![PCE Instruments PCE-HDM 7 User Manual Download Page 6](http://html1.mh-extra.com/html/pce-instruments/pce-hdm-7/pce-hdm-7_user-manual_1516138006.webp)
© PCE Instruments
4
2
Specifications
Insulation
Class 2, double insulation
Case
Double molded, IP 67 waterproof and dust-
proof
Diode test
Test current max. 1.5 mA, open circuit
voltage 3 V typical
Continuity test
Audible signal if the resistance is approx. 50
or less
Low battery indication
“
” is displayed
Display
4000 count LCD
Overrange indication
“OL” is displayed
Polarity
Minus symbol “-“ is displayed for negative
polarity
Measurement rate
2 readings per second, nominal
Auto Power Off
After approx. 15 minutes of inactivity
Input impedance
10 MΩ AC/DC voltage
AC response
True RMS
AC bandwidth
50 … 400 Hz
Batteries
3 x 1.5 V AAA batteries
Fuse
10 A / 600 V (5 x 20 mm) fast blow
Operating conditions
0 … 40 °C / 32 … 104 °F
<70 % RH
Storage conditions
-10 … 60 °C / 14 … 140 °F
<80 % RH
Operating altitude
Max. 2000 m
Dimensions
147 x 68 x 50 mm
Weight
318 g
Safety
Complies with UL 61010-1 V.3 for
measurement category lll 600 V, pollution
degree 2
Accuracy is given at 18 … 28 °C (65 … 83 °F), ≤70 RH.
Function
Range
Resolution
Accuracy
±(% of r digits)
AC voltage
4.000 V
1 mV
±(1.0 % + 5)
40.00 V
10 mV
±(1.2 % + 5)
400.0 V
0.1 V
±(1.5 % + 5)
600 V
1 V
Input protection: 600 V DC or 600 V AC RMS
Input impedance: 10 M
AC Response: 50 … 60 Hz
Function
Range
Resolution
Accuracy
±(% of r digits)
Low Z
AC voltage
4.000 V
1 mV
±(1.5 % +5)
40.00 V
10 mV
40.00 V
0.1 V
All AC voltage ranges are specified from 5 % of range to 100 % of range
Input protection: 300 V AC RMS or 300 V DC
Input impedance: approx. 3 kΩ