5) ⑥ Using Fine adjust handle to control
the target pressure. (CW to increase,
CCW to decrease pressure)
6) After finished the test, Turn ③
Vent / Release valve to CCW to
open the Vent/Release valve.
WARNING
1.
PCS-P100 is pneumatic test and calibration purpose only.
2.
Do not switch between pressure and vacuum mode under pressure.
Open the vent valve before switching between positive to vacuum pressures.
3.
Low pressure gauges can very easily be over pressured if not careful. Please take caution
when applying pressures.
4.
Store the PCS-P100 in a dry non-corrosive environment.
5.
PDK is not responsible for problems or damage resulting from improper use or operation of
the equipment.