©  PCE Instruments 

 

6.2 

Data logger function 

Press  the  logger  key  (3-2)  for  more  than  2  seconds  until  the  display  shows 
"DATALOGGER". Now the data logger starts to save measured values. 
 
If you  want to switch  off  the  data logger  function, you  must  press  the  logger  key  (3-2) 
again for more than 2 seconds. The "DATALOGGER" indicator then disappears from the 
display. 
 
in  chapter  8.3,  it  is  described  how  to  set  the  recording  interval;  in  chapter  8.4,  it  is 
described how to set the beeper. 
 

Note:

 Before removing the SD card, make sure that the data logger function is stopped. 

Otherwise, you may lose data from the SD card. 

6.3 

Time information 

If you press and hold down the time key (3-3) for more than 2 seconds, the following data 
appears in the display: year/month/day, hour/minute/second and the recording interval. 

6.4 

SD card data structure 

1. 

When you first insert the card into the meter, it generates a folder on the memory 
card: 

HBA01

  

2. 

When you start the data logger function for the first time, the meter generates a file 
under the 

HBA01\

 folder with the name HBA01001.xls. The data is then  saved to 

this file. As soon as there are 30,000 data records in this file, a new file is created. 
This file then has the name HBA01002.xls. 

3. 

When 99 files are stored in the 

HBA01

 folder, the machine creates a new folder with 

the name: 

HBA02\...

  

4. 

This results in the following structure: 

HBA01\

 

 HBA01001.xls 
 HBA01002.xls 
 


... 

 HBA01099.xls 
 

HBA02\

 

 HBA02001.xls 
 HBA02002.xls 
 


... 

 HBA02099.xls 
 

HBAXX\

 

... 

 

Note:

 The highest value XX can stand for is 10. 

 

 

Summary of Contents for PCE-AQD 10

Page 1: ...O2 Data Logger User Manual User manuals in various languages fran ais italiano espa ol portugu s nederlands t rk polski can be found by using our product search on www pce instruments com Last change...

Page 2: ...ard to PC 6 8 Settings 6 8 1 SD card format 7 8 2 Setting the time 7 8 3 Setting the recording interval 7 8 4 Setting the beeper 8 8 5 Setting the decimal point 8 8 6 Setting the temperature unit 8 8...

Page 3: ...s The case should only be opened by qualified PCE Instruments personnel Never use the instrument when your hands are wet You must not make any technical changes to the device The appliance should only...

Page 4: ...mperature Relative humidity CO2 0 50 C 10 90 RH 0 4000 ppm Resolution Temperature Relative humidity CO2 0 1 C 0 1 RH 1 ppm Accuracy Temperature Relative humidity CO2 0 8 C 4 of rdg 70 ppm 1000 ppm 5 o...

Page 5: ...pension brackets 3 8 Table stand 3 9 Battery compartment cover 3 10 Safety screw for battery compartment cover 3 11 RESET key 3 12 RS 232 output 3 13 SD card slot 3 14 9 V DC connection 3 15 CO2 senso...

Page 6: ...formatted Refer to chapter 8 1 for more information Note Please do not use an SD card formatted in another device e g digital camera In this case you have to format the SD card again in the data logge...

Page 7: ...wn the time key 3 3 for more than 2 seconds the following data appears in the display year month day hour minute second and the recording interval 6 4 SD card data structure 1 When you first insert th...

Page 8: ...e tester press the SET key 3 5 for more than 2 seconds This takes you to the setting menu and you can navigate through the menu with each further press of the SET key Sd F Format SD card dAtE Setting...

Page 9: ...Now you can move to the next value The sequence then is month day hour minute second Note The value to be set flashes 2 When you have set all the values and confirmed with the Enter key 3 2 all the s...

Page 10: ...ing the RS 232 interface 1 If the display shows rS232 you can use the key 3 3 and the key 3 4 to select yES or no where yES means that the RS 232 interface 3 12 is activated and no means that the inte...

Page 11: ...state This is done as follows While the machine is switched on press the RESET key 3 11 gently with a pointed object The machine is now reset to its original state 12 RS 232 PC interface The device h...

Page 12: ...ed pollutants batteries must not be disposed of as household waste They must be given to collection points designed for that purpose In order to comply with the EU directive 2012 19 EU we take our dev...

Page 13: ...tanbul United Kingdom SO31 4RF Italia T rkiye Tel 44 0 2380 98703 0 Telefono 39 0583 975 114 Tel 0212 471 11 47 Fax 44 0 2380 98703 9 Fax 39 0583 974 824 Faks 0212 705 53 93 info pce instruments co uk...

Reviews: