15 

5.4  Print out results (optional) 

You can print out the measurement results by using the optional PCE-BP 1 printer. To do so, follow these 

steps: 

 

1.  Make sure that the printer is set up properly and both devices are connected to each other. 

2.  During operation, press and hold the 

key until “PrInT” appears in the display. 

3.  After this, the options “L-C” (Chinese print-out) and “L-E” (English print-out) will appear in the 

display. Use the arrow keys 

 and 

 to select the desired option and press the 

 key to 

confirm. 

4.  After this, numbers from 1 to 5 will appear in the display which represent the last 5 

measurements (which are stored in the device). Select a measurement by using the arrow keys 

 and 

 and press the 

 key to confirm. 

5.  The data is now sent to the printer via the RS-232 interface. 

 

Print-out:  

 

 

6  Troubleshooting 

 

Error message 

Error cause 

Possible solution 

-Err 1- 

Sample weight below 1 g 

Put more than 1 g of the sample on the scales 

and start the measurement. 

-Err 2- 

Temperature setting below +40 

°C 

Restart the scales. 

-Err 3- 

Measuring time setting below 30 

seconds. 

Restart the scales. 

-Err 4- 

Halogen lamp problem 

Please contact our technical support. 

-Err 5- 

Time setting for slow heat-up 

mode below 3 minutes 

Restart the scales. 

-Err 6- 

Temperature sensor broken 

Please contact our technical support. 

 

 

 

Summary of Contents for PCE-MA 100

Page 1: ...8 Jupiter FL 33458 USA From outside US 1 Tel 561 320 9162 Fax 561 320 9176 info pce americas com www pce instruments com english www pce instruments com PCE Instruments UK Ltd Units 12 13 Southpoint B...

Page 2: ...ion 9 4 3 Disposable aluminium sample pan 9 4 4 Spreading of the sample 9 4 5 Glass fibre filter circle 10 4 6 Practical advice 10 5 Operation 10 5 1 Measurement 10 5 2 Drying modes 11 5 2 1 Specify h...

Page 3: ...E Instruments or equivalent Before each use inspect the case for visible damage If any damage is visible do not use the device Do not use the instrument in explosive atmospheres The limit values for t...

Page 4: ...e Measurement range moisture 0 100 Readability moisture 0 01 0 1 0 01 0 1 Memory 16 drying programmes parameters Interface RS 232 Display LCD 17 mm digit height Power supply 220 V 50 Hz or 110 V 60 Hz...

Page 5: ...tection plate 1 2 The notch in the side wall of the wind shield 2 must be placed in a way that it faces the user 3 The handle of the insertion aid 3 is placed to rest in this notch 4 Place the sample...

Page 6: ...eat up mode Standard Quick Slow Measuring stop mode constant reading user time lapse Drying temperature 40 C bis zu 199 C Drying time Duration of drying process Percentage moisture Percentage dry weig...

Page 7: ...on off Start stop measurement Tare Leave settings Open settings Switch between absolute and relative measuring result Confirm current settings go to next setting Return to weighing mode short keystro...

Page 8: ...optional PCE BP 1 printer to the scales To do so you have to set the following on the printer The printer comes with an RS 232 cable which is used for connecting both devices SUB D 9 plug for RS 232 i...

Page 9: ...the measurement You should avoid tools with heat conducting properties i e which are able to transfer heat to the sample Improper handling and preparation of the sample will falsify the end result of...

Page 10: ...ure balance saves the drying parameters last used after switching off and back on These are shown in the display Prepare the sample and place it in the drying chamber You can now start the measurement...

Page 11: ...an select one out of three heat up speeds Standard heat up mode This is the default mode which is suitable for most sample types In this heat up mode 120 C are reached after approx 4 minutes Quick hea...

Page 12: ...s 2 mg 60 s 2 mg ATTENTION In this mode it is not possible to set a measuring time Manual user stop mode In this measurement stop mode the measurement can be ended by the user To do so press the key...

Page 13: ...mperature regulation temperature fluctuations of approx 4 C can occur Mind this when selecting the temperature to make sure the sample does not scorch Otherwise the moisture reading will be falsified...

Page 14: ...n be adjusted by means of a 100 g external test weight min M1 Special care must be taken when positioning the scales as little vibration as possible and balanced by means of the level The adjustment o...

Page 15: ...appear in the display which represent the last 5 measurements which are stored in the device Select a measurement by using the arrow keys and and press the key to confirm 5 The data is now sent to th...

Page 16: ...with the EU directive 2012 19 EU we take our devices back We either re use them or give them to a recycling company which disposes of the devices in line with law If you have any questions please cont...

Page 17: ...la pr sente d claration est conforme aux normes cit es ci apr s TR Uygunluk Beyan Bu bildirime ba l r n n a a daki standartlara uygun oldu unu beyan ederiz NL Conformiteit verklaring Wij verklaren hi...

Reviews: