![Pelton & Crane Spirit 3000 Series Installation Instructions Manual Download Page 15](http://html.mh-extra.com/html/pelton-and-crane/spirit-3000-series/spirit-3000-series_installation-instructions-manual_1523569015.webp)
11
047657 r7
Verify the following after installation or servicing of the unit:
All manuals are present.
All labels are present and legible.
No mechanical damage on new installations.
The chair is connected to the correct power source.
The chair is setting on a level surface.
All connections are properly attached.
If applicable, the cover is closed and fasteners tightened (take care not to
pinch tubing on wires).
When depressing the touchpad (if applicable), the unit/chair functions properly.
While running the chair ensure there is nothing leaking from the tubing.
The chair passes a high pot test.
All terminals are connected securely.
The chair passes a ground continuity test.
The internal wiring is in good shape and not frayed.