PeakTech 1280 Operation Manual Download Page 58

 

-58- 

 

1. Safety Precautions 

 
This product complies with the requirements of the following European Community Directives: 
2004/108/EC (Electromagnetic Compatibility) and 2006/95/EC (Low Voltage) as amended by 
2004/22/EC (CE-Marking). 

 

Overvoltage category II; pollution degree 2. 
 
To ensure safe operation of the equipment and eliminate the danger of serious injury due to short-
circuits (arcing), the following safety precautions must be observed. 
 
Damages resulting from failure to observe these safety precautions are exempt from any legal claims 
whatever. 

 

Do not use this instrument for high-energy industrial installation measurement. 

 

Do not place the equipment on damp or wet surfaces. 

 

Do not exceed the maximum permissible input ratings (danger of serious injury and/or destruction 
of the equipment). 

 

The meter is designed to withstand the stated max voltages. If it is not possible to exclude without   

 

that  impulses,  transients,  disturbance  or  for  other  reasons,  these  voltages  are  exceeded  a 
suitable presale (10:1) must be used. 

 

Disconnect test leads or probe from the measuring circuit before switching modes or functions. 

 

Check  test  leads  and  probes  for  faulty  insulation  or  bare  wires  before  connection  to  the 
equipment. 

 

Conduct measuring works only in dry clothing and rubber shoes, i. e. on isolating mats. 

 

Never touch the tips of the test leads or probe. 

 

Comply with the warning labels and other info on the equipment. 

 

The measurement instrument is not to be to operated unattended. 

 

Do not subject the equipment to direct sunlight or extreme temperatures, humidity or dampness. 

 

Do not subject the equipment to shocks or strong vibrations. 

 

Do not operate the equipment near strong magnetic fields (motors, transformers etc.). 

 

Keep hot soldering irons or guns away from the equipment. 

 

Allow  the  equipment  to  stabilize  at  room  temperature  before  taking  up  measurement  (important 
for exact measurements). 

 

Use caution when working with voltages above 35V DC or 25V AC. These Voltages pose shock  
hazard. 

 

Periodically  wipe  the  cabinet  with  a  damp  cloth  and  mid  detergent.  Do  not  use  abrasives  or 
solvents. 

 

The meter is suitable for indoor use only  

 

Do not store the meter in a place of explosive, inflammable substances. 

 

Opening  the  equipment  and  service 

–  and  repair  work  must  only  be  performed  by  qualified 

service personnel 

 

Measuring instruments don’t belong to children hands.

 

 

Cleaning the cabinet 

Prior to cleaning the cabinet, withdraw the mains plug from the power outlet. 
Clean only with a damp, soft cloth and a commercially available mild household cleanser. Ensure that 
no water gets inside the equipment to prevent possible shorts and damage to the equipment. 

 
 
 
 
 
 
 
 
 
 
 
 

Summary of Contents for 1280

Page 1: ...PeakTech 1280 1285 1290 1325 1330 Bedienungsanleitung Operation manual PC Oszilloskope mit USB LAN PC Oscilloscopes with USB LAN ...

Page 2: ...iten nur in trockener Kleidung und vorzugsweise in Gummischuhen bzw auf einer Isoliermatte durchführen Messspitzen der Prüfleitungen nicht berühren Warnhinweise am Gerät unbedingt beachten Gerät darf nicht unbeaufsichtigt betrieben werden Gerät keinen extremen Temperaturen direkter Sonneneinstrahlung extremer Luftfeuchtigkeit oder Nässe aussetzen Starke Erschütterung vermeiden Gerät nicht in der N...

Page 3: ...ndungen eingesetzt werden Lesen Sie die folgenden Sicherheitshinweise bevor Sie das Gerät in Betrieb nehmen Warnung Um Feuer oder einen elektrischen Schlag zu vermeiden benutzen Sie bitte das richtige Netzteil Verwenden Sie nur das Netzteil vom Hersteller Warnung Die Kanäle des Oszilloskops sind nicht elektrisch getrennt Die Kanäle sollten gemeinsame Masse bei der Messung anwenden Um Kurzschlüsse ...

Page 4: ...rischen Schlag zu vermeiden wenn ein Tastkopfeingang mit mehr als 42VSpitze 30 Veff oder mit Schaltungen von mehr als 4800VA verbunden ist Verwenden Sie nur mit dem Messgerät gelieferten oder für das Oszilloskop als angegebene isolierte Tastköpfe und Adapter Vor Gebrauch überprüfen Sie Tastköpfe und Zubehör auf mechanische Beschädigungen und ersetzen Sie diese wenn sie beschädigt sind Entfernen Si...

Page 5: ...n ein Netzwerk 4 Tastkopfkompensation Testsignal Ausgang 3 3 1KHz 5 MULTI Port Trigger Ausgang Pass Fail Ausgang EXT Trigger Eingang 6 Kanal 2 Signal Eingang 7 Kanal 1 Signal Eingang Hinweis Wenn Sie das Oszilloskop in einem Netzwerk betreiben und somit das Gerät über den LAN Port mit einem PC verbinden ist es notwendig das Oszilloskop mit dem AC Netzadapter mit Spannung zu versorgen 1 2 3 4 5 6 7...

Page 6: ...rsorgung über USB Port des PC s Kommunikation mit PC Software 2 Tastkopfkompensation Testsignal Ausgang 5V 1KHz 3 MULTI Port EXT Trigger Eingang Trigger Ausgang Pass Fail Ausgang 4 Kanal 2 Signal Eingang 5 Kanal 1 Signal Eingang 1 2 3 4 5 ...

Page 7: ... PC verbinden ist es notwendig das Oszilloskop mit dem AC Netzadapter mit Spannung zu versorgen 4 Verbindung mit dem PC 1 Installieren Sie die Software auf der mitgelieferten CD 2 Doppelklicken Sie auf die nach der Installation erstellte Desktop Verknüpfung PeakTech_VO 3 Verbindung Schliessen Sie das mitgelieferte USB Kabel an den USB Anschluss des Oszilloskops Abb 3 1 an einen USB Port eines PC s...

