PeakTech 1235 Operation Manual Download Page 19

 

The  start  “bit-select”  can  be  set  with  numbers  in  Hex,  representing  32  input 
channels, if set  bit-select to  0, shows this channel can be  ignored,  without match 
checking, and has  no influence on start  whatever signal level  of this channel  is  0 
or  1.  If  set  bit-select  to  1,  shows  the  channel  is  effective,  and  must  be  made 
matching checking. The default setting of start bit-select is 0000FFFF, means that 
only check 00~07 channels, ignore 08~31 channels. 
 
The  start  “compare  word”  can  be  set  with  numbers  in  Hex,  the  default  setting  of 
start  compare  word  is  00000069  and  means  once  Hex  number  69  appears  in 
00~07  channels  in  the  input  data  streams,  in  other  words  when  the  logic  level  of 
00~07 channels is “01101001”

the trigger conditions is satisfied. 

 
Different  from  start  conditions,  the  trigger  conditions  sets  three  trigger  limit 
switches, <, =, >, which are useful for testing the data fluctuant limit to the tested 
signals.  The  trigger  limit  switch  can  be  set  with  number  keys,  press 

1

to 

connect  the  switch,  but  only  one  is  connected  in  the  three  trigger  limit  switch, 
once  one  switch  is  connected,  the  other  two  switches  are  disconnected.  The 
default setting of the trigger limit switch is "=" connecting. 
 

5.8.5. Event count 

When  the  sampling  process  starts,  the  instrument  starts  to  sample  the  input 
signals, and stores the sampling data in the memory, at the same time, compares 
the  sampling  data  with  the  trigger  conditions  and  the  trigger  limit  switch,  if  the 
sampling  data  satisfy  the  trigger  conditions  and  the  trigger  limit  switch,  means 
that  it  captures  a  trigger  event.  In  some  applications,  the  trigger  event  we  care 
may  appear  many  times,  and  the  thing  we  interested  in  is  the  situations  after  the 
trigger  event  appears  many  times.  If  we  can  capture  the  trigger  event  for  many 
times in one sampling process,  that may  be more convenient for analysis. So the 
instrument  sets  a  trigger  event  counter,  after  the  sampling  process  starts,  the 
count  value  firstly  resets,  then  the  count  value  adds  one  when  meets  the  trigger 
event once, till the count value reaches the setting value of the trigger event , the 
trigger process finishes. 
 
The  trigger  events  count  can  be  set  with  numbers  in  Dec,  the  setting  range  is 
1~999, the default setting of the trigger event count value is 001. 
 

5.8.6. Trigger select 

The trigger select switch can  be set  with  numbers, the key

0

cuts the  switch

the  key

1

connects  the  switch.If  the  trigger  select  switch  is  connected,  the 

trigger  conditions,  the  trigger  limit  switch  and  the  event  counter  are  all  short-
circuit,  having  no  use.In  other  words,  after  sampling  process  starts,  the 
instrument  does not detect the trigger conditions, also does  not count the  event , 
the trigger process finishes unconditionally. 
 
The  default  setting  of  the  “start  select  switch”  is  “connect”  in  order  to  make 
random  sampling  drilling  justly,  and  has  no  specific  sampling  purpose  on 
capturing  what  kind  of  signal,  not  need  to  set  the  trigger  conditions  and  event 
count  too.  If  the  setting  of  the  start  select  switch  is  “cut”,  then  only  understand 
the tested signal clearly and have specific sampling purpose, can set the suitable 
trigger  conditions,  trigger  limit  switch  and  event  count.  Otherwise,  if  sets  these 
parameters unsuitably, the trigger process may not finish. 
 
 

-18- 

Summary of Contents for 1235

Page 1: ...PeakTech 1235 Operation manual 32 Chanel Logic Analyser ...

Page 2: ... find and solve without the logic analyzer such as transmission delay competition risk and burr interference At present digital circuit and bus technique are used in many apparatuses In order to analyze and validate the result of information processing the logic analyzer must be used to find out the error in programming and running measure and compare the state of digital logic circuit With the ra...

Page 3: ...comparator A TTL level digital signal is generated by the comparison of the two signals and then the digital signal is stored synchronously in the data flip latch by sampling clock When sample the internal code the code generator will produce 30 channels of internal digital signals which are stored synchronously in the data flip latch by sampling clock The sampling data in the data flip latch are ...

Page 4: ... diagram 3 Read write data Sampling circuit Sampling rate read write data Keyboard Input Storage address generator High speed comparator Threshold voltage High speed memory MCU Trigger process External signal Clock circuit Internal code generator ...

Page 5: ...keys 4 Data input keys 5 Cursor key 6 Sample key 7 Function keys 8 Signal input 9 Signal input 10 Adjusting knob AC 100 240V 45 65Hz 1A REPLACE FUSE AS SPECIFIED DISCONNECT POWER CORD BEFORE REPLACING FUSE RS 232 USB 3 2 1 Rear panel 1 Power source outlet 2 RS232 interface 3 USB Device interface 4 1 2 3 4 5 6 7 8 9 10 ...

Page 6: ... Number input 11 keys 0 1 2 3 4 5 6 7 8 9 number input key x special characters x only used for data search input 4 2 4 Input control 4 keys select the setting parameters up to down in cycle backspace to delete the input data when the input hasn t finished Shift used for input the English letters above the button 4 2 5 Display control 5 keys Display display timing waveforms or data lists in cycle ...

Page 7: ...eters on the left is the state parameters changes automatically with different operations showing the current state of the instrument The two parameters on the right is the parameters settings of the system can be set with numeric keys 4 3 3 Trigger settings interface The trigger settings interface displays the whole trigger process Adopting the graphical mode makes users can easy to understand th...

