MOST 20-2B

 

  

 

PACIFIC CONSOLIDATED INDUSTRIES

 

350003 Rev D

SECTION II 

 
2-1 FUNCTIONAL 

DESCRIPTION 

 

2-2 

MOST OPERATING PANEL 

 

 
The MOST Operating Panel contains the connecting port 
necessary to refill the MOST with high-pressure oxygen 
and the fittings necessary to provide distribution of the 
oxygen. A valve (V-3) in the filling circuit allows the 
release of pressure in the vent position. Without a vent 
position, the full cylinder pressure would be constantly on 
the fitting, making it impossible to remove or insert the 
quick connect hose fitting. The operating Panel includes 
two pressure gauges. 

The Cylinder Contents Gauge (P1-1), indicates the pressure 
of the remaining oxygen and therefore the remaining 
quantity of oxygen. The Output Pressure Gauge indicates 
the pressure of the oxygen being supplied to the Oxygen 
Output Fittings and Hoses. Three quick connect Oxygen 
Output ports and one quick connect Oxygen Input port are 
located in the center of the panel. A vent port for the 
internal pressure relief devices is also located on this panel 
(See Fig 4-2). 

 

Prior to attaching an Oxygen Hose to the MOST, the Flow 
Selector must be placed on the opposite end of the Oxygen 
Hose. The correct end has a small hand wheel to assist the 
connection of the hose to the threaded fitting on the end of 
the Flow selector, which is marked "INLET". Ensure that 
the Flow Selector is set to "0" before connecting to the 
hose.  
 
There are three Oxygen Output quick connecting ports on 
the MOST Operating Panel. Each port is equipped with a 
protective plug attached to the panel by a lanyard. The 
protective plug should always be in place when the 
respective output fitting is not being used. The Oxygen 
Output fittings are used to connect the tapered end of the 
Oxygen Conducting Delivery Hose(s) to the MOST. 
Pressure is applied on the hose as it is inserted into the 
fitting. When fully engaged, the hose fitting is captured by 
the output fitting. The oxygen is then free to flow through 
the hose. The hose is released by grasping the fitting 
mounted on the panel and twisting the fitting in a 
clockwise direction. A one quarter turn will release the 
hose fitting. 
 

 

2-4 OXYGEN 

INPUT 

 

The MOST high-pressure composite wound cylinders are 
designed for an operating pressure of 3,000 psig. 

The MOST was designed as a functional component of the 
Expeditionary Deployable Oxygen Concentration System 
(EDOCS-120). The EDOCS-120 generates 120 liters per 
minute (lpm) of USP 93% oxygen at 100 psig. It can also 
deliver 60 lpm at 2,250 psig to fill the MOST. Each 
EDOCS-120 is equipped with a stainless steel high-
pressure hose that may be used to connect the EDOCS-120 
to the Oxygen Input fitting on the MOST Operating Panel. 
There is no shut off valve or flow adjustment valve located 
on the panel.

 

In order to release the fittings the filling 

valve must be turned to the ‘vent’ position.  

Although the MOST was designed for use with the 
EDOCS-120, GOX Carts, H-Cylinders or any other 
sources of USP grade oxygen may also be used. The 
EDOCS-120 can deliver USP grade oxygen at 93% while 
the latter systems can deliver USP grade oxygen between 
90 to 100% 

IDENTIFICATION LABEL 

 

 

MOST 20-2B 

Figure 2-1 

2-3 OXYGEN 

OUTPUT 

Summary of Contents for 793080-002

Page 1: ...IONS AND SERVICE MANUAL Including ILLUSTRATED PARTS BREAKDOWN MODEL MOST 20 2B PART NUMBER 793080 002 Protected under US Patent No 2004 0226381 A1 Additional Patents Pending PACIFIC CONSOLIDATED INDUS...

Page 2: ...ustrated Parts Breakdown IPB 13 SECTION VIII 8 1 Trouble Shooting 23 LIST OF FIGURES SECTION IX 9 1 Recommended Spare Parts 24 FIGURE Page Figure 1 1 Cylinder Contents Gauge 1 Figure 1 2 MOST P ID Sch...

Page 3: ...inspected in accordance with Compressed Gas Association pamphlet G 4 1 prior to use to ensure that no combustible materials remain in the connecting piping and fittings If you are not familiar with ox...

Page 4: ...ves regulators gauges and fittings to come into contact with oils greases organic lubricants rubber or any other combustible substance Make sure that any cleaning repair or transfilling of oxygen equi...

Page 5: ...stored and stockpiled ready for immediate use transported with the medical unit and totally self sufficient for as long as required It also has to be easy to operate dependable and as small and light...

Page 6: ...MOST 20 2B 2 PACIFIC CONSOLIDATED INDUSTRIES 350003 Rev D MOST P ID Schematic Figure 1 2...

Page 7: ...G ST STL 7905674 044 CYL 1 STORAGE CYLINDER BRASS LINED 2250 PSI 1 09 CU FT 582515 001 CYL 2 STORAGE CYLINDER BRASS LINED 2250 PSI 1 09 CU FT 582515 001 F 1 FILTER SINTERED STAINLESS STEEL 581530 007...

Page 8: ...urrently To access the Apparatus Kit unlatch the MOST top cover and tilt it to the up position The Apparatus Kit contains Three Oxygen Conducting Delivery Hose s 20 feet long wound on individual holde...

Page 9: ...el Each port is equipped with a protective plug attached to the panel by a lanyard The protective plug should always be in place when the respective output fitting is not being used The Oxygen Output...

Page 10: ...e setting it will result in loss of product 8 Close top lid of case and secure MOST in desired position 9 Connect Flow Selector to Oxygen Delivery Hose Ensure Flow Selector is closed set to 0 10 Remov...

Page 11: ...in VENT position See Figure 4 2 4 Disconnect all hoses from the Oxygen Outlet Connectors 5 SLOWLY OPEN the Cylinder Shutoff Valves See Figure 4 1 6 Remove protective plug on Oxygen Input Port 7 Attac...

Page 12: ...e valve is set to close and when venting is set to vent 3 OXYGEN INPUT C 4 High pressure self sealing quick disconnect fitting for refilling MOST cylinders 4 OUTPUT PRESSURE P1 3 Direct reading gauge...

Page 13: ...interior of MOST 20 2B assembly with a clean dry oil and lint free cloth b Secure the caps on the flow control valve inlets and outlets Stow the valves in the pockets in the accessories tray c Check...

Page 14: ...outlet d Set the flow control valve to the 0 flow rate position e Insert the supply hose male quick disconnect into one of the output quick disconnect fittings on the MOST 20 2B panel f Observe that t...

Page 15: ...50 1000 2010 8960 6 1500 2110 9370 Empty 1915 8530 250 1950 8680 500 1990 8850 750 2025 9010 1000 2060 9180 8 1500 2140 9500 Table 6 1 Capacity of MOST20 2B After Transfilling MOST Operating Panel Fig...

Page 16: ...a rate of about 200 psig per minute NOTE The MOST incorporates a restrictor orifice OR 1 that limits the filling rate to a maximum of 200 psig per minute No special regulators or valves are required...

Page 17: ...EAKDOWN MOBILE OXYGEN STORAGE TANK MODEL MOST 20 2B PART NUMBER 793080 002 Protected under US Patent No 2004 0226381 A1 Additional Patents Pending PACIFIC CONSOLIDATED INDUSTRIES 12201 Magnolia Avenue...

Page 18: ...Figure 7 1 Medical Oxygen Storage Tank MOST Sheet 1 16 Figure 7 1 1 Medical Oxygen Storage Tank MOST Sheet 2 17 Figure 7 2 Door Assembly Panel 18 Figure 7 3 Tubing Assembly 19 Figure 7 4 Panel Assembl...

Page 19: ...OSE ASSEMBLY 3 24 792975 001 75906 FLOWMETER ASSY 3 25 CD13539 25515 DISC HOLDER 2 26 8026250 25515 DISC BURST TAG BD1 2 2 27 50C 0021 0014 19062 HOSE RACK ASSEMBLY 3 28 HHP B 30780 PLUG PIPE SOC HD 1...

Page 20: ...MOST 20 2B 16 PACIFIC CONSOLIDATED INDUSTRIES 350003 Rev D Medical Oxygen Storage Tank MOST Figure 7 1...

Page 21: ...793495 001 REV C PURITY FLUID OXYGEN OUTPUT FLOW 225 LPM FLAMMABLE ATMOSPHERE CAUTION NOT FOR USE IN OIL RICH OR INPUT FLOW 120 LPM OUTPUT PSI 50PSI INPUT PSI 2250 PSI MAX EXT CONTAINER VOLUME 6 1 CU...

Page 22: ...v D Figure Index No Part Number FSCM Description Units Per Assy SMR Code Usable on code 7 2 793257 001 75906 DOOR ASSY PANEL REF 1 793257 003 75906 DOOR PANEL 1 2 M1 61 1 94222 LATCH FLUSH PULL 2 Tabl...

Page 23: ...70 TEE BRANCH MALE 1 6 SS 400 3 4TTF 02570 TEE BRANCH FEMALE 2 7 SS 400 3 02570 TEE UNION 5 8 SS 400 4 02570 CROSS UNION 1 9 SS 400 2 4 02570 ELBOW MALE 2 10 SS 400 8 4 02570 ELBOW FEMALE 2 11 SS 400...

Page 24: ...2A B1 400K1 02570 COUPLING QC 1 13 6153 OUK87 CHECK UNIT SCHRADER 3 14 9833K22 3A054 BREATHER FILTER VENT 1 15 793464 001 75906 PLUG ASSEMBLY PROTECTOR 1 16 793465 001 75906 PLUG ASSEMBLY PROTECTOR 3...

Page 25: ...MOST 20 2B 21 PACIFIC CONSOLIDATED INDUSTRIES 350003 Rev D Panel Assembly Figure 7 4...

Page 26: ...03 75906 BRACKET ANGLE 1 3 793810 001 55996 MOUNTING BRACKET CLAMP 4 4 MS35307 358 96906 BOLT HEX HD 375 16 x 75 LG 8 5 MS15795 814 96906 WASHER FLAT 39 dia 8 6 MS17830 6C 96906 NUT LOCK REG 375 16 8...

Page 27: ...6 SECTION VIII 8 1 TROUBLE SHOOTING Problem Probable Cause Recommendation 1 Oxygen at input output connector leaks or does not flow in out of the system 1 The connectors are not seated properly 1 Dis...

Page 28: ...HOSE ASSEMBLY 1 6 582461 001 01JF9 EDGE CASTER KIT 1 7 793674 036 75906 VALVE CHECK 10PSI 1 8 793674 044 75906 VALVE CHECK 1PSI 1 9 793674 045 75906 VALVE CHECK 5PSI 1 10 792993 001 75906 REGULATOR PR...

Reviews: