10 

 

15  Temperature Calibration 

During  the  measurement  process,  if  the  temperature  reading  displayed  differs  from  that  of  an 
accurate thermometer, the tester needs to be calibrated. 
 

 

 

Press and hold the 

Cal

 key for 3 seconds to 

enter the setup menu. 

 

 

Press the 

Enter

  key  until  the tester  shows 

°C/CAL or °F/CAL. 

 

 

Press the 

Cal

 key to enter the temperature 

calibration mode. 

 

Press  the 

Cal

  key  again  to  modify  the 

temperature value (Resolution: 0.5°C). 

 

 

 

Press  the 

Enter

  key  to  confirm,  the  tester 

shows the next menu item. 

 

Press  the 

Meas

  key  to  return  to  the 

measurement 

mode. 

Calibration 

is 

completed. 

 
 

 

Summary of Contents for PCE-WO2 10

Page 1: ...0 Oxygen Meter User Manual User manuals in various languages fran ais italiano espa ol portugu s nederlands t rk polski can be found by using our product search on www pce instruments com Last change...

Page 2: ...o Use 6 11 Setup Menu 6 11 1 Setting the default option 7 12 Setting the salinity coefficient 8 13 Setting the ambient pressure 8 14 DO Calibration in Saturation Mode 9 14 1 Single point calibration 1...

Page 3: ...s The case should only be opened by qualified PCE Instruments personnel Never use the instrument when your hands are wet You must not make any technical changes to the device The appliance should only...

Page 4: ...ure compensation 0 40 C 32 104 F automatic Air pressure correction 60 0 112 5 kPa 450 850 mmHg Salinity correction 0 35 g l Sensor type polarography Environmental conditions 0 60 C 32 140 F max 80 r H...

Page 5: ...lay PCE WO2 10 pocket dissolved oxygen tester is equipped with an easy to read LCD display that used to show the measured values and mode icons The following table describes the function of each icon...

Page 6: ...hes the tester ON OFF Locks the measured value press the key again to resume measuring Exits the calibration or setting and returns to the measurement mode CAL Press the key to start the calibration P...

Page 7: ...cap onto the probe excess electrolyte solution will drain out Be sure the cathode of probe makes contact with membrane cap the electrolyte solution in membrane cap should be without an air bubble 8 C...

Page 8: ...uirements MENU DESCRIPTION OPTIONS DEFAULT Set the salinity coefficient of sample Range 0 35ppt 0 0ppt Set the barometric pressure coefficient Range 450 850mmHg 760mmHg Set the number of calibration p...

Page 9: ...the steps above until the tester returns to the measurement mode Setting is completed During the setting process press the Meas key the tester will exit the setup menu and return to the measurement m...

Page 10: ...in the unit mmHg The following table will help to convert your ambient pressure into the required unit mmHg on the basis of your current height ALTITUDE m kPa mmHg ALTITUDE m kPa mmHg 0 101 3 760 160...

Page 11: ...e into the air saturated water for 15 minutes Press the Enter key Wait for the reading to stabilize the tester automatically shows END Single point calibration is completed 14 2 2 points calibration M...

Page 12: ...ss and hold the Cal key for 3 seconds to enter the setup menu Press the Enter key until the tester shows C CAL or F CAL Press the Cal key to enter the temperature calibration mode Press the Cal key ag...

Page 13: ...the membrane on the probe will be blocked damaged or eroded Connect the dissolved oxygen probe to tester and wait for 15 minutes to polarize the probe If necessary to set the barometric pressure and...

Page 14: ...ECTIVE ACTION DO probe does not connect to tester Check the connector of probe Measured value is out of range Check the DO membrane whether clogged dirty or broken Electrolyte solution is depleted Ref...

Page 15: ...ed pollutants batteries must not be disposed of as household waste They must be given to collection points designed for that purpose In order to comply with the EU directive 2012 19 EU we take our dev...

Page 16: ...0583 975 114 Fax 49 0 2903 976 99 9971 Fax 44 0 2380 98703 9 Fax 39 0583 974 824 info pce instruments com info industrial needs com info pce italia it www pce instruments com deutsch www pce instrumen...

Reviews: