User Manual

 

MODEL : PCS-P100 

Portable Pneumatic Pressure Calibrator

 

PDK CO., LTD.

 

www.pdk.co.kr

 

 

Summary of Contents for PCS-P100

Page 1: ...User Manual MODEL PCS P100 Portable Pneumatic Pressure Calibrator PDK CO LTD www pdk co kr...

Page 2: ...al items 6 2 Main Menu 7 2 1 Power on off 7 2 2 Unit Resolution Zero Cal 8 2 3 Slide 10 2 4 mA 2wire 24V out mA 3wire 24V out 11 2 5 Vdc 3wire 24V out Screen switching 12 3 TASK Menu 13 3 1 TASK Menu...

Page 3: ...ine pressure control PCS P100 also has Pressure vacuum switching valve to generate up to 97 kPa in vacuum mode which has the best degree of a vacuum in its class Full graphic touchscreen display with...

Page 4: ...alibrate pressure gauge for pressure measuring instruments retail business 4 Pressure generating and controlling for pressure test lab 5 Pressure calibration for laboratory 6 Pneumatic high pressure c...

Page 5: ...l color touch screen display Hand strap Fine pressure adjustment Electrical connections for measuring mA V and detecting switch contacts Pressure range Vent release valve Test port On Off Full color t...

Page 6: ...re Range 30 to 80 Temperature Compensated Range 10 to 50 Pressure units Pa hPa kPa MPa bar mbar kg cm2 psi mmH20 cmH20 inH20 mmHg inHg mSW fSW mSW and fSW units are applied on a water temperature of 1...

Page 7: ...8 Optional Items Item Description Dirt Moisture Trap Up to 10 MPa Pneumatic Quick connection adaptor set CA P Series 1 4 3 8 1 2 1 8 PT PF NPT Flexible tubes FTH Series 30 cm to 300 cm Traceable calib...

Page 8: ...7 www pdk co kr 2 Main Menu 2 1 Power on off 2 1 1 Power on Press and hold down Power button until screen shows It will changes screen to screen 2 1 2 Power off Press and hold down Power button until...

Page 9: ...r unit mmH2O will not indicates because after converted the unit 7 digits cannot indicate on display 2 2 Unit Resolution Zero Cal 2 2 1 Unit 2 2 2 Resolution User can select to digit number 4 5 or 6 O...

Page 10: ...PCS P100 User Manual 9 www pdk co kr 2 2 3 Zero cal If pressure does not indicate zero when fully vented condition touch to zeroing Zeroing function only works when 0 5 F S range...

Page 11: ...PCS P100 User Manual 10 www pdk co kr 2 3 Slide Touch and slide the screen to move page as below figure for Electrical Signal measurement P a h P a k P a M P a b a r m b...

Page 12: ...2wire 24V out Loop Power Measure up to 24mA Supply up to 24 VDC 2 4 2 mA 3wire 24V out Measure up to 15V Supply up to 24 VDC Touch Help to see the reference connection Touch Exit to change the screen...

Page 13: ...5 Vdc 3wire 24V out Screen switching 2 5 1 Vdc 3wire 24V out Measure up to 15V Supply up to 24 VDC 2 5 2 Screen switching Touch Exit to change the screen back Touch Help to see the reference connectio...

Page 14: ...CS P100 User Manual 13 www pdk co kr Touch to go to task set up menu Max Min test Leak test Switch test and Setting can be selectable 3 TASK Menu 3 1 TASK menu Max Min Avg Switch Test Leak Test Settin...

Page 15: ...uch then select Max Min 2 Touch to start measure 3 Touch to reset the values This function only can use in pressure measure mode The leak test function is provided as a means of checking and quantifyi...

Page 16: ...tween the change of states i e OPEN CLOSE or CLOSE OPEN value shows in Hysteresis 1 Touch then select Switch Test 2 Touch to reset the values 1 Time Set time 2 Cal Pressure Pressure Calibration 3 Cal...

Page 17: ...PCS P100 User Manual 16 www pdk co kr 3 5 1 Time User can set time in Time page 3 5 2 Cal Pressure User can calibrate the pressure values in Cal Pressure page...

Page 18: ...PCS P100 User Manual 17 www pdk co kr 3 5 3 About PCS User can see the current PCS information in About PCS page...

Page 19: ...ption On Off Test port Vent release Lever type hand pump Charging port Fine pressure adjustment Pressure Vacuum switching valve Hand Strap Pin Test port Vent release Charging port Fine pressure adjust...

Page 20: ...00 User Manual 19 www pdk co kr 4 2 Pressurize Test port Charging port Fine pressure adjustment Pressure Vacuum switching valve Lever type hand pump On Off Vent release 1 Connect the UUT to test port...

Page 21: ...ce without volume will apply approximately 1 6 to 2 0 bar of pressure Short stroke will generate small pressure Vacuum mode also 2 Turn Pressure Vacuum switching valve to CCW to Switching to P 3 Turn...

Page 22: ...st and calibration purpose only 2 Do not switch between pressure and vacuum mode under pressure Open the vent valve before switching between positive to vacuum pressures 3 Low pressure gauges can very...

Page 23: ...P100 User Manual 22 www pdk co kr 4 3 Vacuum Test port Charging port Fine pressure adjustment Pressure Vacuum switching valve Lever type hand pump On Off Vent release 1 Connect the UUT to test port 1...

Page 24: ...et pressure with Hand pump Pressing the pump up once without volume will apply high vacuum Short stroke will generate small vacuum 2 Turn Pressure Vacuum switching valve to CW to Switching to V 3 Turn...

Page 25: ...tic test and calibration purpose only 2 Do not switch between pressure and vacuum mode under pressure Open the vent valve before switching between positive to vacuum pressures 3 Low pressure gauges ca...

Page 26: ...ifying the reason for PCS P100 behavior and determining whether the behavior is due to normal operation or an internal or external problem If the failure cannot be fixed please stop system operation i...

Page 27: ...k contact the manufacturer or authorized agent Poor pressure control Pressure supply too low Verify pressure supply External leak Verify Test port and UUT are leak tight Internal leak contact the manu...

Page 28: ...LTD Head Office Postal Code 34122 10 6 Expo ro 339beon gil Yuseong gu Daejeon Korea Tel 042 862 6880 Fax 042 862 6881 E mail pdk pdk co kr PDK CO LTD Seoul Office Postal Code 08506 1 610 IT Castle 98...