5.1.2 Locating and tracing of lines and sockets
ATTENTION:
-
The circuit must not be live.
-
The neutral cable and the protective grounding cable must be connected and fully functional.
-
Connect the transmitter to the phase cable and the protective grounding cable as shown in the
picture below:
In case of a single-pole application the lateral circuit branches can be tracked (the fuse must be removed
in this example).
If the power cable that is supplied with the signals from the transmitter is located, for example, directly
parallel to the other conductors (e.g. cable slot or channel) or if these conductors are crossed, the signals
are also sent to the other conductors.
During the search and tracing, the closer the locator gets to the cables to be tracked, the stronger the
signal is displayed.
Set the transmission power in order to adapt it to different search radii.
The target position can be located accurately by using the manual mode and adjusting the sensitivity.