GRL-PCIE-TX Quick Start/User Guide/MOI
Rev1.0
© PCI-SIG 2022
Version 1.0, Mar 2022. Updated 03.26.2022
Page.
6
2
Resource Requirements
Note: Equipment requirements may vary according to the lab setup and DUT board. Below are the
recommended lists of equipment for the typical test setup.
2.1
Equipment Requirements
T
ABLE
1.
E
QUIPMENT
R
EQUIREMENTS
–
S
YSTEMS
Equipment
Qty.
Description/Key Spec Requirement
High Performance Real-time Oscilloscope
[a]
1
≥
33 GHz bandwidth (For PCIe Gen5)
Sample rate: ≥
128 Gb/s with 2x interpolation
“GRL
-
P1” PCIe Compliance Test Hardware
Controller (Optional)
1
For PCIe compliance toggle control
Advanced Technology eXtended (ATX) Power
Supply
1
For power supply to the DUT
GRL Switch (Optional)
1
For multi-lane automated testing
PCI-SIG Compliance Base Board (CBB)
or
PCI-SIG Compliance Load Board (CLB)
1
For add-in cards
For hosts
Computer (laptop or desktop)
1
Windows 7+ OS
For automation control of the DUT state
[a]
Oscilloscope with scope software requirements
as specified in vendor specific MOI’s.
For example, when
using the Keysight Scope, scope software such as Keysight InfiniiSim / EZ-JIT / Serial Data Analysis / Serial
Data Equalization that are required for testing and signal processing must be pre-installed on the Scope.
Similarly, the Tektronix Scope shall be used with DPOJET (Jitter and Eye Analysis Tools)
software for making
measurements.
T
ABLE
2.
E
QUIPMENT
R
EQUIREMENTS
–
C
ABLES
Cable
Qty.
Power Control Adapter Cable
1
USB Type-A to Type-A/B Micro Cable
1
Test Fixture Control Cable (2-conductor power control only cable or
6-conductor fixture control cable)
1
Clock Compliance Toggle Cables (matched SMA-to-SMP cables)
1 pair
SMA-to-SMP Cables
4 pairs
[a]
[a]
Based on the standard test configuration. May require more or less cables depending on the DUT
type.