©  PCE  Instruments

 

 

36 

 

 

 

5.4 

Data logger

 

The  internal  memory  of the device can save up  to 1,800  values. (Each item  selected in  M50  is

 

considered a value). To set or start the logger, do the follow ing: 

 

(1)  Via w indow  M51, you can set the start time, the saving interval and the run time. This means

 

that measurements can be taken for a maximum of 24 hours. 

 

(2)  Window  M52 is used to determine w here data are saved. According to the standard setting,

 

the data are saved to the device. You can also transfer the data directly to the RS-232 / USB

 

interface w ithout saving them to the internal memory. 

 

(3)  Via w indow  M50, you can set the logger and select the values you w ould like to save.

 

You can view  the saved data via w indow M53. 

You can read out the saved data from the buffer via the RS-232 / USB interface and delete the internal

 

memory using the functions in w indow  M52. 

 

To transfer data to a computer, you w ill need the softw are SOFT-PCE-TDS. 

Calibration

 

There is a  (calibration) factor betw een the real flow  velocity and the flow  velocity displayed by the

 

device. This calibration factor can be determined by carrying out a calibration. How ever, you will need

 

flow  calibration equipment to do so. 

 

Please send the device to PCE Instruments for calibration. Our contact details can be found at the end

 

of this manual. 

Maintenance

 

 

7.1 

Troubleshooting

 

Error m essages after switching on

 

The  device runs a  self-diagnostic programme  w hen sw itched on.  This  diagnostic programme  is

 

supposed to find hardw are errors. The follow ing chart shows possible error messages. 

 

Error m essage

 

Reason

 

Countermeasure

 

„ROM Testing Error” 
„Segment Test Error” 

Softw are problem 

(1) Restart the device 
(2) Contact PCE  Instruments 

„Stored Data Error“ 

The   parameters  entered  by

 

the user are not integrated 

Press   the  ENTER    key.   All

 

values are reset to default. 

„Timer Slow  Error” 

„Timer Fast Error” 

Problems w ith the timekeeper 

(1) Restart the device 
(2) Contact PCE  Instruments 

„Date Time Error” 

Number error in the calendar 

Initialise    the    calendar   via

 

w indow  M61 

Repeated reboot 

Hardw are problem 

Contact PCE  Instruments 

E

n

g

li

s

h

 

Summary of Contents for PCE-TDS 100 H

Page 1: ...Debimetre Ultrasonic Flow Meter PCE TDS 100 H HS User manualsin various languages fran ais italiano espa ol portugu s nederlands t rk polski can be downloaded here www pce instruments com Letzte nderu...

Page 2: ...aksesuarlar 3 3 Sistem A klamas 4 3 1 Cihaz 4 3 2 Fonksiyon tu lar 6 4 Haz rl k 6 4 1 Dahili pil 6 4 2 A ma 7 4 3 Men pencereleri 7 5 Kullan m 8 5 1 l m prensibi 8 5 2 Parametre konfig rasyonu 9 5 3 S...

Page 3: ...ries 23 3 System description 24 3 1 Device 24 3 2 Function keys 26 4 Getting started 26 4 1 Internal battery 26 4 2 Power on 26 4 3 Menu windows 27 5 Operation 28 5 1 Principle of measurement 28 5 2 C...

Page 4: ...dir G l darbelerden ka n n Metreyia nd r c ya da patlay c gazlar n etraf nda kullanmay n Servis antas sadece kalifiye PCE Teknik Cihazlar personeli taraf ndan a lmal d r Onar m ve bak m al malar sadec...

Page 5: ...i 5 ve ak 0 03 m saat olan t m s v lar Ak Birimleri Metre k p m Litre l Galon ABD gal mparatorlukGalonu NG igl Milyon ABD Galonu mgl Fit k p cf Varil ABD bal ngiliz varili NG ib Ya varili ob Zaman g n...

Page 6: ...A rl k 75 g 260 g 250 g 1080 g 2 2 Teslimat i eri i 1 x Ultrasonik Debimetre PCE TDS 100 2 x Sens r modele g re 2 x 5 m D n t r c Ba lant Kablolar 2 x C rtl Kablo Ba lar 1 x G Adapt r 1 x Ultrasonik...

Page 7: ...ments 4 3 Sistem A klamas 3 1 Cihaz stten g r n m nden g r n m Alttan g r n m 1 Sens r ba lant s alt 2 Sens r ba lant s st 3 Ekran 4 LED arjlambas 5 Dokunma duyarl Membran klavye 6 arj yuvas 7 USB ara...

