4.13 Interface protocol:
DATA FORMAT
HEAD1, HEAD2,
DATA
UNIT CR
1 2 3 4th 5 6
78910111213 14
15 16 17 18 19 20
HEAD1 (2BYTES)
HEAD2 (2BYTES)
OL - overload
NT - net weight mode
ST - stable
US - unstable
DATA
(8BYTE)
2D (HEX) = “
-
” (negative sign) 20 (HEX) = “
”(Blank)
2E (HEX) = “. ”(Decimal point)
UNIT (4
bytes)
g =
20 (HEX) ; 20 (HEX); 20 (HEX)
; 67 (HEX)
kg = 20 (HEX); 20 (HEX); 6B (HEX) ; 67 (HEX)
ct =
20 (HEX); 20 (HEX); 63 (HEX) ; 74 (HEX)
ozt = 20 (HEX); 6F (HEX); 7A (HEX); 74 (HEX)
CR = OA (HEX); OD (HEX);
Transmission example
1. Ex.: Stable net + 0.168 g
HEAD, HEAD,
DATA
UNIT
CR
ST,
NT, + 0.168
G
OA, OD