T
ROUBLESHOOTING
Combination Blanket/Fluid Warmer
▪
Operator’s Manual
▪
MN-39930
▪
Rev 2
▪
4/20
45
T
ROUBLESHOO
TING
What to do if a Power Interruption Occurs
Background
You may need to reset the warmer in the case that a power interruption occurs.
When the power is restored, the ON/OFF status indicator LED decimal
flashes
.
The controller stores all settings and continues operation using these settings in
the event of a power interruption.
Procedure
To continue operation of the warmer, do the following.
Step
Action
1.
Press
the standby button
. The status indicator LED decimal goes out.
If an error code displays, refer to topic
Error Codes
.
2.
Make sure
the fluids and blankets are at the correct temperature before
using.
Result
Resume operation of the warmer.
LED
NOTE:
Inspect the fluids for their expiration date before
use.