  U N P A C K I N G   A N D   S E T - U P

Transport and Storage Environmental Conditions (not to exceed 15 days)
 

• Ambient temperature range of -40  to +159°F (-40° to +70°C)

 

• Relative humidity range of 10% to 100%, including condensation

 

• Atmospheric pressure range of 50KPa to 106KPa

DELIVERY

 

The Pedigo Blanket Warming Cabinet has been 

thoroughly tested and inspected to insure only the 
highest quality unit is provided.  Upon receipt, 
check for any possible shipping damage and report it 
at once to the delivering carrier.  

See Transportation 

Damage and Claims section located in this manual.

 

This appliance, complete with unattached items 

and accessories, may have been delivered in one or 
more packages.  Check to ensure that all standard 
items and options have been received with each 
model as ordered.
 

Save all the information and instructions 

packed with the appliance.  Complete and return 
the warranty card to the factory as soon as possible 
to assure prompt service in the event of a warranty 
parts and labor claim.  
 

This manual must be read and understood by 

all people using or installing the equipment model.  
Contact the Pedigo service department if you have 
any questions concerning installation, operation, 
or maintenance.

NOTE:

  All claims for warranty must include the 

full model number and serial number of
the unit.

UNPACKING

1.  Carefully remove the

appliance from the 
carton or crate.

 

NOTE:

   Do not discard the 

carton and other 
packaging material 
until you have 
inspected the unit 
for hidden damage 
and tested it for 
proper operation.

2.  Read all instructions in this manual carefully before 

initiating the installation of this appliance.

 

DO NOT DISCARD THIS MANUAL. 

 

This manual is considered to be part of the 
appliance and is to be provided to the owner
or manager of the business or to the person 
responsible for training operators.  

Additional 

manuals are available from the Pedigo 
service department.

3.  Remove all protective plastic film, packaging 

materials, and accessories from the appliance 
before connecting electrical power.

T R A N S P O R T   A N D   S T O R A G E

T R A N S P O R T   A N D   S T O R A G E

1

B L A N K E T   W A R M I N G   C A B I N E T

Summary of Contents for P-2010-S

Page 1: ...OPERATION CARE MANUAL OPERATION CARE MM 1070 05 21 P R I N T E D I N U S A BLANKET WARMING CABINET P 2010 S 120V P 2010 S...

Page 2: ...and Preventative Maintenance 9 Protecting Stainless Steel Surfaces 9 Cleaning Agents 9 Cleaning Materials 9 Care and Cleaning 10 Clean the Unit Regularly 10 Troubleshooting Guide 11 Service Parts Lis...

Page 3: ...ent of a warranty parts and labor claim This manual must be read and understood by all people using or installing the equipment model Contact the Pedigo service department if you have any questions co...

Page 4: ...information that is important but not hazard related 1 Pedigo blanket warmers are intended for warming cotton blankets ONLY No other use for this device is authorized or recommended 2 This device is i...

Page 5: ...LIANCE NAMEPLATE NOTE An electrical wiring diagram for this unit can be found in a sleeve on the bottom side of the top cover of the control Safety Class I Equipment Protective Earth Ground Symbol Med...

Page 6: ...IEC 61000 3 3 Complies Guidance and manufacturer s declaration electromagnetic immunity The Model P 2010 S is intended for use in the electromagnet environment specified below The customer or the end...

Page 7: ...not be predicted theoretically with accuracy To assess the electromagnetic environment due to fixed RF transmitters an electromagnetic site survey should be considered If the measured field strength i...

Page 8: ...tch and a temperature control knob that is adjustable from 16 C 60 F to 92 C 200 F The unit includes a digital display and a temperature display light that indicates that the cabinet has reached the d...

Page 9: ...7 D I M E N S I O N S C L ELECTRICAL CONNECTION 1 5 38mm FROM TOP 16 25 411mm Cavity 13 25 337mm Cavity 14 365mm Cavity 21 5 546mm 20 3 516mm 20 8 528mm 27 685mm 41 1035mm...

Page 10: ...tion and then show the temperature in C 3 Repeat to toggle to Fahrenheit Note With a power failure the control will retain the C or F setting selected by the user when power is restored 4 The temperat...

Page 11: ...use a non abrasive scouring pad on dif cult areas and make certain to scrub with the visible grain of surface metal to avoid surface scratches Never use wire brushes metal scouring pads or scrapers to...

Page 12: ...e on stainless steel surfaces Replace blanket support assembly 5 Clean the exterior of the cabinet with a cleaner recommended for stainless steel surfaces Spray the cleaner on a clean cloth and wipe w...

Page 13: ...te relay Bad wire connections or open heating cable If none of the above contact Pedigo Service Department E 31 Over temperature Unit will shut down Unit has exceeded its maximum allowable set tempera...

Page 14: ...OR HANDLE WITH STRIKER 1 HD 2007 17 HINGE SET 1 PAIR OF 2 HINGES 1 HG 2015 18 THERMOSTAT KNOB 1 KN 26568 19 TOP WELD ASSEMBLY 1 5011144 OPTIONS AND ACCESSORIES 20 CASTER STAND ASSEMBLY WITH 5 127mm CA...

Page 15: ...13 1 2 3 4 5 6 19 16 14 12 17 11 PART NUMBERS AND DRAWINGS ARE SUBJECT TO CHANGE WITHOUT NOTICE 10...

Page 16: ...14 REFER TO WIRE DIAGRAM UNDER TOP COVER OF P 2010 S FOR MOST CURRENT VERSION PART NUMBERS AND DRAWINGS ARE SUBJECT TO CHANGE WITHOUT NOTICE...

Page 17: ...rs rst The parts warranty on all other parts remains in effect three 3 years from installation or thirty nine 39 months from the shipping date whichever occurs rst This warranty does not apply to 1 Ca...

Page 18: ...MN 29621 05 21...