PeakTech 1235 Operation Manual Download Page 21

 

The  trigger  cursor  is  the  same  as  the  above  description  in  data  listing  interface, 
displaying  in  a  row  of  red  data.  The  left  line  is  address  value  of  cursor  in 
Decimalization, middle  line  is data  value of cursor in Hex,  and the  Bin  data  value 
of cursor is on the right line. 
 

5.9 Save/Recall  

 
The  use  of  the  logic  Analyzer  is  complicated,  there  is  a  lot  of  parameters  to  set, 
through  the  descriptions  in  the  foregoing  sections,  we  can  know  that  if  some 
parameters  set  improper,  the  sample  doesn’t  start,  even  it  starts,  maybe  it  can’t 
sample  the  data  we  need.  And  even  it  samples  the  data  we  need,  it  may  be  not 
easy  to  analyze.  Therefore,  we  need  to  read  the  descriptions  of  foregoing 
sections  to  understand  the  settings  of  these  parameters.  The  logic  analyzer  isn’t 
as  popular  as  the  oscilloscopes;  the  operations  of  the  logic  analyzer  may  have 
big  differences  from  different  manufacturers,  for  users’  convenience,  this  
instrument sets an internal code generator and a set of typical default parameters, 
which can demonstrate the various functions of the instrument better. 
 

5.9.1. Parameters storage 

The  application  area  of  the  logic  analyzer  is  broad;  the  parameters  settings  may 
be  different  in  various  application  occasions,  so  just  using  the  default  settings  is 
obviously  not  enough.  To  set  parameters  users  need  every  time  repeatedly  is 
very  troublesome  in  each  use,  so  it  sets  parameter  “save”  function  and  save  the 
present  parameters  settings  of  the  instrument  for  the  further  use

press

save

key. The instrument will query firstly “Store? 0: prameter, 1: waveform, 2: cancel”. 
Press

0

key  to  save  the  all  currently  setting  parameters

even  not  lose  when 

power-off.  
 

5.9.2. Waveform storage 

In  practical  application,  catching  an  interesting  sample  data  is  not  so  easy  for 
some  little  probability  affairs.  So  user  can  set  waveform  storage  function  to  save 
caught sample data for further analysis. The length of wave is 16348 sample data. 
Start  address  is  current  displayed  window  address.  User  can  change  the 
displayed  window  address  by  method  of  setting  parameters  or  moving  figure. 
Then  press

save

key.  The  instrument  will  query  firstly  “Store?  0:  prameter,  1: 

waveform,  2:  cancel”.  Press

1

save  data  with  length  of  16k,  current  displayed 

window as start address.  
 

5.9.3. Cancel the storage 

If user press the

save

key by mistake and need not  to save the current setting 

parameter  or  waveform,  you  can  press

2

key  cancel  operation.  Original 

parameters or waveform won’t be destroyed. 
 

5.9.4. Recall 

Press

Recall

to recall the saved parameters and waveform data. After recalling, 

you  can  display  them  with  window  and  data  list,  also  can  measure  and  analyze 
them in different ways. Recalled wave always exits except that user sample again. 
But  after  pressing

Single

or

Run/Stop

,  new  data  will  be  shown.  If  needed, 

saved waveform data will be recalled by pressing

Recall

again. 

 
 
 
 

-20- 

Summary of Contents for 1235

Page 1: ...PeakTech 1235 Operation manual 32 Chanel Logic Analyser ...

Page 2: ... find and solve without the logic analyzer such as transmission delay competition risk and burr interference At present digital circuit and bus technique are used in many apparatuses In order to analyze and validate the result of information processing the logic analyzer must be used to find out the error in programming and running measure and compare the state of digital logic circuit With the ra...

Page 3: ...comparator A TTL level digital signal is generated by the comparison of the two signals and then the digital signal is stored synchronously in the data flip latch by sampling clock When sample the internal code the code generator will produce 30 channels of internal digital signals which are stored synchronously in the data flip latch by sampling clock The sampling data in the data flip latch are ...

Page 4: ... diagram 3 Read write data Sampling circuit Sampling rate read write data Keyboard Input Storage address generator High speed comparator Threshold voltage High speed memory MCU Trigger process External signal Clock circuit Internal code generator ...

Page 5: ...keys 4 Data input keys 5 Cursor key 6 Sample key 7 Function keys 8 Signal input 9 Signal input 10 Adjusting knob AC 100 240V 45 65Hz 1A REPLACE FUSE AS SPECIFIED DISCONNECT POWER CORD BEFORE REPLACING FUSE RS 232 USB 3 2 1 Rear panel 1 Power source outlet 2 RS232 interface 3 USB Device interface 4 1 2 3 4 5 6 7 8 9 10 ...

Page 6: ... Number input 11 keys 0 1 2 3 4 5 6 7 8 9 number input key x special characters x only used for data search input 4 2 4 Input control 4 keys select the setting parameters up to down in cycle backspace to delete the input data when the input hasn t finished Shift used for input the English letters above the button 4 2 5 Display control 5 keys Display display timing waveforms or data lists in cycle ...

Page 7: ...eters on the left is the state parameters changes automatically with different operations showing the current state of the instrument The two parameters on the right is the parameters settings of the system can be set with numeric keys 4 3 3 Trigger settings interface The trigger settings interface displays the whole trigger process Adopting the graphical mode makes users can easy to understand th...

Page 8: ...ease the key then input the letter you need If there is no letter on the key user can input space The shift key is a one kick So if the user needs to input the next letter please repeat the operations above 5 3 Channels setting 5 3 1 Channel order setting The double digits number 00 31 on the most left of the waveforms display frame is the channel serial number Press Channel key to select the chan...

Page 9: ...ltage the analyzer shows the number 1 otherwise it shows the number 0 Then the signals are sampled and saved and displayed on the screen Waveforms displayed by this means only reflect the timing logic when the input signals are higher or lower than the threshold voltage It doesn t reflect the real amplitude of the input signals and may be far different from the real waveforms of the input signals ...

Page 10: ...el is at the bottom This is to enable the highest order digit of the byte to be on the top and the lowest order digit on the bottom 5 5 2 Waveform rolling Up Down The waveform rolling can be used to observe all the waveforms in 32 channels Press and turn the knob to browse the waveforms If there is a need to compare the waveforms the method of channel order setting 5 3 1 can be used to display the...

Page 11: ...direction Zoom ns div in the sixth row on the left bottom of the wave frame will change this ratio coefficient stands for time amount that each lattice of scale line represents in the above and underside two rows of waveform frame The essence of waveform amplification is that a number in sampling memory is displayed continuously with several points in waveform display So the amplifying waveform ca...

Page 12: ...inuously from the top to the bottom The middle is the address value of the sampling data in Hex the right one is the Bin address value a parting line between each 8 bits Bin code so as to read conveniently Use two different colors to distinguish the two adjacent lines making display clear beautiful and not easy to confuse The storage depth of the instrument is 260 000 storing addresses but the scr...

Page 13: ...ncide with each other their parameters are 0 If want to measure the time difference between two points in present waveform interface can move cursor1 and cursor2 to the two aim points respectively it s easy to read the time difference or the address difference between the two points But if the address difference between the two points to be measured exceeds 280 the two points can t be displayed in...

Page 14: ...t result may be more difficult Because the Inner Code Generator s logic relation is simple waveform is pure and standard one can quickly master logic analyzer s use with it Second in practical application if the test result is different from expectant one sometimes it is hard to make clear that the problem comes from the tested circuit or logic analyzer here as long as use inner code generator to ...

Page 15: ...e discrete event that is take the event sequences as independent variable the data listing after sampling reflects the logic state relation between the system clock and the other signals in the system this mode is known as state analysis Here the sampling clock is synchronous with the tested system it is also called synchronous sampling If take sample using inner clock to inner code generator this...

Page 16: ... sampling cycle in order to observe the particular changes of the tested signals that is to increase the sampling rate Generally speaking the sampling cycle should be less 3 5 times than the narrowest pulse width of the tested signals In other words even the narrowest pulse of the tested signals should include three sampling points at least which can truly reflect input signals change as time The ...

Page 17: ...y which requires corresponding sampling velocity for logic analyzer However the space of memory in the instrument is limited so in practical the effective sampling time is very short it can say sampling process completes instantaneously If start up sampling process manually this needs to press Single in an extreme accurate time but it is too hard to do moreover the data needed sampling and storing...

Page 18: ...n the input data streams in other words when the logic level of 00 15 channels is 0001001000110100 the sampling process starts 5 8 3 Start select The start select can be set with numbers the key 0 cut the switch the key 1 connects the switch If the start select switch is connected the start conditions will be short circuit having no use In other words after pressing Single the sampling process sta...

Page 19: ... that it captures a trigger event In some applications the trigger event we care may appear many times and the thing we interested in is the situations after the trigger event appears many times If we can capture the trigger event for many times in one sampling process that may be more convenient for analysis So the instrument sets a trigger event counter after the sampling process starts the coun...

Page 20: ...nals appear and displays sample is processing press any key to stop To release from this state just press any key to stop the sampling process manually Then one must study the tested signals carefully reset the trigger process to make sure the sampling process can run normally 5 8 9 Trigger cursor Sometimes there may be one or several vertical red lines in the timing waveforms interface they are t...

Page 21: ...repeatedly is very troublesome in each use so it sets parameter save function and save the present parameters settings of the instrument for the further use press save key The instrument will query firstly Store 0 prameter 1 waveform 2 cancel Press 0 key to save the all currently setting parameters even not lose when power off 5 9 2 Waveform storage In practical application catching an interesting...

Page 22: ...s Input Channel 32 data sample channels two external clock channels Threshold voltage 6 independently adjustable threshold voltages Adjusting range 6V to 6V Resolution 0 1V Input impedance Resistance 100kΩ Capacitance 8pf Input range 500mVpp to 20Vpp Input protection maximum input voltage 40V no damage 6 2 Sample Storage characteristics Timing Sample internal clock Sampling Rate 1Hz to 100MHz 10ns...

Page 23: ...gether with it Cursor measure move the position of the cursor can dynamically display the data values the address distance and the interval of the two cursors Trigger cursor the sample point accords with the trigger conditions Search cursor the sample point accords with the search conditions 6 6 Internal code generator Code Type 00 to 15 channels are counters with adding one 16 to 29 channels are ...

Page 24: ...al knowing Technical changings which are in the interest of progress reserved We herewith confirm that the units are calibrated by the factory according to the specifications as per the technical specifications We recommend to calibrate the unit again after one year PeakTech 05 2012 Ho PeakTech Prüf und Messtechnik GmbH Kornkamp 32 DE 22926 Ahrensburg Germany 49 0 4102 42343 44 49 0 4102 434 16 in...

Reviews: