Assembly
Using the provided screw (3), fix the water jacket (1) onto the
rear support (2).
Close the sample container (4) with the lower cap
(5). Ensure
that the lower cap is properly fixed.
Fill the sample container (4). Ensure that there are not any
bubbles in the container. Tilt the container and fill it with a large
syringe. The needed sample volume is relatively small (8-13
ml).
Check the sample volume. The spindle should be covered
completely.
Hang the desired spindle (6) onto the spindle hook (7). Attach
the hook to the screw (8).
Insert the spindle with hook and screw into the sample
container.
Insert the sample container (4) into the water jacket (1) from
below.
Fix the sample container (4) in the water jacket (1). The groove
of the lower cap has to be aligned with the fastening screw (9)
of the water jacket. Turn the sample container to do so.
Attach the upper cap (10). Make sure that the spindle does not
fall into the sample container.
Using the provided screw (11), fix the rear support (2) onto the
viscometer.