 112 RPU 352 RPU 353 RPC 80 RPC 50...
Page 108: ...HAFFMANS RPU 352 RPU 353 RPC 80 RPC 50 2 Page 108 of 112 RPU 352 RPU 353 RPC 80 RPC 50...
Page 110: ...HAFFMANS RPU 352 RPU 353 RPC 80 RPC 50 2 Page 110 of 112 RPU 352 RPU 353 RPC 80 RPC 50...