Page 8: ...ease the key then input the letter you need If there is no letter on the key user can input space The shift key is a one kick So if the user needs to input the next letter please repeat the operations above 5 3 Channels setting 5 3 1 Channel order setting The double digits number 00 31 on the most left of the waveforms display frame is the channel serial number Press Channel key to select the chan...

Page 9: ...ltage the analyzer shows the number 1 otherwise it shows the number 0 Then the signals are sampled and saved and displayed on the screen Waveforms displayed by this means only reflect the timing logic when the input signals are higher or lower than the threshold voltage It doesn t reflect the real amplitude of the input signals and may be far different from the real waveforms of the input signals ...

Page 10: ...el is at the bottom This is to enable the highest order digit of the byte to be on the top and the lowest order digit on the bottom 5 5 2 Waveform rolling Up Down The waveform rolling can be used to observe all the waveforms in 32 channels Press and turn the knob to browse the waveforms If there is a need to compare the waveforms the method of channel order setting 5 3 1 can be used to display the...

Page 11: ...direction Zoom ns div in the sixth row on the left bottom of the wave frame will change this ratio coefficient stands for time amount that each lattice of scale line represents in the above and underside two rows of waveform frame The essence of waveform amplification is that a number in sampling memory is displayed continuously with several points in waveform display So the amplifying waveform ca...

Page 12: ...inuously from the top to the bottom The middle is the address value of the sampling data in Hex the right one is the Bin address value a parting line between each 8 bits Bin code so as to read conveniently Use two different colors to distinguish the two adjacent lines making display clear beautiful and not easy to confuse The storage depth of the instrument is 260 000 storing addresses but the scr...

Page 13: ...ncide with each other their parameters are 0 If want to measure the time difference between two points in present waveform interface can move cursor1 and cursor2 to the two aim points respectively it s easy to read the time difference or the address difference between the two points But if the address difference between the two points to be measured exceeds 280 the two points can t be displayed in...

Page 14: ...t result may be more difficult Because the Inner Code Generator s logic relation is simple waveform is pure and standard one can quickly master logic analyzer s use with it Second in practical application if the test result is different from expectant one sometimes it is hard to make clear that the problem comes from the tested circuit or logic analyzer here as long as use inner code generator to ...

Page 15: ...e discrete event that is take the event sequences as independent variable the data listing after sampling reflects the logic state relation between the system clock and the other signals in the system this mode is known as state analysis Here the sampling clock is synchronous with the tested system it is also called synchronous sampling If take sample using inner clock to inner code generator this...

Page 16: ... sampling cycle in order to observe the particular changes of the tested signals that is to increase the sampling rate Generally speaking the sampling cycle should be less 3 5 times than the narrowest pulse width of the tested signals In other words even the narrowest pulse of the tested signals should include three sampling points at least which can truly reflect input signals change as time The ...

Page 17: ...y which requires corresponding sampling velocity for logic analyzer However the space of memory in the instrument is limited so in practical the effective sampling time is very short it can say sampling process completes instantaneously If start up sampling process manually this needs to press Single in an extreme accurate time but it is too hard to do moreover the data needed sampling and storing...

Page 18: ...n the input data streams in other words when the logic level of 00 15 channels is 0001001000110100 the sampling process starts 5 8 3 Start select The start select can be set with numbers the key 0 cut the switch the key 1 connects the switch If the start select switch is connected the start conditions will be short circuit having no use In other words after pressing Single the sampling process sta...

Page 19: ... that it captures a trigger event In some applications the trigger event we care may appear many times and the thing we interested in is the situations after the trigger event appears many times If we can capture the trigger event for many times in one sampling process that may be more convenient for analysis So the instrument sets a trigger event counter after the sampling process starts the coun...

Page 20: ...nals appear and displays sample is processing press any key to stop To release from this state just press any key to stop the sampling process manually Then one must study the tested signals carefully reset the trigger process to make sure the sampling process can run normally 5 8 9 Trigger cursor Sometimes there may be one or several vertical red lines in the timing waveforms interface they are t...

Page 21: ...repeatedly is very troublesome in each use so it sets parameter save function and save the present parameters settings of the instrument for the further use press save key The instrument will query firstly Store 0 prameter 1 waveform 2 cancel Press 0 key to save the all currently setting parameters even not lose when power off 5 9 2 Waveform storage In practical application catching an interesting...

Page 22: ...s Input Channel 32 data sample channels two external clock channels Threshold voltage 6 independently adjustable threshold voltages Adjusting range 6V to 6V Resolution 0 1V Input impedance Resistance 100kΩ Capacitance 8pf Input range 500mVpp to 20Vpp Input protection maximum input voltage 40V no damage 6 2 Sample Storage characteristics Timing Sample internal clock Sampling Rate 1Hz to 100MHz 10ns...

Page 23: ...gether with it Cursor measure move the position of the cursor can dynamically display the data values the address distance and the interval of the two cursors Trigger cursor the sample point accords with the trigger conditions Search cursor the sample point accords with the search conditions 6 6 Internal code generator Code Type 00 to 15 channels are counters with adding one 16 to 29 channels are ...

Page 24: ...al knowing Technical changings which are in the interest of progress reserved We herewith confirm that the units are calibrated by the factory according to the specifications as per the technical specifications We recommend to calibrate the unit again after one year PeakTech 05 2012 Ho PeakTech Prüf und Messtechnik GmbH Kornkamp 32 DE 22926 Ahrensburg Germany 49 0 4102 42343 44 49 0 4102 434 16 in...