ECN #: 43809

Performance

ENGLISH

SI 

Measurement Range

50 klbf

222 kN

[6][7]

Sensitivity(± 10 %)(RO)

4.0 mV/V

4.0 mV/V

[3][1]

Non-Linearity

≤ 0.05 % FS

≤ 0.05 % FS

[6][8]

Hysteresis

≤ 0.06 % FS

≤ 0.06 % FS

[6][8]

Non-Repeatability

≤ 0.01 %RO

≤ 0.01 %RO

[3][8]

Eccentric Load Sensitivity

0.25 %/in

0.25 %/25.4mm

[8]

Resonant Frequency

15 kHz

15 kHz

Side Load Sensitivity

0.25 %/in

0.25 %/25.4mm

[8]

Creep(in 20 minutes)

0.025 %

0.025 %

[8]

Static Error Band

≤ 0.05 % FS

≤ 0.05 % FS

[6][8]

Environmental

Overload Limit

75 klbf

334 kN

Load Limit(Side Force, F

X

 or F

Y

)

25 klbf

111 kN

[4]

Load Limit

(Bending Moment, M

X

 or M

Y

)

25 klbf-in

2825 Nm

[4]

Load Limit

(Axial Torque, M

Z

)

25 klbf-in

2825 Nm

[4]

Temperature Range(Operating)

-65 to +200 °F

-54 to +93 °C

Temperature Range

(Compensated)

+15 to +115 °F

-9 to +46 °C

Temperature Effect on Output(Maximum)

± 0.0008 %Reading/°F

± 0.0018 %Reading/°C

[5]

Temperature Effect on Zero Balance(Maximum)

± 0.0008 %FS/°F

± 0.0018 %FS/°C

[5]

Electrical

Bridge Resistance

350 Ohm

350 Ohm

[1]

Excitation Voltage(Recommended)

10 VDC

10 VDC

[2]

Insulation Resistance

>5K GOhm

>5K GOhm

Zero Balance

± 1 %RO

± 1 % FS

[3]

Output Polarity

+Tension

+Tension

Physical

Size (Diameter x Height)

6.06 in x 3.50 in

153.9 mm x 88.9 mm

[9]

Weight

21 lb

9.53 kg

Mounting Thread

No English Equivalent

M33 x 2.0 4H

Housing Material

Painted Steel

Painted Steel

Sensing Element

Strain Gage

Strain Gage

Mounting Torque

1400 lbf-in

158 Nm

Deflection at Full Scale Capacity

0.002 in

0.05 mm

Electrical Connector

PT02E-10-6P

PT02E-10-6P

All specifications are at room temperature unless otherwise specified.
In the interest of constant product improvement, we reserve the right to change specifications without notice.

M1204-13B

- PC04E-10-6P Connector

M1204-13C

- PT02E-12-8P Connector

M1204-13D

- PC01E-12-8P Connector

NOTES:

[1]Nominal.
[2]Calibrated at 10 VDC, usable 5 to 20 VDC or VAC RMS.
[3]RO - Rated Output.
[4]Singularly applied, i.e. no other extraneous loads.
[5]Over compensated operating temperature range.
[6]FS - Full Scale.
[7]Calibrated in tension and compression.
[8]Performance specifications guaranteed with factory installed optional Tension Base 084A101
[9]See Outline Drawing 49932 for Complete Dimensions

OPTIONAL ACCESSORIES: 

Model 084A90 Connector Shroud
Model 181-012A PT06A-10-6S(SR)
Model 8311-01-10A CABLE
Model M084A101 TENSION LOAD BASE W/ M33X2 THD.

PCB Load & Torque, Inc.
24350 Indoplex Circle
Farmington Hills, MI 48335
UNITED STATES
Phone: 866-684-7107
Fax: 716-684-0987
E-Mail: [email protected]
Web site: 
http://www.pcbloadtorque.com

OPTIONAL VERSIONS

Optional versions have identical specifications and accessories as listed for the standard model 

except where noted below. More than one option may be used.

Entered: AP

Engineer: PE

Sales: JC

Approved: JSD

Spec Number:

Date: 2/16/2015

Date: 2/16/2015

Date: 2/16/2015

Date: 2/16/2015

59224 

Summary of Contents for M1204-13A

Page 1: ...oad cell Installation and Operating Manual For assistance with the operation of this product contact PCB Piezotronics Inc Toll free 800 828 8840 24 hour SensorLine 716 684 0001 Fax 716 684 0987 E mail...

Page 2: ...repair User servicing or repair is not recommended and if attempted may void the factory warranty Calibration Routine calibration of sensors and associated instrumentation is recommended as this helps...

Page 3: ...ry customer service representative Warranty All equipment and repair services provided by PCB Piezotronics Inc are covered by a limited warranty against defective material and workmanship for a period...

Page 4: ...isclosure Table Pb Hg Cd Cr VI PBB PBDE O O O O O O PCB X O O O O O O O O O O O X O O O O O O O O O O O O O O O O O O O O O O O O O X O O O O O O O O O X O O O O O O O O O O O X O O O O O X O O O O O...

Page 5: ...e O O X O O O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indica...

Page 6: ...nted Steel Sensing Element Strain Gage Strain Gage Mounting Torque 1400 lbf in 158 Nm Deflection at Full Scale Capacity 0 002 in 0 05 mm Electrical Connector PT02E 10 6P PT02E 10 6P All specifications...

Page 7: ...4 5 0 ELECTRICAL INSTALLATION 5 5 1 Electrical Drawing Western Regional Std 5 5 2 Cable Grounding Considerations 5 6 0 CALIBRATION 5 6 1 Calibration Certificate Description 5 6 1 1 Measured Output 6...

Page 8: ...evaluated for potential failure modes of the fixtures It is important to use threaded fasteners properly when assembling test fixtures used with the load cells Fixtures with inadequately designed bol...

Page 9: ...modified 4 2 Mounting Load Cell to a Standard Base or Custom Fixture If the load cell is to be mounted to a custom base or fixture it is necessary to consider the following 1 The thickness of the new...

Page 10: ...f Connector Protector 1 Remove the 4 40 x 1 4 button head cap screws from the connector using a 1 16 hex wrench as shown in Figure 7 2 Carefully pull the connector from the load cell Take care to not...

Page 11: ...A simple test with a voltmeter connected between the power cord ground and the structure on which the load cell is mounted can confirm that the structure has been properly grounded If the power cord g...

Page 12: ...into the load cell structure The strain gage bridges on general purpose low profile load cells are adjusted such that the error due to eccentric loading is less than 0 25 per inch of offset for 1200 s...

Page 13: ...s on the load cell If possible remove all loads 2 Power up the host signal conditioner and connect it to the load cell via appropriate cable and allow for a 30 minute warm up 3 Set the load indicator...

Page 14: ...e load cell is mounted on 5 Check the signal conditioning electronics for proper setup 6 Check the insulation resistance of shielded conductors for short circuits 7 Check isolation resistance load cel...

Page 15: ...der Please request a return material authorization RMA before sending a load cell back to the factory for any reason For calibration services if possible a copy of the purchase order covering the requ...

Reviews: