11.1
Setting the default option
•
Press and hold the
Cal
key for 3 seconds to enter the setup menu, the tester shows a
menu item and default option.
•
If necessary, press the
Cal
key again to select the desired option.
•
Press the
Enter
key to confirm, the tester moves to the next menu item.
•
Repeat the steps above until the tester returns to the measurement mode. Setting is
completed.
•
During the setting process, press the
Meas
key, the tester will exit the setup menu and
return to the measurement mode.
•
If you do not want to calibrate the temperature, press the
Enter
key to skip the °C/CAL
or °F/CAL option.
•
The Reset function will restore the tester back to factory default settings. If enabled, all
of the calibration data and selected parameters will be lost or reset, the tester must
be recalibrated.