PCS-P100 User Manual
5
www.pdk.co.kr
1.5 General Specifications
Pressure range
-0.097 to 10 MPa
Accuracy
± 0.02% F.S
Electrical measurement
and source
±15 V.dc, ±24 mA.dc,
Accuracy ±(0.01% of R 1digit), Loop Power
Display
Full color touch screen LCD
, 110 mm(4.3") diagonal,
480 x 272 pixels., 6 digit
Power
Lithium-ion rechargeable battery, Charger 9V, 6A
(5 hrs charge, 30 hrs battery life)
Workable
Temperature Range
-20 to 70 ℃
Storage
Temperature Range
-30 to 80 ℃
Temperature
Compensated Range
-10 to 50 ℃
Pressure units
Pa, hPa, kPa, MPa, bar, mbar, kg/cm
2
, psi, mmH
2
0,
cmH
2
0, inH
2
0, mmHg, inHg, mSW, fSW
(mSW and fSW units are applied on a water temperature of 15°C
with reference to U.S. Navy Dividing Manual, Revision 7, Table 2-
10. Pressure Equivalent.)
Test Port
1/8" PF Female
Weight
4.2 kg including batteries
Size
201 mm(W) x 360 mm(D) x 110 mm(H)
1.6 Ordering Information
Model (PCS-P100_Range)
Description (
Pressure Range
)
200 kPa
-97 kPa to 200 kPa
350 kPa
-97 kPa to 350 kPa
700 kPa
-97 kPa to 700 kPa
2 MPa
-97 kPa to 2 MPa
3.5 MPa
-97 kPa to 3.5 MPa
7 MPa
-97 kPa to 7 MPa
10 MPa
-97 kPa to 10 MPa