I
NSTALLATION
Combination Blanket/Fluid Warmer
▪
Operator’s Manual
▪
M N-39930
▪
Rev 2
▪
4/20
24
Continued from previous page
Position the warmer
To position the warmer, do the following.
Step
Action
1.
Make sure
that:
▪
The location where the warmer is being installed is rated to support the
weight of the warmer;
▪
The warmer is within five feet of the appropriate electrical outlet;
▪
You follow the warmer clearance guidelines shown below. Refer to topic
Specification Information
for clearance requirements.
▪
You reference the built-in specification page for cutout dimensions
(P-2150 only).
2.
Move
the warmer to the installation location and onto the final resting
surface.
3.
Lock
the casters, if equipped.
The warmer is now correctly positioned.
Mount the warmer to
floor (P-2150 only)
To mount the warmer to the floor, do the following.
4.
Bolt
the warmer to the floor.
Continued on next page
WARNING:
Tipping hazard.
Mount the P-2150 to the floor before use.
NOTE:
Reference OPM-0561-19 for California seismic conformance.
EC-TS-009211