Part No.  IG-Genius 3023  

Rev. 

01 

 

Page 

Page 6 of 9 

 

18. 

          

 

19. 

     

 

The desired system which is to be used 

must run for a minimum of 30 mins before 

connecting to application. 

Select desired system. Using highlighted 

switches. 

20. 

       

 

21.  

         

  

Switch the generator back 

ON

Leave the generator switched on for a period 

of 30 minutes to purge the system. 

22. 

23. 

 

The 

‘NITROGEN’

 pressure gauge, on the 

selected system(s), on the front of the 

generator will gradually increase from 

0

 to 

100psi

 as the pressure in the internal 

storage tanks increases. 

After 30 minutes of running, the generator can 

be turned off. Allow the stored gas in the 

internal tanks to dissipate. 

Summary of Contents for Genius 3023

Page 1: ...cientific com Part No IG Genius 3023 Rev 01 Page Page 1 of 9 x 1 x 2 Installation Guide Genius 3023 Generator FITTINGS KIT 8mm Hex Key Mains Cable Euro Silencer Fitting Compression Fitting Mains Cable...

Page 2: ...e the RH side panel Using the Hex Key provided in the fittings kit turn the locks 90o to the right An earth lead is connected to the inside of the panel This can be disconnected by pulling it off the...

Page 3: ...e unit is to be transported from its current location 8 9 From the fittings kit select one of the 6mm push fit fittings and connect it to the port marked DRAIN on the rear of the unit using a 16mm or...

Page 4: ...ngle phase power supply NOTE Do not touch anything inside the generator while the side panels are removed and the mains power is connected to the unit 12 13 Switch the generator power ON at the switch...

Page 5: ...Switch the generator power OFF at the switch on the front panel Remove the mains cable completely from the power supply 16 17 Re attach both the RH LH side panels to the generator taking care to ensur...

Page 6: ...e connecting to application Select desired system Using highlighted switches 20 21 Switch the generator back ON Leave the generator switched on for a period of 30 minutes to purge the system 22 23 The...

Page 7: ...fitting or the remaining 6mm push fit fittings and attach to the NITROGEN ports If the compression fittings are used it is recommended that PTFE tape is applied to the thread to ensure a leak tight f...

Page 8: ...nsformer at your earliest convenience Mains voltages between 220V and 253V no transformer is required The 6m length of Teflon tube supplied in the fittings kit is to allow the generator to be installe...

Page 9: ...41 812 8200 PRODUCT WARRANTY REGISTRATION COMPANY CONTACT NAME ADDRESS EMAIL ADDRESS CITY TOWN GENERATOR SERIAL NUMBER POSTCODE COUNTRY MODEL TYPE TELEPHONE INSTALLATION DATE DD MM YYYY Important Plea...

Reviews: