4.11 Calibration (external)
1. Press the
Key and keep it pressed while switching on the scale,
until the scale has completed the self-test. "CAL" now appears in the display.
Now press
again
and "0" appears in the display.
2. Now you can use the number keys to enter the weight with which you want to calibrate. A weight of 2/3
of the path area is recommended. (PCE-PCS 6 -> 4 kg; PCE-PCS 30 -> 20 kg) The unit of entry is
kilograms. Now place the appropriate weight on the scale.
3. Press the
Key to perform the calibration. The calibration is now complete.
(The calibration is invalid if the calibration range deviates by more than 10% from the maximum
weighing range.
4.12 Recharge the battery
If the battery charge is low, a corresponding message appears on the display. Now switch off the scale
and then charge the battery. A yellow LED lights up during the charging process. When the scale is fully
charged, it lights up red. The charging process takes about 12 hours to fully charge the battery.