PeakTech 1235 Operation Manual Download Page 11

 

Parameter  in  second  row  on  the  right  bottom  of  the  wave  frame  is 

scroll-step, 

turn  knob  one  step;  the  window-address  will  increase  a  scroll-step  value.  The 
higher  the  value  of  scroll-step  is,  the  faster  the  speed  of  waveform  rolling  is,  but 
this  may  miss  needed  part;  the  lower  the  value  of  scroll-step  is,  the  slower  the 
speed  of  rolling  is,  and  the  finer  observe  the  waveforms.  Press

System

to 

select 

scroll-step

,  the  value  of  the  scroll-step  can  be  set  by  number  keys  with 

the range of 1~ 260 000, after the setting, press

←→

 and turn the knob to roll 

the waveform left and right, so that the speed of waveform rolling will change. 
 
Setting  the  window-address  with  number  keys  directly  enables  display  window  to 
locate  in  the  data  block  needing  to  visit,  and  rolling  waveforms  with  knob  turning 
can observe the change situations of a section of waveform continuously. 
 

5.5.4. Waveform zoom 

In  digital  systems,  the  changing  rate  of  the  logic  level  may  be  very  different  in 
different  data  channels,  in  a  timing  waveform  displaying  graph;  it  may  include  8 
channels with big difference in changing rate of logic level. This brings a problem, 
that  is,  a  large  number  of  pulse  waveforms  crowds  together  in  channel  with  fast 
changing  rate  of  logic  level  and  cannot  be  seen  clearly,  but  in  the  channel  with 
slow  changing  rate,  it  is  a  line  on  screen.  To  solve  this  problem,  can  stretch  the 
waveforms in level direction and unfold the crowded pulse wave to visit waveform 
characteristics  with  fast  changing,  also  can  compress  the  waveforms  and  gather 
sparse  pulse  waveform  together  to  visit  waveform  characteristics  with  slow 
changing.  
 
Press

Zoom

key  and  turn  the  knob  to  make  the  display  waveforms  stretch  or 

compress  in  horizontal  direction. 

“Zoom=  ns/div” 

in  the  sixth  row  on  the  left 

bottom of the wave frame will change, this ratio coefficient stands for time amount 
that each  lattice  of scale line represents,  in the  above  and  underside two  rows of 
waveform  frame.  The  essence  of  waveform  amplification  is  that  a  number  in 
sampling  memory  is  displayed  continuously  with  several  points  in  waveform 
display.    So,  the  amplifying  waveform  can  make  dense  waveform  be  seen  more 
clearly  and  let  measurement  become  easy,  it  won’t  generate  any  distortion  to 
waveform.  But  waveform  amplification  is  unable  to  increase  resolution;  the 
original  invisible  contents  are  still  not  seen  after  amplification.  If  want  to  observe 
the  detailed  change,  need  to  increase  sampling  velocity.    The  essence  of 
compression is taking one number to display from several data points in sampling 
memory,  therefore  after  compression,  waveform’s  observing  scope  is  broadened, 
but  may  lead  to  distortion  and  miss  small  changing  part,  more  compression  rate 
results in more serious distortion. So compression should be made gradually, one 
should  not  continue  compressing  when  finds  distortion.  The  amplifying  or 
compression  state  will  be  displayed  in  present  waveform,  with  the  turn  of  knob; 
character indication displays on top of interface at any moment. 
 
Because  waveform  zoom  is  taken  sampling  clock  as  the  time  unit,  so  the  zoom 
ratio  change  is  spaced  by  step,  if  measure  waveform’s  interval  with  scale  of 
waveform  frame,  may  have  error,  just  a  rough  measurement.  If  need  to  measure 
the  interval  of  the  waveform  accurately,  should  use  the  method  of  cursor 
measurement. 
 
The  zoom  function  of  the  waveform  is  applied  in  the  situation  that  the  changing 
rate  difference  of  the  logic-level  in  every  channel  is  very  big  in  a  group  of 
waveforms,  if  the  changing  rate  of  the  logic  level  in  a  group  of  waveforms  is 
generally bigger or smaller, then should solve it by adjusting the sampling rate. 

-10- 

Summary of Contents for 1235

Page 1: ...PeakTech 1235 Operation manual 32 Chanel Logic Analyser ...

Page 2: ... find and solve without the logic analyzer such as transmission delay competition risk and burr interference At present digital circuit and bus technique are used in many apparatuses In order to analyze and validate the result of information processing the logic analyzer must be used to find out the error in programming and running measure and compare the state of digital logic circuit With the ra...

Page 3: ...comparator A TTL level digital signal is generated by the comparison of the two signals and then the digital signal is stored synchronously in the data flip latch by sampling clock When sample the internal code the code generator will produce 30 channels of internal digital signals which are stored synchronously in the data flip latch by sampling clock The sampling data in the data flip latch are ...

Page 4: ... diagram 3 Read write data Sampling circuit Sampling rate read write data Keyboard Input Storage address generator High speed comparator Threshold voltage High speed memory MCU Trigger process External signal Clock circuit Internal code generator ...

Page 5: ...keys 4 Data input keys 5 Cursor key 6 Sample key 7 Function keys 8 Signal input 9 Signal input 10 Adjusting knob AC 100 240V 45 65Hz 1A REPLACE FUSE AS SPECIFIED DISCONNECT POWER CORD BEFORE REPLACING FUSE RS 232 USB 3 2 1 Rear panel 1 Power source outlet 2 RS232 interface 3 USB Device interface 4 1 2 3 4 5 6 7 8 9 10 ...

Page 6: ... Number input 11 keys 0 1 2 3 4 5 6 7 8 9 number input key x special characters x only used for data search input 4 2 4 Input control 4 keys select the setting parameters up to down in cycle backspace to delete the input data when the input hasn t finished Shift used for input the English letters above the button 4 2 5 Display control 5 keys Display display timing waveforms or data lists in cycle ...

Page 7: ...eters on the left is the state parameters changes automatically with different operations showing the current state of the instrument The two parameters on the right is the parameters settings of the system can be set with numeric keys 4 3 3 Trigger settings interface The trigger settings interface displays the whole trigger process Adopting the graphical mode makes users can easy to understand th...

Page 8: ...ease the key then input the letter you need If there is no letter on the key user can input space The shift key is a one kick So if the user needs to input the next letter please repeat the operations above 5 3 Channels setting 5 3 1 Channel order setting The double digits number 00 31 on the most left of the waveforms display frame is the channel serial number Press Channel key to select the chan...

Page 9: ...ltage the analyzer shows the number 1 otherwise it shows the number 0 Then the signals are sampled and saved and displayed on the screen Waveforms displayed by this means only reflect the timing logic when the input signals are higher or lower than the threshold voltage It doesn t reflect the real amplitude of the input signals and may be far different from the real waveforms of the input signals ...

Page 10: ...el is at the bottom This is to enable the highest order digit of the byte to be on the top and the lowest order digit on the bottom 5 5 2 Waveform rolling Up Down The waveform rolling can be used to observe all the waveforms in 32 channels Press and turn the knob to browse the waveforms If there is a need to compare the waveforms the method of channel order setting 5 3 1 can be used to display the...

Page 11: ...direction Zoom ns div in the sixth row on the left bottom of the wave frame will change this ratio coefficient stands for time amount that each lattice of scale line represents in the above and underside two rows of waveform frame The essence of waveform amplification is that a number in sampling memory is displayed continuously with several points in waveform display So the amplifying waveform ca...

Page 12: ...inuously from the top to the bottom The middle is the address value of the sampling data in Hex the right one is the Bin address value a parting line between each 8 bits Bin code so as to read conveniently Use two different colors to distinguish the two adjacent lines making display clear beautiful and not easy to confuse The storage depth of the instrument is 260 000 storing addresses but the scr...

