14
3.2.2
Status LEDs
With the default configuration the status LEDs for the thermocouple
inputs are preconfigured according to the following table:
LED
LED Indication
Meaning
Lights red
An intact thermocouple is connected.
If the corresponding LED does not light up,
despite the thermocouple being connected,
the cable connection or the thermocouple
may be broken.
Thermocouple
inputs
Flashes red quickly
(4 Hz)
No thermocouple is plugged in, the
thermocouple is not connected correctly,
or the cable at the thermocouple is broken.