5.7
Capacitance measurement
•
Slide the cover of the input sockets upwards.
•
Set the rotary switch to position
Ω.
•
Press the MODE key repeatedly until "F" appears on the display.
•
Plug the test leads into the measuring input sockets (black to COM and red
to
Ω).
•
Touch the measuring spots with the measuring tips.
•
Read the measured value.
The stabilisation of the measured value can take several seconds for larger measured
values.
5.8
Frequency and duty cycle
•
Slide the cover of the input sockets upwards.
•
Set the rotary switch to position Hz.
•
Plug the test leads into the measuring input sockets black to COM and
red to CAP).
•
Touch the measuring spots with the measuring tips.
Read the measured values in the display (frequency in the middle / duty cycle at
the bottom).
5.9
µA current measurement AC / DC
•
Slide the cover of the input sockets upwards.
•
Set the rotary switch to position µA.
•
Use the MODE key to select AC or DC.
•
Plug the test leads into the measuring input sockets (black to COM and
red to µA).
•
Touch the measuring spots with the measuring tips.
•
Read the measured value in the display.
5.10
Temperature measurement
•
Slide the cover of the input sockets downwards.
•
Set the rotary switch to position Temp.
•
Use the MODE key to select °C or °F.
•
Plug the thermocouple(s) into the measuring input sockets.
•
Place it/them at the desired measuring spots.
•
Read the measured value in the display.
A measured value overflow or unconnected temperature sensor is indicated in the
display by "OL". During the T1-T2 measurement, a missing sensor is signalled by ----- in the main
display.
With the RANGE key, you can choose between different types of displaying the measured values:
Key
Display
RANGE
Middle
Bottom
T1
T2
1 x
T2
T1
2 x
T1-T2
T1
3 x
T1- T2
T2