background image

Technical Note 

Technical Note:  PendoTECH Process Control Systems: Pump Calibration 
Revision 0 
Page 3 of 5 

 

Figure 3.2 TFF GUI "Maintenance View" Tab 

 

Note: The input fields for the “Filtrate Pump” will only appear if “Yes” is selected from the drop-
down menu located below the text reading “Enable filtrate Pump”. 

 
The following table lists some of  the most popular  nominal pump calibration paramters for PendoTECH’s 
standard offering of pumps used with either the NFF or TFF control system. As shown below, the mL/rot 
parameter for peristaltic pumps is dependent on tubing size while it is constant for diaphragm pumps 
irregardless of the size of tubing connected to the inlet/outlet. The nominal values listed below are taken 
directly from manufacturer documentation.    
 

 

Manufacturer 

Model 

Max 

RPM 

ml/rot 

(Size 14) 

ml/rot 

(Size 16) 

ml/rot 

(Size 17) 

ml/rot 

(Size 24) 

ml/rot 

(Size 36) 

Quattroflow 

Q150 

3000 

Quattroflow 

Q1200 

1200 

17 

Quattroflow 

Q4400 

583 

99 

Masterflex 

L/S 

600 

0.22 

0.8 

2.8 

2.8 

4.8 

Watson-Marlow 

120U 

200 

0.14 

0.47 

N/A 

N/A 

N/A 

 
Please refer to vendor documentation provided on PendoTECH.com or contact PendoTECH customer 
service if you do not see your pump or tubing size listed above.  In most cases, the mL/rot is the maximum 
flow in mL/min divided by the maximum RPMs. 
 

4. Flow Verification 

This  section  will  illustrate  how  to  further  calibrate 
pumps if the nominal settings described in section 3 
are  insufficient.  The  PendoTECH  NFF  and  TFF 
systems  control  pumps,  set  to  remote  control,  by 
sending analog signals, generally 4-20 mA, that are 
proportional to the flowrate entered by the user. The 
user  enters  the  unique  calibration  values  for  the 
pump they are using as described in section 3. The 
system  then  scales  the  ouput  linearly  from  the 
minimum  flow  value  (0  LPM)  to  the  calculated 
maximum  flowrate  which is simply equal to MAX 
RPM * mL/rotation. The following graph shows the 
control logic for a Masterflex L/S pump with size 16 
tubing installed.    

Reviews: