![Peak IPEH-004005-J User Manual Download Page 18](http://html1.mh-extra.com/html/peak/ipeh-004005-j/ipeh-004005-j_user-manual_1517165018.webp)
MU-Thermocouple1 CAN FD – User Manual
18
The larger measurement error applies. In this case, the resulting
measurement error is
1.5 K
.
Total Accuracy
The total accuracy results from the measurement error components
of the measurement electronics (1 K), reference temperature
sensors (1 K), and the thermocouples (1.5 K):
Total accuracy
= 1 K + 1 K + 1.5 K
=
3.5 K