Page 8: ...zustellen Die Standardeinstellung des Dämpfungskoeffizienten des Tastkopfes in der Software ist 10X 2 Stellen Sie den Schalter des Oszilloskop Tastkopfes auf 10X und verbinden Sie den Tastkopf mit CH1 Kanal Verbinden Sie die Tastkopfspitze und die Erdungsklemme mit dem Anschluss der Tastkopfkompensation am Oszilloskop 3 Klicken Sie auf oben rechts in der Softwareoberfläche um den AUTO Set Modus zu...

Page 9: ...ändert oder überprüft werden soll drücken Sie die Taste für das Funktionsmenü des jeweiligen Kanals und dann die dem Tastkopf entsprechende Auswahltaste bis der richtige Wert angezeigt wird Diese Einstellung bleibt gültig bis sie wieder geändert wird Hinweis Der Dämpfungsfaktor des Tastkopfes im Menü ist werksmäßig auf 10X voreingestellt Stellen Sie sicher dass der am Dämpfungsschalter des Tastkop...

Page 10: ...öpfe zu vermeiden darf es nicht allen Sprays Flüssigkeiten bzw Lösungsmitteln ausgesetzt werden Reinigung Überprüfen Sie das Gerät und die Tastköpfe so oft wie die Betriebsbedingungen es erfordern Um das Gerät äußerlich zu reinigen führen Sie die folgenden Schritte aus 1 Wischen Sie den Staub auf dem Gerät und den Tastköpfen mit einem weichen Tuch ab 2 Trennen Sie das USB Kabel vor dem Reinigen de...

Page 11: ...eis Wenn Sie ein USB Kabel verwenden das nicht von uns geliefert wurde könnten einige Probleme wie Verbindungsfehler und Signalstörungen auftreten Für Windows Vista oder Windows 7 Während der gesamten Installation stellen Sie bitte sicher dass das Gerät läuft und mit den USB Ports des PC s verbunden ist Klicken Sie mit der rechten Maustaste auf Computer Dieses Icon finden Sie auf dem Desktop oder ...

Page 12: ... 12 Klicken Sie mit der rechten Maustaste auf das unbekannte Gerät und klicken Sie auf Treibersoftware aktualisieren Im folgenden Fenster wählen Sie dann Auf dem Computer nach Treibersoftware suchen ...

Page 13: ...n Verzeichnis Pfad für die Installation der Treiber Software und klicken Sie auf Weiter Hinweis Die Treibersoftware befindet sich im Installationsverzeichnis der Geräte Software mit dem Namen usbdrv und der Inhalt wird wie folgt dargestellt ...

Page 14: ...ateien Sie können die Datei in verschiedenen Verzeichnisses finden wie x86 ia64 oder amd64 in Abhängigkeit von der CPU Meistens wird x86 oder amd64 benötigt Zurück zur Treiberinstallation nach dem letzten Weiter Schritt wird das System die Treiber Software für Sie installieren ...

Page 15: ...e folgende Abb Wählen Sie einfach Installation dieser Treibersoftware fortsetzen um fortzufahren Für Windows 7 64 Bit Manchmal kann ein Fenster mit dem Namen Windows Sicherheit angezeigt werden wie folgende Abb Klicken Sie einfach auf Installieren um fortzufahren Und fahren Sie mit der Installation fort Installation abgeschlossen ...

Page 16: ...ter angezeigt mit der Information Windows hat erfolgreich Ihre Treiber Software aktualisiert Schließen Sie das Fenster Im Windows Geräte Manager werden Sie ein Gerät LibUSB Win32Devices finden Jetzt arbeitet der USB Treiber ...

Page 17: ...wird erscheint Neue Hardware gefunden auf dem Windows Bildschirm Oder Sie klicken mit der rechten Maustaste auf Computer und wählen Verwalten im linken Bereich Computerverwaltung wird die Option Geräte Manager angezeigt Doppelklicken Sie auf den Eintrag USB Device mit im mittleren Bereich zum Öffnen des Installations Assistenten ...

Page 18: ... or specific location Advanced Wählen Sie Search for the best driver in these locations und wählen dann Include this location in the search und geben Sie ein Verzeichnis Standort für USB Treiber an der als usbdrv und unter dem Verzeichnis in dem Sie das Programm installiert haben benannt ist ...

Page 19: ...ie Installation ist abgeschlossen und fertiggestellt Windows zeigt an dass die Installation abgeschlossen wurde Im Geräte Manager wird angezeigt dass der Treiber installiert ist Jetzt können Sie das Programm für USB Kommunikation verwenden ...

Page 20: ...10 Zurück zum Hauptmenü 11 Menü ausblenden 12 Spannungsmessung mit Hilfe des Cursors 13 Die rote Markierung zeigt die Position des Triggerpegels für Kanal 1 gelb für Kanal 2 Es kann nach oben und unten verschoben werden 14 Funktionsmenü Klicken Sie um das Menü auszublenden 15 Kurzwahltaste zum Zurücksetzen auf Werkseinstellungen Kurzwahltaste zum Exportieren von Wellenformen Wechseln zwischen dem ...

Page 21: ...p wieder AutoSet In Prozess der Selbstregistrierung Der Status wenn das Oszilloskop nicht verbunden ist Offlin Keine USB Verbindung mit dem Oszilloskop USB verfügbare USB Geräte gefunden USBDrvEr Fehler bei USB Treiberinstallation MachineNotSuppor Kann Gerät nicht erkennen Der Status wenn das Oszilloskop verbunden ist Linking Verbindung mit Oszilloskop aktiv Connec Verbindung erfolgreich Matc Anpa...

Page 22: ...Tastkopfes im Menü ist werksmäßig auf 10X voreingestellt Stellen Sie sicher dass der am Dämpfungsschalter des Tastkopfes eingestellte Wert dem am Oszilloskop eingestellten Dämpfungswert entspricht Abb 10 Dämpfungsschalter Hinweis Wenn der Dämpfungsschalter auf 1X eingestellt ist begrenzt der Tastkopf die Bandbreite des Oszilloskops auf 5 MHz Sie müssen den Schalter auf 10X stellen wenn Sie die ges...

