PCE Instruments PCE-PCS Series Manual Download Page 5

4.0 operation 

 

4.01 Switch on / off 

 

Remove all objects from the platform and switch the scale on or off using the switch on the lower 
right side. 

 
 

4.02 zeroing 

 

Zero range: ± 2% of the max. Weighing range. 

 

If the display of the scale does not jump to "0" without loading the weighing platform, please press the 
"ZERO" button. 

 
 

4.03 Tare function 

 

Place the weight to be tared on the weighing platform and press the tare key to tare this weight. The 
tare  function  is  now  activated.  Now  take  the  weight  down,  the  display  now  shows  the  weight  as  a 
negative value. Press the tare key again to exit this function. 

 
 

4.04 piece counting 

 

If the weight of a piece is known, you can enter it directly using the number keys 

 

 
 

and confirm the entry by pressing the key. If the weight is not known, this can also be entered using the 
objects. Enter the number of objects using the number key, the number entered is shown in the Unit 
Weight display. Now press 

 

 
 

the

 

Button, now the weight and the number are shown in the display. 

 
 

4.05 alarm 

 

The user has the option of specifying a number of pieces. If this number is exceeded, an alarm signal 
sounds. To do this, enter the number of pieces using the number keys and press 

 

The button 

. To delete press the button

 

 

 
 

4.06 memory 

 

Press the key 

after the display has stabilized to store the current value in the internal memory. 

The weight display now shows the number of stores (max. 99) and the total number of all stores in the 
Unit Weight Display. If "OL" appears on the display, the internal memory is full and must be deleted. To 
clear the memory, press the button. 

 

 
 
 

Important: The balance must be unloaded after each saving process so that a new value can be 
saved. 

 

- 5 - 

Summary of Contents for PCE-PCS Series

Page 1: ... Germany Tel 01805 976 990 Fax 029 03 976 99 29 info warensortiment de www warensortiment de 14 cents per minute from the German Landline max 42 cents per minute from the German mobile network operation manual Piece counting scale PCE PCS series Version 1 0 11 04 2011 ...

Page 2: ...ERATION 5 4 01 TURN ON 5 4 02 NULLATION 5 4 03 TARE FUNCTION 5 4 04 DEALING 5 4 05 NOISE 5 4 06 MEMORY 5 4 08 ADJUST THE TIME 6 4 09 ADJUSTING THE BACKLIGHT 6 4 10 DATA ISSUE RS232 TO ADJUST 6 4 11 CALIBRATION EXTERNAL 7 4 12 BATTERY RECHARGE 7 4 13 INTERFACE PROTOCOL 8 2 ...

Page 3: ... Static electricity can falsify the result Discharge the static charge Use for example anti static spray to counteract the effect 2 0 specifications Scale type PCE PCS 6 PCE PCS 30 Weighing range 6000g 30000g Readability 0 1 g 0 5 g reproducibility 0 2 g 0 5 g Minimum load 0 1 g 0 5 g Minimum unit weight 0 1 g 0 5 g Memory locations 99 for total piece counting Tare range 100 Weighing unit Grams di...

Page 4: ... container By pressing this button you can save the current value in the memory To clear the memory press this button To clear the alarm for certain quantities press this button With this button you can set a certain number of pieces when exceeded an alarm signal should sound Press this button to delete the entered values from the display You can manually enter the piece weight by pressing this bu...

Page 5: ...e key If the weight is not known this can also be entered using the objects Enter the number of objects using the number key the number entered is shown in the Unit Weight display Now press the Button now the weight and the number are shown in the display 4 05 alarm The user has the option of specifying a number of pieces If this number is exceeded an alarm signal sounds To do this enter the numbe...

Page 6: ... on the balance until the balance has completed the self test The current setting now appears on the display AUTO the backlight is switched on and off automatically A Backlight is always on OUT Backlight is always off Now press the button to set the value Press the button to confirm the set value 4 10 Adapt data output RS232 Press the 2 key and keep it pressed while switching on the scale until th...

Page 7: ...unit of entry is kilograms Now place the appropriate weight on the scale 3 Press the Key to perform the calibration The calibration is now complete The calibration is invalid if the calibration range deviates by more than 10 from the maximum weighing range 4 12 Recharge the battery If the battery charge is low a corresponding message appears on the display Now switch off the scale and then charge ...

Page 8: ...ATA UNIT CR ST NT 0 168 G OA OD If you have any questions please contact PCE Deutschland GmbH You can find an overview of our measurement technology here http www warensortiment de messtechnik htm An overview of our measuring devices can be found here http www warensortiment de messtechnik messgeraete htm You can find an overview of our scales here http www warensortiment de messtechnik messgeraet...

Reviews: