Notes:
*
If the measured resistance value exceeds the maximum
value of the range selected, 1 appears flashes. This
indicates an overload. Select a higher range. In this mode,
the beeper does not sound.
*
When you short the test leads in the 200
range, your
meter displays a small value (no more than 0.3
). This
value is due to your
meter’s and test leads internal
resistance. Make a note of this value and subtract it from
small resistance measurements for better accuracy.
6.4. Capacitance measurements
Warning !
To avoid electric shock, disconnect power to the unit under test
and discharge all capacitors before taking any capacitance
measurements. Remove the batteries and unplug the line cords.
1. Set the function switch to the CAP position.
2. Press the LC-button
3.
Plug the black test lead into your meter’s V/
/Hz- terminal
and the red test lead into your meter’s mA-terminal.
4. Touch the test leads to the capacitor to be tested. The
display will indicate the proper decimal point value.
6.5. Inductance measurements
1. Switch to a proper range an
d insert test leads into “mA” and
“V/
/Hz” jack.
2. Press LC-button
3. Connect the test leads crossly to the two ends of inductor.