PCE Instruments PCE-HDM 7 User Manual Download Page 15

 

 

©  PCE Instruments 

13 

 

Operation 

6.1 

Auto Power Off 

The meter automatically turns off after 15 minutes of activity. To reset the meter after it shuts off, 

turn rotary function switch to the off position and then set the switch to the desired function. To 

disable Auto Power Off, turn the rotary function switch to the off position. Press and hold the 

MODE button and set the rotary function switch to the desired function. Release the MODE button 

when the “

” symbol disappears from the display. Auto Power Off is now disabled. Auto Power 

Off will be restored when the meter is turned off and back on. 

6.2 

MODE button 

Used to select AC or DC voltage, Hz or % duty cycle, ohms, diode Test, continuity or capacitance, 

and ° C and ° F. 

6.3 

MAX/MIN button 

1. 

Momentarily press the MAX/MIN button to activate MAX/MIN mode. The “MAX” indicator 

will appear on the LCD. The meter will display and hold the maximum reading and will 

update when a higher “max” occurs. 

2. 

Momentarily  press  the  MAX/MIN  button  again  to  view  the  lowest  reading.  The  “MIN” 

indicator will appear on the LCD display. The meter will display and hold the minimum 

reading and will update when a lower “min” occurs. 

3. 

Press  and  hold  the  MAX/MIN  button  to  end  MAX/MIN  mode  and  return  to  normal 

operation. 

 

NOTE

: MAX/MIN does not work with frequency, duty cycle, diode test, continuity and capacitance. 

6.4 

Bluetooth/Flashlight button 

Momentarily  press  the  Bluetooth/Flashlight  button  to  turn  the  flashlight  on  and  off.  Bluetooth 

allows readings to be displayed and stored on mobile devices. To activate Bluetooth, press and 

hold  the  Bluetooth/Flashlight  button  until  the  “

”  symbol  appears  on  the  display.  Bluetooth 

should be disabled when not connected to a mobile device in order to conserve battery power. 

To turn off Bluetooth, press and hold the Bluetooth/Flashlight until the “

” symbol no longer 

appears on the display. Use the “Meterbox Pro” app from your Android or iOS store to connect 

the meter to your phone.. 

6.5 

HOLD/Backlight button 

To freeze the reading on the display, momentarily press the HOLD button. “HOLD” will appear 

on the LCD while the reading is being held. Momentarily press the HOLD button again to return 

to normal operation. The backlight illuminates the LCD when the ambient light is too low to view 

the displayed readings. To switch on the backlight, press and hold the HOLD button until the 

backlight turns on. To turn it off, press and hold the HOLD button until the backlight turns off. 

 

 

Summary of Contents for PCE-HDM 7

Page 1: ...ster PCE HDM 7 User Manual User manuals in various languages fran ais italiano espa ol portugu s nederlands t rk polski can be found by using our product search on www pce instruments com Last change...

Page 2: ...AC DC voltage measurements 8 5 3 AC current measurements 9 5 4 DC current measurements 9 5 5 Frequency and duty cycle measurements 10 5 6 Resistance measurements 10 5 7 Diode test 11 5 8 Continuity 1...

Page 3: ...spect the case for visible damage If any damage is visible on the meter or its test leads do not use the device Do not use the meter if it operates incorrectly Protection may be compromised Do not ope...

Page 4: ...ge with respect to earth ground exceeds the maximum safety rating of the meter This symbol can appear next to another symbol or connection and indicates the user must refer to the manual for important...

Page 5: ...ear and distribution panels Lighting circuits in commercial buildings Feeder lines in industrial plants Any device or branch circuit that is close to a Cat III source The measurement category CAT rati...

Page 6: ...V AAA batteries Fuse 10 A 600 V 5 x 20 mm fast blow Operating conditions 0 40 C 32 104 F 70 RH Storage conditions 10 60 C 14 140 F 80 RH Operating altitude Max 2000 m Dimensions 147 x 68 x 50 mm Weigh...

Page 7: ...solution Accuracy of reading digits AC current 4 000 A 1 mA 2 5 3 10 00 A 10 mA Overload protection 10 A 600 V fuse AC response 50 60 Hz Function Range Resolution Accuracy of reading digits DC current...

Page 8: ...gits Duty cycle 1 0 99 9 0 1 1 2 2 Input protection 600 V AC RMS or 600 V DC Pulse width 0 1 100 mS Frequency range 5Hz 10kHz Sensitivity 8 V RMS Function Range Resolution Accuracy of reading digits T...

Page 9: ...wer Off Low battery AUTO Autoranging HOLD Display hold LOZ Low Z impedance MAX MIN Maximum Minimum Bluetooth 3 1 Delivery contents 1 x multimeter PCE HDM 7 1 x test lead 1 x storage bag 1 x K type the...

Page 10: ...ges Do not connect to circuits that exceed 300 V AC RMS or 300 V DC when the meter is set to Low Z Low Z is used to check for ghost voltage Ghost voltages are present when non powered wires are in clo...

Page 11: ...wer to the circuit 6 Read the current on the LCD 5 4 DC current measurements WARNINGS Observe all safety precautions when working on live circuits Do not measure current on circuits that exceed 600 V...

Page 12: ...lead probes to the circuit under test 5 Read the frequency or duty cycle on the LCD 5 6 Resistance measurements WARNING Never test resistance on a live circuit 1 Set the rotary function switch to the...

Page 13: ...the display Reverse voltage will indicate OL Shorted devices will indicate near 0 and an open device will indicate OL in both polarities 5 8 Continuity WARNING Never test continuity on a live circuit...

Page 14: ...minute to get a stable reading on large capacitors 5 10 Temperature measurements WARNING Do not touch the temperature probe to live circuits 1 Set the rotary function switch to the F C position 2 Pre...

Page 15: ...ill update when a lower min occurs 3 Press and hold the MAX MIN button to end MAX MIN mode and return to normal operation NOTE MAX MIN does not work with frequency duty cycle diode test continuity and...

Page 16: ...o avoid electric shock do not operate meter until the battery fuse cover is securely fastened to the meter Fuse replacement WARNING To avoid electric shock remove the test leads from the meter before...

Page 17: ...d pollutants batteries must not be disposed of as household waste They must be given to collection points designed for that purpose In order to comply with the EU directive 2012 19 EU we take our devi...

Page 18: ...dustrial needs com info pce italia it www pce instruments com deutsch www pce instruments com english www pce instruments com italiano The Netherlands Chile Hong Kong PCE Brookhuis B V PCE Instruments...

Reviews: