Head space correction
The RPU-353 use a pressure sensor inside the instrument that connects to the container via a
quick release coupling and a fine flexible tube. This causes a slight increase in volume. If the
theoretical volume of the head space of the container is known and entered into the RPC-80
then the unit can make a first order correction for the error. The pressure sensor connections
must be cleaned out after each run so that the pipe and coupling are empty if this error
correction is to operate correctly. The error is only significant when head space volume is less
than about 20 millilitres. The maximum correction allowed will increase the recorded pressure
by 5%.
HEAD SPACE CORRECTION?
Make the correction Yes or No.
HEAD SPACE VOLUME enter the theoretical volume in millilitres (same as cubic centimetres).
Channel names
SET CHANNEL NAMES
sets up names for the three possible channels of the P.U. monitor.
The names can have up to 15 characters (letters or numbers). Channels 1 and 2 are
temperature channels, channel 3 is the pressure channel. For instance channel 1 "PRODUCT"
an d channel 2 "SPRAY". These names appear on the printed results.
ATTENTION
!
After you have set up the graph format for the P.U. monitor user you will
be returned to the SET GRAPH FORMATS display where you can exit
this part of the setting section by pressing the CANCEL button. If you
continue by pressing the OK button you can move on to set up the
formats for the other user.
23.2 P.U. CALCULATION SET-UP
This section covers the selection of P.U. calculation constants and of the P.U. Cut-off
temperature. By choosing
CUSTOM CALCULATION
the P.U. calculation can be changed for
the 100 and 200 series. For the base- and Z value see general product information.
ATTENTION!
The P.U. value calculated by the 100 series can be different form the
P.U. value calculated by the RPC-80.when a custom calculation is used.
P.U. cut-off and the graph
The P.U. Cut-off temperature is printed on the graph as a dashed line so that it is easy to see
the area where the temperature trace exceeds the Cut-off and where P.U.s are being added to
the total.
In the case of two different Cut-off levels being set, they cannot both be shown on the graph
when the RPC-80 is used with a multi-channel monitor.
Summary of Contents for HAFFMANS RPU-352
Page 2: ......
Page 4: ......
Page 12: ...HAFFMANS RPU 352 RPU 353 RPC 80 RPC 50 2 Page 12 of 112 RPU 352 RPU 353 RPC 80 RPC 50...
Page 14: ...HAFFMANS RPU 352 RPU 353 RPC 80 RPC 50 2 Page 14 of 112 RPU 352 RPU 353 RPC 80 RPC 50...
Page 18: ...HAFFMANS RPU 352 RPU 353 RPC 80 RPC 50 2 Page 18 of 112 RPU 352 RPU 353 RPC 80 RPC 50...
Page 22: ...HAFFMANS RPU 352 RPU 353 RPC 80 RPC 50 2 Page 22 of 112 RPU 352 RPU 353 RPC 80 RPC 50...
Page 24: ...HAFFMANS RPU 352 RPU 353 RPC 80 RPC 50 2 Page 24 of 112 RPU 352 RPU 353 RPC 80 RPC 50...
Page 32: ...HAFFMANS RPU 352 RPU 353 RPC 80 RPC 50 2 Page 32 of 112 RPU 352 RPU 353 RPC 80 RPC 50...
Page 34: ...HAFFMANS RPU 352 RPU 353 RPC 80 RPC 50 2 Page 34 of 112 RPU 352 RPU 353 RPC 80 RPC 50...
Page 36: ...HAFFMANS RPU 352 RPU 353 RPC 80 RPC 50 2 Page 36 of 112 RPU 352 RPU 353 RPC 80 RPC 50...
Page 42: ...HAFFMANS RPU 352 RPU 353 RPC 80 RPC 50 2 Page 42 of 112 RPU 352 RPU 353 RPC 80 RPC 50...
Page 50: ...HAFFMANS RPU 352 RPU 353 RPC 80 RPC 50 2 Page 50 of 112 RPU 352 RPU 353 RPC 80 RPC 50...
Page 52: ...HAFFMANS RPU 352 RPU 353 RPC 80 RPC 50 2...
Page 54: ...HAFFMANS RPU 352 RPU 353 RPC 80 RPC 50 2 Page 54 of 112 RPU 352 RPU 353 RPC 80 RPC 50...
Page 56: ...HAFFMANS RPU 352 RPU 353 RPC 80 RPC 50 2 Page 56 of 112 RPU 352 RPU 353 RPC 80 RPC 50...
Page 58: ...HAFFMANS RPU 352 RPU 353 RPC 80 RPC 50 2 Page 58 of 112 RPU 352 RPU 353 RPC 80 RPC 50...
Page 60: ...HAFFMANS RPU 352 RPU 353 RPC 80 RPC 50 2 Page 60 of 112 RPU 352 RPU 353 RPC 80 RPC 50...
Page 64: ...HAFFMANS RPU 352 RPU 353 RPC 80 RPC 50 2 Page 64 of 112 RPU 352 RPU 353 RPC 80 RPC 50...
Page 74: ...HAFFMANS RPU 352 RPU 353 RPC 80 RPC 50 2 Page 74 of 112 RPU 352 RPU 353 RPC 80 RPC 50...
Page 76: ...HAFFMANS RPU 352 RPU 353 RPC 80 RPC 50 2 Page 76 of 112 RPU 352 RPU 353 RPC 80 RPC 50...
Page 78: ...HAFFMANS RPU 352 RPU 353 RPC 80 RPC 50 2 Page 78 of 112 RPU 352 RPU 353 RPC 80 RPC 50...
Page 86: ...HAFFMANS RPU 352 RPU 353 RPC 80 RPC 50 2 Page 86 of 112 RPU 352 RPU 353 RPC 80 RPC 50...
Page 88: ...HAFFMANS RPU 352 RPU 353 RPC 80 RPC 50 2...
Page 90: ...HAFFMANS RPU 352 RPU 353 RPC 80 RPC 50 2 Page 90 of 112 RPU 352 RPU 353 RPC 80 RPC 50...
Page 92: ...HAFFMANS RPU 352 RPU 353 RPC 80 RPC 50 2 Page 92 of 112 RPU 352 RPU 353 RPC 80 RPC 50...
Page 94: ...HAFFMANS RPU 352 RPU 353 RPC 80 RPC 50 2 Page 94 of 112 RPU 352 RPU 353 RPC 80 RPC 50...
Page 96: ...HAFFMANS RPU 352 RPU 353 RPC 80 RPC 50 2 Page 96 of 112 RPU 352 RPU 353 RPC 80 RPC 50...
Page 100: ...HAFFMANS RPU 352 RPU 353 RPC 80 RPC 50 2 Page 100 of 112 RPU 352 RPU 353 RPC 80 RPC 50...
Page 102: ...HAFFMANS RPU 352 RPU 353 RPC 80 RPC 50 2 Page 102 of 112 RPU 352 RPU 353 RPC 80 RPC 50...
Page 104: ...HAFFMANS RPU 352 RPU 353 RPC 80 RPC 50 2 Page 104 of 112 RPU 352 RPU 353 RPC 80 RPC 50...
Page 106: ...HAFFMANS RPU 352 RPU 353 RPC 80 RPC 50 2 Page 106 of 112 RPU 352 RPU 353 RPC 80 RPC 50...
Page 108: ...HAFFMANS RPU 352 RPU 353 RPC 80 RPC 50 2 Page 108 of 112 RPU 352 RPU 353 RPC 80 RPC 50...
Page 110: ...HAFFMANS RPU 352 RPU 353 RPC 80 RPC 50 2 Page 110 of 112 RPU 352 RPU 353 RPC 80 RPC 50...