Page 13: ...ncide with each other their parameters are 0 If want to measure the time difference between two points in present waveform interface can move cursor1 and cursor2 to the two aim points respectively it s easy to read the time difference or the address difference between the two points But if the address difference between the two points to be measured exceeds 280 the two points can t be displayed in...

Page 14: ...t result may be more difficult Because the Inner Code Generator s logic relation is simple waveform is pure and standard one can quickly master logic analyzer s use with it Second in practical application if the test result is different from expectant one sometimes it is hard to make clear that the problem comes from the tested circuit or logic analyzer here as long as use inner code generator to ...

Page 15: ...e discrete event that is take the event sequences as independent variable the data listing after sampling reflects the logic state relation between the system clock and the other signals in the system this mode is known as state analysis Here the sampling clock is synchronous with the tested system it is also called synchronous sampling If take sample using inner clock to inner code generator this...

Page 16: ... sampling cycle in order to observe the particular changes of the tested signals that is to increase the sampling rate Generally speaking the sampling cycle should be less 3 5 times than the narrowest pulse width of the tested signals In other words even the narrowest pulse of the tested signals should include three sampling points at least which can truly reflect input signals change as time The ...

Page 17: ...y which requires corresponding sampling velocity for logic analyzer However the space of memory in the instrument is limited so in practical the effective sampling time is very short it can say sampling process completes instantaneously If start up sampling process manually this needs to press Single in an extreme accurate time but it is too hard to do moreover the data needed sampling and storing...

Page 18: ...n the input data streams in other words when the logic level of 00 15 channels is 0001001000110100 the sampling process starts 5 8 3 Start select The start select can be set with numbers the key 0 cut the switch the key 1 connects the switch If the start select switch is connected the start conditions will be short circuit having no use In other words after pressing Single the sampling process sta...

Page 19: ... that it captures a trigger event In some applications the trigger event we care may appear many times and the thing we interested in is the situations after the trigger event appears many times If we can capture the trigger event for many times in one sampling process that may be more convenient for analysis So the instrument sets a trigger event counter after the sampling process starts the coun...

Page 20: ...nals appear and displays sample is processing press any key to stop To release from this state just press any key to stop the sampling process manually Then one must study the tested signals carefully reset the trigger process to make sure the sampling process can run normally 5 8 9 Trigger cursor Sometimes there may be one or several vertical red lines in the timing waveforms interface they are t...

Page 21: ...repeatedly is very troublesome in each use so it sets parameter save function and save the present parameters settings of the instrument for the further use press save key The instrument will query firstly Store 0 prameter 1 waveform 2 cancel Press 0 key to save the all currently setting parameters even not lose when power off 5 9 2 Waveform storage In practical application catching an interesting...

Page 22: ...s Input Channel 32 data sample channels two external clock channels Threshold voltage 6 independently adjustable threshold voltages Adjusting range 6V to 6V Resolution 0 1V Input impedance Resistance 100kΩ Capacitance 8pf Input range 500mVpp to 20Vpp Input protection maximum input voltage 40V no damage 6 2 Sample Storage characteristics Timing Sample internal clock Sampling Rate 1Hz to 100MHz 10ns...

Page 23: ...gether with it Cursor measure move the position of the cursor can dynamically display the data values the address distance and the interval of the two cursors Trigger cursor the sample point accords with the trigger conditions Search cursor the sample point accords with the search conditions 6 6 Internal code generator Code Type 00 to 15 channels are counters with adding one 16 to 29 channels are ...

Page 24: ...al knowing Technical changings which are in the interest of progress reserved We herewith confirm that the units are calibrated by the factory according to the specifications as per the technical specifications We recommend to calibrate the unit again after one year PeakTech 05 2012 Ho PeakTech Prüf und Messtechnik GmbH Kornkamp 32 DE 22926 Ahrensburg Germany 49 0 4102 42343 44 49 0 4102 434 16 in...

Reviews: