8. Standard-M1 (PCE-TDS 100-2)
9. Plug-in Type C
10. Standard-HS (PCE -TDS 100-3)
11. Standard-HM (PCE-TDS 100-4)
12. Standard-S1 (PCE-TDS 100-1)
13.
π -Pipe
14. Standard-L1
15. Clamp-On TL-1
16. Standard-M
M24
Select sensor installation:
0. V method, 1. Z method, 2. N method, 3. W method
M25
View distance betw een sensors; should be stuck to as exactly as possible
M26
Save parameters to internal memory (18 memory locations)
M27
Read out saved parameters
M28
Select w hether or not device should hold last good value w hen signal is
insufficient; standard setting is YES
M29
Low limit signal strength (default 35)
M30
Select unit system: metric or English
M31
Select flow rate unit:
Cubic Meters
[m³]
Liters
[l]
USA Gallons
[gal]
Imperial Gallons
[igl]
USA M Gallons
[mgl]
Cubic Feet
[cf]
USA Barrels
[bal]
Imperial Barrels
[ib]
Oil Barrels
[ob]
Time can be per day, per hour, per minute or per second, w hich means 36
different unit combinations are possible
M32
Select operating unit of totaliser (options as in M31)
M33
Multiplier totaliser (default 1)
M34
Enable / disable net totaliser
M35
Enable / disable positive totaliser
M36
Enable / disable negative totaliser
M37
1. Reset totaliser
2. Reset device to factory default settings by pressing Reset key, follow ed by
Back key. Be careful w ith this function and note dow n your personal
settings before using it
M38
Start or stop totaliser by means of ENTER key
M39
Select the language (English or French)
M40
Set damping w hich can be a value betw een 0 and 999 seconds; if
“0“ is
selected, damping is disabled
M41
Set low cutoff value
M42
Set zero point; make sure no liquid passes through the pipe
M43
Delete zero point, reset to default zero point
M44
Set flow value manually (offset value); this value should normally be
“0“
M45
Set scale factor; this value is set to the delivered sensors by PCE before
shipping and should only be changed after a calibration by PCE
M46
View netw ork IDN
M47
Lock device; parameters cannot be changed anymore
M48
Not in use
M49
Comm tester
M50
Enable or disable data logging function