Page 8: ...PCE Instruments 5 Kablo 5 m 2 x 1 Turuncu ba lant kablosu 2 Mavi ba lant kablosu G adapt r T rk e...

Page 9: ...07 penceresinde g sterilir Voltaj3 V un alt na d t nde pil neredeyse bo al r Ayr ca cihaz kalan pil mr n de g sterir Kalan pil mr g stergesisadecekabaca bir de erdir Tu sim Fonksiyon Yukar tu u Farkl...

Page 10: ...oru malzemesini se mek i in M14 3 Se im yapmadan veriyig r nt lemek cihaz n al ma s resini g r nt lemeki in M 1 1 hakk nda Verigiri ii in penceredeysenizverigiri iniyapabilir ve ENTER tu u ile onaylay...

Page 11: ...olarakhizmet veren iki sinyald n t r c sens r ile al r Sens rler borunun d duvar na belirli bir mesafede biri di erinin alt nda olacak ekilde monte edilir Sens rler Z eklinde Z y ntemi monte edilebili...

Page 12: ...in M14 Standart boru malzemesini se in e er boru malzemesi buraya eklenmi se boru ses h z gerekli de ildir 0 Karbonlu elik 1 Paslanmaz elik 2 D kme demir 3 S nek d kme demir 4 Bak r 5 PVC 6 Al minyum...

Page 13: ...k yap mal d r M26 Parametreleri dahili haf zaya kaydedin 18 haf za yeri M27 Kaydedilen parametreleri okutun M28 Sinyal yetersiz oldu unda cihaz n son iyi de eritutup tutmas n gerekti ini se in standar...

Page 14: ...ay c ba lat n ya da durdurun M39 Dili se in ngilizce ya da ince M40 0 ve 999 saniye aras nda bir de erde olabilen s n m ayarlay n E er 0 se ilmi se s n mdevre d b rak l r M41 D k durdurma noktas n aya...

Page 15: ...68 En d kfrekans i in debiyi ayarlay n M69 En y ksekfrekans i in debiyi ayarlay n M70 Ekran arka plan ayd nlatmas n ayarlay n Tu vuru u olmadan arka plan ayd nlatmas n n ka saniye al aca n se in M72 a...

Page 16: ...bulmak olmal d r Bu hassas l m sonu lar i in gereklidir Boru tesisat sistemlerihakk nda baz temel bilgilere sahip olmak gereklidir deal konum sonsuz uzun d z bir boru olacakt r s v da s k m hava hava...

Page 17: ...ribulunmaktad r Ultrasonik dalgalar n boru duvarlar ndan ve s v dan ge mesi i in ge en s re ak h z sonu lar n sa lar Ultrasoniksinyallerin ge i s resi okk sa oldu undan sens rler en y kseksistemhassas...

Page 18: ...bilgiler girilmelidir 1 Borunun d ap M11 2 Borunun malzeme kal nl M12 3 Boru malzemesi M14 4 Borunun kaplamas M16 5 S v tipi M20 6 Ba l sens r tipi M23 7 Sens rlerin kurulum y ntemi M24 8 Bo lu u M25...

Page 19: ...i in nerilir Borunun stten g r n m 1 Ak a kar d n t r c 2 Bo luk 3 Ak y n 4 Ak y n nde d n t r c W y ntemi W y ntemi ap 20 100 mm olan plastikborular n l mleri i in uygundur Borunun stten g r n m 1 A...

Page 20: ...z ne do rudan transferedebilirsiniz 3 M50 penceresiile kaydediciyiayarlayabilir ve kaydetmekistedi iniz de erlerise ebilirsiniz Kaydedilen verileriM53 penceresinde inceleyebilirsiniz Ara bellekte kay...

Page 21: ...at n 2 PCE Teknik Cihazlar ile ileti ime ge in Data Time Error Takvimde say hatas M61 penceresiile takvimi ba lat n Repeated reboot Donan m sorunu PCE Teknik Cihazlar ile ileti ime ge in Hata kodlar v...

Page 22: ...frekans izin verilen aral n d nda M67 M68 ve M69 pencerelerindeki de erlerikontrol edin M69 penceresindeki y ksekde erleri girin F SystemRAM Error Date Time Error CPU or IRQ Error ROM Parity Error 1 R...

Page 23: ...ebimanuel olarak girildi Bu de eri 0 girin b Sens r kurulumu ile ilgili bir hata c M42 penceresiile mevcut bir debi olmas na ra men ekran 0 a ayarlanm olabilir S f r noktas ayar n tekrarlay n ve borud...