Page 23: ...le Triggermodus ist es ein Ein Kanal Cymometer und kann nur die Frequenz des getriggerten Kanals messen Im ALT Triggermodus ist es ein Zwei Kanal Cymometer und kann die Frequenz von zwei Kanälen messen 8 3 Einstellung des horizontalen Systems Sie können die entsprechenden Parameter des horizontalen Systems im Erfassungs und Zeitbasis Fenster 17 in 8 Einführung in die Benutzeroberfläche einstellen ...

Page 24: ...u erfassen und genügend Daten gesammelt hat wird links vom Triggerpunkt eine Wellenform dargestellt Die Datenerfassung arbeitet auch weiter während der Wartezeit bis zum nächsten Triggerpunkt Sobald ein Triggerpunkt erkannt wird sind ausreichend kontinuierliche Daten vorhanden um die Wellenform auf der rechten Seite des Triggerpunktes darstellen zu können Klicken Sie auf um das Hauptmenü aufzurufe...

Page 25: ...ger Das Oszilloskop beginnt die Triggerung entsprechend der Anstiegs bzw Abfallgeschwindigkeit des Signals Impuls Trigger Tritt bei Impulsen mit bestimmten Breiten ein Die vier Triggermodi im Single Triggermenü im Einzelnen Edge Ein Edge Trigger tritt beim Trigger Schwellenwert des Eingangssignals auf Wählen Sie den Edge Triggermodus um auf der ansteigenden oder abfallenden Flanke des Signals zu t...

Page 26: ...ofeld Even gerades Videofeld oder Line Number in erstellter Videozeile Wenn Line Number ausgewählt wurde kann die Zeilennummer mit den Up Down Tasten von 1 525 eingestellt werden 3 SET Hold Off Über diese Einstellung können Wellenform erfasst werden auch wenn kein Trigger auftritt Slope Trigger Der Slope Modus lässt das Oszilloskop innerhalb eines festgelegten Zeitraums auf der ansteigenden abfall...

Page 27: ...on Alternate die ausgewählte Schaltfläche erhält eine Markierung 2 Auswählen der Triggerquelle Source 3 Trigger Modus auswählen Hinweis Im alternierenden Trigger wird nur höchsten ein Kanal im Video Modus dargestellt Sie können den Video Modus nicht auf beiden Kanälen gleichzeitig anwenden Einstellungen im Trigger Fenster Einführung der Symbole im Trigger Fenster Aufsteigende Flanke Abfallende Fla...

Page 28: ...cken Sie CH1 oder CH2 und setzen Sie ein Häkchen vor Invertieren um das Signal zu invertieren bzw entfernen Sie das Häkchen um die Wellenform im Normalzustand anzuzeigen Kanalkopplung einstellen DC Lässt AC und DC Komponenten im Eingangssignal passieren AC Blockiert die DC Komponente im Eingangssignal Ground Eingangssignal ist unterbrochen 9 1 Einstellen des Tastkopf Dämpfungsfaktors Für korrekte ...

Page 29: ...nen Werte der beiden Kanäle können gleichzeitig angezeigt werden 10 2 Entfernen einer automatischen Messung Entfernen Sie das Häkchen vor der Art der Messung die Sie entfernen möchten Klicken Sie auf Alle entfernen um alle Messungen zu entfernen 10 3 Die automatische Messung von Spannung Parameter Das PeakTech Oszilloskop stellt automatische Spannungs Messungen einschließlich Vpp Vmax Vmin Vavg Va...

Page 30: ...igt um von 10 auf 90 ihrer Amplitude zu steigen Fall Time Abfallzeit Die Zeit die die Vorderflanke des ersten Impulses in der Kurve benötigt um von 90 auf 10 ihrer Amplitude zu fallen Width Die Breite des ersten positiven Impulses am 50 Amplitudenpunkt Width Die Breite des ersten negativen Impulses am 50 Amplitudenpunkt Delay 1 2 Die Verzögerung zwischen den beiden Kanälen an der Anstiegsflanke De...

Page 31: ...rades und der Möglichkeit einer Reduzierung der Störungen verwendet Durchschnitt Dient zur Verringerung von willkürlich auftretenden Störungen jeder Art mit einer optionalen Anzahl von Mittelwertbildungen 1 128 Normal Sampling Modus in dem kein Grat erkannt werden konnte Spitzenwert Erkennungs Modus unter dem die Grate auf der fallenden Flanke des Rechtecksignals und ein starkes Rauschen erkannt w...

Page 32: ...ie auf die Zeit Option es werden dann zwei hellrote Linien entlang der vertikalen Richtung des Bildschirms angezeigt die Cursor 1 und Cursor 2 darstellen Platzieren Sie den Mauszeiger über Cursor 1 oder Cursor 2 und ziehen Sie den Mauszeiger nach dem Wechsel auf und legen Sie die Positionen des Cursor 1 und Cursor 2 an Hand der Wellenform die gemessen werden soll fest Im Fenster für Cursormessunge...

Page 33: ...nten 12 4 Die Cursor Messung im FFT Modus Art der Messung auswählen Wählen Sie entweder Amplitudenmessung Frequenzmessung oder beides zur gleichen Zeit bei der Art der FFT Messung Frequenzmessung 1 Klicken Sie auf Frequenz 2 Gehen Sie zurück ins Hauptmenü und wählen Sie die Funktion Mathe Funkt und dann FFT Es werden zwei hellrote Linien für Cursor 1 und Cursor 2 entlang der vertikalen Richtung im...

Page 34: ...uszeiger nach dem Wechsel auf und legen Sie die Position des Cursor 1 und Cursor 2 an Hand der Wellenform die gemessen werden soll fest 3 Im Fenster für Cursormessungen wird links unten die aktuelle Position der beiden Cursor und die absolute Amplitudendifferenz der beiden Cursors angezeigt Siehe Bild unten 13 Einstellen des Anzeigesystems Klicken Sie um das Hauptmenü anzuzeigen und wählen Sie Anz...

Page 35: ...n Persist können Sie den Nachleuchteffekt eines Röhrenoszilloskops simulieren die gespeicherten Originaldaten werden verblasst die neuen Daten in kräftiger Farbe dargestellt Verschiedene Nachleuchtzeiten können gewählt werden Aus 0 5 Sekunde 1 Sekunde 2 Sekunden 5 Sekunden und unendlich Wenn die unendlich Option ausgewählt wird werden die Messpunkte gespeichert bis der Wert verändert wird Klicken ...

Page 36: ...w Kanal 2 1 Schalten Sie CH1 und CH2 ein 2 Klicken Sie um das Hauptmenü anzuzeigen und wählen Sie Mathe Funkt 3 Setzen Sie ein Häkchen bei Mathe Funkt 3 Wählen Sie die Faktoren und die Berechnungsfunktion aus Wählen Sie die Spannungsteilung M Die Software wandelt die Wellenformen entsprechend der beiden Faktoren um die ausgewählte Spannungsteilung M zu berechnen Die grüne berechnete Kurve M wird a...

Page 37: ...loskops kann 2048 Punkte des Zeitbereichsignals in dessen Frequenzanteile umwandeln Die Endfrequenz enthält 1024 Punkte von 0Hz bis zur Nyquist Frequenz Zur Verwendung der FFT Funktion verfahren Sie beispielsweise wie folgt 1 Klicken Sie um das Hauptmenü anzuzeigen und wählen Sie Mathe Funkt 2 Setzen Sie ein Häkchen vor FFT Das Anzeigefenster wechselt in die dreifach Skalenansicht 3 Machen Sie die...

Page 38: ...die genaue Messung der Amplitude dieser Frequenzen Es ist das beste Fenster für die Messung des Frequenz spektrums von nicht repetitiven Signalen und der Messung von Frequenzanteilen nahe DC Verwenden Sie das Rechteckfenster für die Messung von Transienten oder Spitzen bei denen das Signal niveau vor und nach dem Ereignis fast gleich ist Auch verwendbar für Sinuswellen mit gleicher Amplitude und m...

Page 39: ...hstehenden Bildern sind Beispiele für die Messung einer Sinuswelle mit einer Frequenz von 1 kHz unter der Auswahl von vier verschiedenen FFT Fenstern Hamming Fenster Rechteck Fenster Blackman Fenster Hanning Fenster ...

Page 40: ...enannt Werden nicht genug Abtastpunkte erfasst und liegt die Frequenz über der Nyquistfrequenz tritt das Phänomen der falschen Wellenform auf Beachten Sie daher mehr die Beziehung zwischen der abgetasteten und gemessenen Frequenz Hinweis Im FFT Modus können nur Einstellungen der Kopplung Spannungsteilung und Nullstellung im Kanal Fenster vorgenommen werden andere Einstellungen sind nicht möglich W...

Page 41: ...zwei Cursor definiert wurde wird auf die volle Bildschirmgröße expandiert Sie können die Zeitbasis W und die horizontale Trigger Position T im Menü Zoom einstellen Sie können auch den roten Zeiger auf der horizontalen Trigger Position anpassen 6 Einführung in die Benutzeroberfläche Sie können auch im Zeit Fenster sämtliche Einstellungen vornehmen wie im folgenden Bild dargestellt ...

Page 42: ...aren Ausgangsport auch Fail oder Passsignale ausgeben So führen Sie eine Pass Fail Prüfung durch 1 Wählen Sie den Eingangskanal CH 1 CH 2 oder Math 2 Stellen Sie die horizontale und vertikale Toleranz ein Hierfür können Sie die Schaltfläche oder den Wert direkt eingeben 3 Klicken Sie auf die Schaltfläche Erstellen 4 Einstellen der Ausgabetyps Wählen Sie Pass oder Fail Markieren Sie die Optionen Ri...

Page 43: ...hme 18 1 Aufnahme 1 Wählen Sie die Record Schaltfläche 2 Klicken Sie auf Speichern unter um den Speicherort einzugeben Wellenform Dateien haben die Endung cap 3 Einstellen der Intervallzeit und End Frame Die Intervallzeit bezieht sich auf den Abstand zwischen den aufgenommenen Bildern Die Intervallzeit reicht von 0 bis 100 000ms und in Schritten von 10ms 4 Stellen Sie die Wellenform in den Run Sta...

Page 44: ...Wiedergabe Schleife auszuwählen Wird Wiederholung deaktiviert wird die Wellenform nur einmal wiederholt 6 Klicken Sie Play Anzahl zeigt die aktuelle Frame Nummer die abgespielt wird an 7 Klicken Sie Pause um die Wiedergabe anzuhalten 8 Ziehen Sie den Schieberegler um die Frame Nummer anzuzeigen die Sie benötigen Hinweis 1 Wenn die Software im RUN Modus ist wird die aktuelle Datenerfassung während ...

Page 45: ...gen und um die Farbe der Vorschau Seite auszuwählen Bild speichern Klicken Sie Bild Speich um einen Screenshot als Image Datei png bmp oder gif zu speichern Speichern Anzeigen Mit dieser Funktion können bis zu 8 Referenzwellenformen gespeichert werden Diese Wellenformen können mit der aktuellen Wellenform gleichzeitig angezeigt werden Die aufgerufene Wellenform kann nicht eingestellt oder veränder...

Page 46: ...Sie können diese Selbst Kalibrierung jederzeit durchführen Eine Selbst Kalibrierung muss ausgeführt werden wenn die Umgebungstemperatur sich um 5 C oder mehr ändert Vor der Durchführung einer Selbst Kalibrierung müssen sämtliche Tastköpfe oder Anschlussleitungen von den Eingängen getrennt werden Wenn das Gerät soweit vorbereitet ist klicken Sie auf Self Cal 19 3 Vorgabe Werkseinstellungen Klicken ...

Page 47: ...n sind Auto Set Run Stop Single Trigger Siehe auch 7 Einführung in die Benutzeroberfläche AutoSet Diese Taste dient zur automatischen Einstellung aller für die Erzeugung einer sichtbaren Wellenform benötigten Steuerwerte des Geräts Drücken Sie die Taste AUTOSET das Oszilloskop führt dann eine schnelle automatische Messung des Signals durch Die folgende Tabelle zeigt die Parameterwerte der Funktion...

Page 48: ...szilloskop mit einem PC zu verbinden Nachdem die Verbindung erfolgreich aufgebaut ist klicken Sie auf um das Hauptmenü anzuzeigen und wählen Sie Einstellungen klicken Sie dann auf Netzwerk 2 Einstellen der IP Adresse und des Ports des Oszilloskops Im Netzwerk Menü klicken Sie auf OK um Geräte Netzwerk Einstellungen einzugeben Wählen Sie das Oszilloskop welches verbunden werden soll aus der Liste a...

Page 49: ... das LAN Kabel mit der LAN Schnittstelle des Oszilloskops stecken Sie das andere Ende in die LAN Schnittstelle des Computers 3 Stellen Sie die Parameter im Netzwerk Menü ein Klicken Sie auf um das Hauptmenü anzuzeigen und wählen Sie Einstellungen klicken Sie auf Netzwerk Stellen Sie die IP Adresse und den Port auf den gleichen Wert des Oszilloskops wie in Schritt 2 ein 4 Klicken Sie auf Verbinden ...

Page 50: ...sein wie der Router Angenommen IP Adresse 192 168 1 71 Subnet Maske 255 255 255 0 Standard Gateway 192 168 1 1 22 4 Einstellen der Netzwerk Parameter des Oszilloskops Eine Verbindung über USB aufbauen und das Hauptmenü aufrufen Verwenden Sie das mitgelieferte USB Kabel um das Oszilloskop mit einem PC zu verbinden Nachdem die Verbindung erfolgreich aufgebaut ist klicken Sie auf um das Hauptmenü anz...

Page 51: ...fgelistet ist klicken Sie auf Refresh um die Liste zu aktualisieren Geben Sie die IP Adresse des Oszilloskops ein Die ersten drei Bytes sollten dieselben sein wie die IP Adresse des Computers Beispiel 192 168 1 71 Das letzte Byte sollte unterschiedlich sein Hier stellen wir sie auf 192 168 1 72 Der Bereich der Port Werte ist 0 4000 hier setzen wir es auf 3000 Der Standard Gateway und die Subnet Ma...

Page 52: ...rwenden Sie ein LAN Kabel Patch Kabel um das Oszilloskop mit dem Router zu verbinden Der Computer sollte auch mit dem Router verbunden sein 3 Stellen Sie die Parameter im Netzwerk Menüein Klicken Sie auf um das Hauptmenü anzuzeigen und wählen Sie Einstellungen klicken Sie auf Netzwerk Stellen Sie die IP Adresse und den Port auf den gleichen Wert des Oszilloskops wie in Schritt 2 ein 4 Klicken Sie ...

Page 53: ... 1 Extern P 1325 1330 4 1 Extern Abtastung Modus Normal Spitzenwerterkennung Durchschnitt Abtastrate Echtzeit P 1280 Dual CH 250 MSa s Single CH 500 MSa s P 1285 Dual CH 500 MSa s Single CH 1 GSa s P 1290 Dual CH 100 MSa s Single CH 100 MSa s P 1325 Four CH 125 MSa s Dual CH 250 MSa s Single CH 500 MSa s P 1330 Four CH 250 MSa s Dual CH 500 MSa s Single CH 1 GSa s Eingang Eingangskopplung DC AC Gr...

Page 54: ...Geschwindigkeit Sa div P 1280 1290 1325 5 ns div 100 s div step by 1 2 5 P 1285 1330 2 ns div 100 s div step by 1 2 5 Abtastrate Zeitverzögerungs genauigkeit 100 ppm Intervall T Genauigkeit DC 100 MHz Single 1 Zeitintervall 100 ppm x Messwert 0 6 ns Durchschnitt 16 1 Zeitintervall 100 ppm x Messwert 0 4 ns Vertikales System A D Wandler P 1280 P 1285 P 1290 8 bits Auflösung 2 Kanäle gleichzeitig P ...

Page 55: ...ang Ausgang Trigger Leistungsmerkmale Bemerkungen Triggerpegel Bereich Intern 5 div vom Bildschirmzentrum Triggerpegel Genauigkeit typisch Intern 0 3 div EXT TTL Trigger Verdrängung Entsprechend der Speicherlänge und Zeitbasis Trigger Holdoff Bereich 100 ns 10 s 50 Pegeleinstellung typisch Eingangsfrequenz 50 Hz Flankentrigger Slope Aufsteigend Fallend Pulstrigger Triggerbedingung Positiver Puls N...

Page 56: ...W Umgebungsbedingungen Temperatur Betriebstemperatur 0 C 40 C Lagertemperatur 20 C 60 C Relative Luftfeuchtigkeit 90 Höhe ü n N Im Betrieb 3 000 m Nicht im Betrieb 15 000 m Kühlung Natürliche Konvektion Mechanische Spezifikationen Abmessungen B x H x T P 1280 1285 1325 1330 120 x 18 x 190 mm P 1290 120 x 18 x 170 mm Gewicht 260 g Anhang A mitgeliefertes Zubehör Tastkopf 1 2 m 1 1 10 1 umschaltbar ...

Page 57: ...nur mit schriftlicher Genehmigung des Herausgebers gestattet Letzter Stand bei Drucklegung Technische Änderungen des Gerätes welche dem Fortschritt dienen vorbehalten Hiermit bestätigen wir dass alle Geräte die in unseren Unterlagen genannten Spezifikationen erfüllen und werkseitig kalibriert geliefert werden Eine Wiederholung der Kalibrierung nach Ablauf von 1 Jahr wird empfohlen PeakTech 01 2015...

Page 58: ... measuring works only in dry clothing and rubber shoes i e on isolating mats Never touch the tips of the test leads or probe Comply with the warning labels and other info on the equipment The measurement instrument is not to be to operated unattended Do not subject the equipment to direct sunlight or extreme temperatures humidity or dampness Do not subject the equipment to shocks or strong vibrati...

Page 59: ...the products High Voltage Please Consult the Manual Protective Ground End Earth End on the Shell Grounding End for Measurement To avoid body damage and prevent product and connected equipment dam This product can only be used in the specified applications Carefully read the following safety information before using the test tool Warning To avoid fire or electrical shock please use proper power ada...

Page 60: ... and accessories for mechanical damage and replace when damaged Remove all probes and accessories that are not in use Do not apply voltages that differ more than 400 V from earth ground to any input when measuring in a CAT II environment Do not apply voltages that differ more than 400 V from each other to the isolated inputs when measuring in a CAT II environment Do not apply input voltages above ...

Page 61: ...ommunicate with PC 3 LAN port optional Network port which can be used to connect with PC 4 Probe Compensation Measurement signal 3 3 1KHz output 5 MULTI input trigger output or Pass Fail output and EXT trigger input 6 Signal input of Channel 2 7 Signal input of Channel 1 Note If you use LAN port to communicate with PC the oscilloscope must be powered by the adapter 1 2 3 4 5 6 7 ...

Page 62: ...ower by PC USB or the adapter communicate with PC 2 Probe Compensation Measurement signal 5V 1KHz output 3 MULTI port EXT trigger input trigger output or Pass Fail output 4 Signal input of Channel 2 5 Signal input of Channel 1 2 3 4 5 1 ...

Page 63: ...LAN port to communicate with PC the oscilloscope must be powered by the adapter 4 How to Connect with PC 1 Install the Software on the supplied CD 2 Double click the created desktop shortcut PeakTech_VO 3 Connection Use the supplied USB cable to connect the USB port of the Oscilloscope in Fehler Verweisquelle konnte nicht gefunden werden to the USB port of a PC Note If you use a USB cable which is...

Page 64: ...X 2 Set the Switch in the Oscilloscope Probe as 10X and Connect the Oscilloscope with CH1 Channel Align the slot in the probe with the plug in the CH1 connector BNC and then tighten the probe with rotating it to the right side Connect the probe tip and the ground clamp to the connector of the probe compensator 3 Click on the right top of the software interface to auto set The square wave of 1 KHz ...

Page 65: ...pensated Compensated correctly Under compensated Displayed Waveforms of the Probe Compensation Repeat the steps mentioned if needed Adjust Probe The set values of the probe switch are 1X and 10X Attenuation Switch ...

Page 66: ...id any damage to the instrument or probe do not expose it to any sprays liquids or solvents Cleaning Inspect the instrument and probes as often as operating conditions require To clean the instrument exterior perform the following steps Wipe the dust from the instrument and probe surface with a soft cloth Disconnect the USB cable before cleaning your Oscilloscope Clean the instrument with a wet so...

Page 67: ...ied by us some problems such as connection error and signal disturbing might occur For Windows Vista or Windows 7 The Microsoft Windows Systems since Windows Vista or Windows 7 change a lot which require a new installation guide of USB driver Here it is During the whole installation please assure that the device is running well and plugged into PC from USB Right click Computer you can find it on t...

Page 68: ... 68 Right click the unknown device icon in the popup menu click Update Driver Software In the open window select Browse my computer for driver software ...

Page 69: ...ndow select a directory path for the driver software location and click Next Note The driver software location is a directory that is under the software setup folder named USBDRV and the contents inside are like these ...

Page 70: ...le and to the sys or dll file you can indicate then in different directories like x86 ia64 or amd64 depending on the C PU but most of time just x86 and amd64 are enough OK back to the driver installing after last Next step the system is installing driver software for you as follows ...

Page 71: ... x 86 may open a window named Windows Security as below and just select Install this driver software anyway to continue Or sometimes it for Windows 7 x 64 may open a window named Windows Security as below and just click Install to continue And then continue installing and finish ...

Page 72: ... opens with information Windows has successfully updated you r driversoftware Close the window have a look at the Computer Management window you will find a device under LibUSB Win32 Devices it should be like this Now the USB driver will work ...

Page 73: ...ell device to open Found New Hardware Wizard dialog Or you can right click My Computer and select Manage in the left area of opened Computer Manage ment select Device Manager double click the item USB Device with in the middle area to open the Wizard In the Wizard select No not this time ...

Page 74: ...earch for the best driver in these locations then select Include this location in the search and indicate a directory location for USB driver which is named as USBDRV and under the directory where you installed the program at Then the installation is running And complete ...

Page 75: ... installed in Device Management Now you can use the program and use if for USB communication If there is a nearly version of USB driver in your computer you could try running reinstall bat to fix the file is under the directory of USBDRV ...

Page 76: ...to Use Executive Buttons 8 Run Stop see 20 How to Use Executive Buttons 9 Single Trigger see 20 How to Use Executive Buttons 10 Back to Home menu 11 Hide the menu 12 Measure voltage with cursors 13 The red pointer shows the trigger level position for CH1 yellow for CH2 It can be dragged up and down 14 Function menu click to show hide 15 Shortcut icon of resetting to factory settings see Default Sh...

Page 77: ...ed Error Error occurred ReSyncing Synchronize with the oscilloscope again AutoSet In process of auto setting The statuses of not staying connected with the oscilloscope Offline No connection with the oscilloscope USBFound Find available USB devices USBDrvErr USB drivers installed error MachineNotSupport Can not recognize the device The statuses in the process of connecting with the oscilloscope Li...

Page 78: ... is changed again Caution The default attenuation coefficient of the probe in the software is preset to10X Make sure that the set value of the attenuation switch in the probe is the same as the menu selection of the probe attenuation coefficient in the software The set values of the probe switch are 1X and 10X Figure Attenuation Switch Caution When the attenuation switch is set to 1X the probe wil...

Page 79: ... measure frequencies from 2 Hz to the full bandwidth Only if the measured channel has triggering signal and in Edge mode it can measure frequency correctly In the Single trigger mode it is a one channel cymometer and it can only measure the frequency of the triggering channel In the ALT trigger mode it is a two channel cymometer and it can measure the frequency of two channels Keyboard Shortcuts Q...

Page 80: ...nding parameters of horizontal system in the Capture period window 17 in 7 User Interface The further from the center the higher the changing speed is See also 16 How to zoom the waveform Keyboard Shortcuts Time base decreases one level Time base increases one level ...

Page 81: ...pes single trigger and alternate trigger Single trigger Use a trigger level to capture stable waveforms in two channels simultaneously Alternate trigger Trigger on non synchronized signals The Single Trigger and Alternate Trigger menus are described respectively as follows Single trigger 1 In Trigger menu choose Single the chosen label has a mark 2 Choose Source 3 Choose Mode Single trigger has fo...

Page 82: ... the first digit Eg When is chosen click the last digit will be increased by 1 Click Reset to set Holdoff time as default value 100 ns Set in Trigger window Force Force to create a trigger signal and the function is mainly used in Normal and Single mode Video Trigger Choose Mode as Video to trigger on fields or lines of NTSC PAL or SECAM standard video signals 1 Select video modulation NTSC PAL or...

Page 83: ...igger modes for different channels The options are as follows edge video pulse or slope Set in Trigger menu 1 In Trigger menu choose Alternate the chosen label has a mark 2 Choose Source 3 Choose Mode About the setting method of Mode and parameters refer to Single trigger Note In alternate trigger only one channel at most can be set as Video mode You cannot choose Video in both channels Set in Tri...

Page 84: ...he earth potential Click to choose CH1 or CH2 check Opposite to invert the waveform uncheck it to display normally To set channel coupling DC Pass both AC and DC components of the input signal AC Block the DC component of the input signal Ground Disconnect the input signal You can also click the coupling mode to switch it See figure below 9 1 To adjust the probe attenuation For correct measurement...

Page 85: ...ly 10 2 Remove measurement Uncheck the measurement type to remove it Click Remove All to remove all measurements 10 3 The automatic measurement of voltage parameters The PeakTech oscilloscopes provide automatic voltage measurements including Vpp Vmax Vmin Vavg Vamp Vrms Vtop Vbase Overshoot and Preshoot Figure below shows a pulse with some of the voltage measurement points Vpp Peak to Peak Voltage...

Page 86: ...ulse in the waveform takes to rise from 10 to 90 of its amplitude Fall Time Time that the falling edge of the first pulse in the waveform takes to fall from 90 to 10 of its amplitude Width The width of the first positive pulse in 50 amplitude points Width The width of the first negative pulse in the 50 amplitude points Delay 1 2 The delay between the two channels at the rising edge Delay 1 2 The d...

Page 87: ...ction of the jamming burr and the possibility of reducing the confusion Average It is used to reduce the random and don t care noises with the editable number 1 number 128 of averages Figure Normal sampling mode display in which no burr can be detected Figure Peak Detect mode under which the burrs on the falling edge of the square wave can be detected and the noise is heavy Figure The displayed wa...

Page 88: ... cursor measurement Tick on Time option then two light red lines display along the vertical direction of the screen which represent Cursor 1 and Cursor 2 Place the mouse pointer over Cursor 1 or Cursor 2 drag after the mouse pointer changing to adjust the positions of Cursor 1 and Cursor 2 according to the waveform to be measured The cursor increment window at the left bottom shows current time of...

Page 89: ...ee Figure below 12 4 The cursor measurement for FFT mode Check measurement type choose either Amplitude Measurement or Frequency Measurement at the mode of FFT or both at the same time Frequency Measurement Tick on Frequency enter Home page Math FFT then two light red lines represent Cursor 1 and Cursor 2 show along the vertical direction on FFT window Place the mouse pointer over Cursor 1 or Curs...

Page 90: ...and Cursor 2 according to the waveform to be measured The cursor increment window on the bottom left of FFT shows current position of the two cursors absolute voltage amplitude difference of the two cursors See Figure below 13 How to Set the Display System Click to show Function menu choose Display Display type Click the button to choose to choose the display type the chosen button has a mark Vect...

Page 91: ...ersistence display effect of the picture tube oscilloscope can be simulated The reserved original data is displayed in fade color and the new data is in bright color Different persistence time can be chosen Off 0 5 second 1 second 2 seconds 5 seconds and Infinite When the Infinite option is set the measuring points will be stored till the controlling value is changed Click Clear button the persist...

Page 92: ... and Channel 2 and the FFT operation of Channel 1 or Channel 2 Math of dual waveform 1 Turn on CH1 and CH2 2 Click to show Function menu choose Math Check Math 3 Choose the factors and operator Select the voltage division of M The software transforms the waveforms data of the two factors into selected voltage division of M and calculate The green calculated waveform M is displayed in the screen ...

Page 93: ...es FFT function in this oscilloscope transforms 2048 data points of the time domain signal into its frequency components mathematically and the final frequency contains 1024 points ranging from 0 Hz to Nyquist frequency Taking the FFT operation for example the operation steps are as follows 1 Click to show Function menu choose Math Check FFT The user interface is switched into Three View mode 2 Se...

Page 94: ...tude Best type for measuring the frequency spectrum of non repetitive signals and measuring frequency components near DC Recommend to use for Transients or bursts the signal level before and after the event are nearly equal Equal amplitude sine waves with frequencies those are very close Broadband random noise with a relatively slow varying spectrum Blackman Best solution for magnitude worst for f...

Page 95: ... 95 The figures below are examples for measuring sine wave with a frequency of 1 kHz under the selection of four different window for FFT Hamming window Rectangle window Blackman window Hanning window ...

Page 96: ...ncies above the Nyquist frequency will be under sampled which causes aliasing So pay more attention to the relation between the frequency being sampled and measured Note In FFT mode only support to set coupling voltage division and zero position in Channel window other operations are prohibited If you want to enter other menus uncheck FFT first 16 How to zoom the waveform Click to show Function me...

Page 97: ...ed to the full screen size You can adjust the time base W and horizontal trigger position Tw of zoom window in the menu You can also drag the red pointer to adjust the horizontal trigger position 3 in 7 User Interface You can also switch and set in Capture Period window as shown in the below ...

Page 98: ... or pass signal by built in and configurable output port 17 1 To run the test read the following steps 1 Choose channel Options include CH 1 CH 2 and Math 2 Set horizontal tolerance and vertical tolerance You can adjust by or input directly 3 Click Create Rule 4 Set output type Choose Pass or Fail Check or uncheck Ring Message Show and Stop Once as needed Ring The bell rings when it meets the rule...

Page 99: ...on menu choose Record 18 1 Record 1 Choose Record tab on the top 2 Click Preset Save Path to specify the save location Waveform files have the suffix cap 3 Set Interval Time and End Frame Interval Time refers to the interval between recorded frames Interval Time ranges from 0 to 100000 ms and increment by 10 ms 4 Make the waveform in Run status 5 Click Begin Record Counter starts to count the fram...

Page 100: ...eplaying 5 Check Cycle to play back the waveform in a loop Uncheck it to replay just once 6 Click Play Numbers shows the current frame number that is replayed 7 Click Pause to pause it 8 Drag slider to display the frame you need Note 1 If the software is in Run status playing back will make the current acquisition stopped 2 When replaying entering other menu items will make the replay paused ...

Page 101: ...color uncheck to display preview paper background Set Preview Page Background Click to display color picker dialog box to set the color of preview page background Save Image Click to save screen shot as an image file in png bmp or gif format Save Refer This function allows to store 8 reference waveforms These waveforms can be displayed with the current waveform simultaneously The recalled waveform...

Page 102: ...ady click Self Cal Default Click Default to call out the factory settings You can also click icon on the user interface Help Open this help document You can press F1 as shortcut key 20 Network The oscilloscope could be connected to a PC via LAN port For more details see 22 Use LAN Port Tips Window Choose a label among Channel Capture Period or Trigger to view the guide for the new users Click anyw...

Page 103: ...wn as the following table Function items Setting Acquisation Mode Current Vertical Coupling DC Vertical Scale Adjust to the proper division Horizontal Level Middle Horizontal Scale Adjust to the proper division Trigger Type Current Trigger Source Show the minimum number of channels Trigger Coupling Current Trigger Slope Current Trigger Level Mid point Setting Trigger Mode Edge Display Format YT Ru...

Page 104: ...he supplied USB cable to connect the oscilloscope with a PC through their USB ports After connecting successfully click to show Function menu choose Utility click Network 2 Set IP and port of the oscilloscope In Network menu click OK to enter Machine Net Setting Choose the oscilloscope to be connected from the list If it s not in the list click Refresh to refresh the list Set IP the first three by...

Page 105: ...e adapter into an electrical outlet to power the oscilloscope 2 Connection Plug in the LAN line to the LAN port of the oscilloscope plug the other end into the LAN interface of the computer 3 Set parameters in menu Click to show Function menu choose Utility click Network Set IP and Port to the same value of the oscilloscope in step 2 4 Click Connect ...

Page 106: ...rs of the computer View the network parameters of the computer to connect to The Default gateway and Subnet mask should be set according to the router Assume the parameters are as below IP address 192 168 1 71 Subnet mask 255 255 255 0 Default gateway 192 168 1 1 ...

Page 107: ...f the oscilloscope In Network menu click OK to enter MachineNetSetting Choose the oscilloscope to be connected from the list If it s not in the list click Refresh to refresh the list Set IP the first three bytes should be same as the IP address of the computer in step 1 the last byte should be different Here we set it to 192 168 1 72 The range of the port value is 0 4000 here we set it to 3000 The...

Page 108: ... adapter into an electrical outlet to power the oscilloscope 2 Connect to the router Use a LAN cable to connect the oscilloscope with a router the computer should be connected to the router too 3 Set parameters in menu Click to show Function menu choose Utility click Network Set IP and Port to the same value of the oscilloscope in step 2 4 Click Connect ...

Page 109: ...P 1325 60 MHz P 1330 100 MHz Channel P 1280 1285 1290 2 1 External P 1325 1330 4 1 Extern Acquisition Mode Normal Peak detect Averaging Sample rate real time P 1280 Dual CH 250 MSa s Single CH 500 MSa s P 1285 Dual CH 500 MSa s Single CH 1 GSa s P 1290 Dual CH 100 MSa s Single CH 100 MSa s P 1325 Four CH 125 MSa s Dual CH 250 MSa s Single CH 500 MSa s P 1330 Four CH 250 MSa s Dual CH 500 MSa s Sin...

Page 110: ... div P 1280 1290 1325 5 ns div 100 s div step by 1 2 5 P 1285 1330 2 ns div 100 s div step by 1 2 5 Sampling rate relay time accuracy 100 ppm Interval T Accuracy DC 100 MHz Single 1 interval time 100 ppm x reading 0 6 ns Average 16 1 interval time 100 ppm x reading 0 4 ns Vertical system A D converter P 1280 P 1285 P 1290 8 bits resolution 2 Channels simultaneously P 1325 P 1330 8 bits resolution ...

Page 111: ...al 5 div from the screen center Trigger level Accuracy typical Internal 0 3 div EXT TTL Trigger displacement According to Record length and time base Trigger Holdoff range 100 ns 10 s 50 level setting typical Input signal frequency 50 Hz Edge trigger slope Rising Falling Pulse trigger Trigger condition Positive pulse Negative pulse Pulse Width range 30 ns 10 s Video Trigger Modulation NTSC PAL and...

Page 112: ...1325 1330 5 W P 1290 2 5 W Environment Temperature Working temperature 0 C 40 C Storage temperature 20 C 60 C Relative Humidity 90 Height Operating 3 000 m Non operating 15 000 m Cooling Method Natural convection Mechanical Specifications Dimensions WxHxD P 1280 1285 P 1325 1330 120 x 18 x 190 mm P 1290 120 x 18 x 170 mm Weight 260 g Enclosure Standard Accessories Passive probes 1 2 m 1 1 10 1 swi...

Page 113: ...technical knowing Technical changes which are in the interest of progress reserved We herewith confirm that the units are calibrated by the factory according to the specifications as per the technical specifications We recommend to calibrate the unit again after one year PeakTech 01 2015 Po PeakTech Prüf und Messtechnik GmbH Kornkamp 32 DE 22926 Ahrensburg Germany 49 0 4102 42343 44 49 0 4102 434 ...