Guidance and manufacturer's declaration-electromagnetic immunity
The Model SP3000 Dental Chairs are intended for use in the electromagnetic environment specified
below. The customer or the user of the SP3000 should assure that it is used in such an environment.
electromagnetic
environment Guidance
Compliance
level
IeC60601
Test level
Immunity Test
ELECTROSTATIC
DISCHARGE (ESD) IEC
61000-4-2
61000-4-2
ELECTRICAL FAST
TRANSIENT/BURST
IEC 61000-4-4
suRGe IeC61000-4-5
VOLTAGE DIPS, SHORT
INTERRUPTIONS AND
VOLTAGE VARIATIONS
ON POWER SUPPLY
INPUT LINES
IEC 61000-4-11
POWER FREQUENCY
(50/60 HZ) MAGNETIC
FIELD IEC61000-4-8
U
T
is the AC. mains voltage prior to application of the test level.
+/-6 kV contact
+/-8 kV air
+/-6 kV contact
+/-8 kV air
+/-2 kV for power
supply lines
+1-1 kV for input
output lines
+/-2 kV for power
supply lines
Not applicable,
No I/O lines
+/-1 kV differential mode
+/-2 kV common mode
+/-1 kV
differential mode
+/-2 kV
common mode
<5% U
T
(>95% dip in U
T
)
for 0.5 cycle
40% U
T
(60% dip in U
T
)
for 5 cycles
70% U
T
(30% dip in U
T
)
for 25 cycles
<5% U
T
(>95% dip in U
T
)
for 5 seconds
<5% U
T
(>95% dip in
U
T
) for 0.5 cycle
40% U
T
(60% dip in U
T
)
for 5 cycles
70% U
T
(30% dip in U
T
)
for 25 cycles
<5% U
T
(>95% dip in
U
T
) for 5 seconds
3 A/m
3 A/m
Floors should be wood, concrete or
ceramic tile. If floors are covered
with synthetic material the relative
humidity should be at least 30%
Where labeled, a ground strap
(connected to ground lug) should be
worn to reduce the possibility of
damaged to the unit when servicing.
Mains power quality should be that of
typical commercial or hospital
environment.
Mains power quality should be that of
typical commercial or hospital
environment.
Mains power quality should be that of
typical commercial or hospital
environment. If the user of the
SP3000 requires continued operation
during power mains interruptions, it is
recommended that the SP3000 be
powered by an uninterrupted power
supply or battery.
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital environment.
Summary of Contents for Spirit 3000 Series
Page 1: ...Pelton and Crane Series 3000 Dental Chair Collins Bay Institution Warehouse ...
Page 2: ...SPIRIT 3000 CHAIR TECHNICAL DIMENSIONAL DATA ...
Page 3: ...SPIRIT 3000 CHAIR Dimensions w Junction Box Plumbing Diagram ...
Page 11: ...7 047657 r7 ...
Page 16: ...Spirit 3000 Series Dental Chair USE AND CARE Model 3001 Model 3004 Model 3002 Model 3003 ...