Page 24: ...bilgilendirmek zorunday z Kullan lm piller evsel at k olarak at lmamal d r T keticiler pilleri t ketici i in cretsiz olan uygun bir ticari veya belediye toplama noktas na vermekle y k ml d r Kullan lm...

Page 25: ...arrying the device fromcold to w armrooms and vice versa Avoid strong shocks Do not use the meter around corrosive or explosive gases The case should only be opened by qualified PCE Instruments person...

Page 26: ...03m h Flow units Cubic metre m Litre l Gallon USA gal Imperial Gallon UK igl Million USA Gallons mgl Cubic foot cf Barrel USA bal Imperial Barrel UK ib Oil Barrel ob The time can be per day d per hour...

Page 27: ...250 g 1080 g 2 2 Delivery contents 1 x ultrasonic flow meter 2 x sensor depending on model 2 x 5 m connection cable 2 x releasable cable tie 1 x mains adaptor 1 x contact gel 1 x PCE measuring tape 1...

Page 28: ...ments 24 3 System description 3 1 Device Top view Front view Bottom view 1 Sensor connection down 2 Sensor connection up 3 Display 4 LED charging lamp 5 Membrane keypad 6 Charging socket 7 USB interfa...

Page 29: ...PCE Instruments 25 Cable 5 m 2 x 1 Plug orange 2 Plug blue Mains adaptor English...

Page 30: ...the remaining battery life which is determined internally via the voltage The indication of the remaining battery life is just a rough guide value 4 2 Power on The device can be sw itched on via the...

Page 31: ...9 There are tw o waysto select these windows 1 Directly using the MENU key and the tw o digits 2 Via the Up Dow n keys each keystroke changesto the next higher or previouswindow w hereasthe window M00...

Page 32: ...distance one below the other The sensors can be installed in Z shape Z method In this case the ultrasound w ill pass through the pipe once If the sensors are installed in W shape W method the ultraso...

Page 33: ...er of pipe M14 Select standard pipe material if pipe material is included here no pipe sound velocity is needed 0 Carbon Steel 1 Stainless Steel 2 Cast Iron 3 Ductile Iron 4 Copper 5 PVC 6 Aluminium 7...

Page 34: ...ns are possible M32 Select operating unit of totaliser options as in M31 M33 Multiplier totaliser default 1 M34 Enable disable net totaliser M35 Enable disable positive totaliser M36 Enable disable ne...

Page 35: ...value for alarm 2 M76 Set high value for alarm 2 M77 Set up acoustic signal M82 Date totaliser by day month or year M85 Activate or deactivate Auto Power Off function device powersoff after 4 minutes...

Page 36: ...rapped air air bubbles in the liquid The pipes can either run vertically or horizontally To avoid inaccuracies due to turbulence in the liquid a straight flow calming section before and behind the mea...

Page 37: ...asonic w aves take to pass through the pipe w alls and the liquid allows conclusions about the flow velocity As the transit time of the ultrasonic pulses is very short the sensors should be installed...

Page 38: ...d 1 Outer diameter of the pipe M11 2 Material thickness of the pipe M12 3 Material of the pipe M14 4 Liner of the pipe M16 5 Type of liquid M20 6 Type of sensorsconnected M23 7 Mounting method of sens...

Page 39: ...mm Top view of pipe 1 Upstream transducer 2 Spacing 3 Flow direction 4 Dow nstreamtransducer W method The W method is suitable for measurements of plastic pipes with diameters of 20 to 100 mm Top view...

Page 40: ...tion There is a calibration factor between the real flow velocity and the flow velocity displayed by the device This calibration factor can be determined by carrying out a calibration However you will...

Page 41: ...not properly connected 1 Change measuring location 2 Clean measuring location 3 Check the cables 4 Check the contact gel Q Frequ OutputOver The output frequency is outside the permitted range Check t...

Page 42: ...a M42 despite an existing volume flow Repeat the zero point setting and make sure that there is no flow in the pipe 3 The real battery life is shorter than the value stated in M07 a The battery has ex...

Page 43: ...utants batteries must not be disposed of as household w aste They must be given to collection points designed for that purpose In order to comply w ith the EU directive 2012 19 EU w e take our devices...

Page 44: ...kish English Germany PCE Deutschland GmbH France PCE Instruments France EURL Spain PCE Ib rica S L Im Langel 4 23 rue de Strasbourg Calle Mayor 53 D 59872 Meschede 67250 Soultz Sous Forets 02500 Toba...

Page 45: ...r manualsin various languages fran ais italiano espa ol portugu s nederlands t rk polski can be downloaded here www pce instruments com Specifications are subject to change without notice PCE Instrume...

Reviews: