background image

Model Number

482C15

FOUR-CHANNEL, ICP SENSOR SIGNAL CONDITIONER

Revision: J

ECN #: 47052

[6]

Performance

ENGLISH

SI

Channels

4

4

Sensor Input Type(s)

ICP®, Voltage

ICP®, Voltage

[2]

Voltage Gain(± 1 %)(at 500 Hz)

x1, x10, x100

x1, x10, x100

[2]

Output Range(Maximum)

± 10 V

± 10 V

Low Frequency Response(-5 %)

<0.05 Hz

<0.05 Hz

[5][3]

High Frequency Response(-3 dB)(x100)

>50 kHz

>50 kHz

High Frequency Response

(-5 %)(x100)

>17 kHz

>17 kHz

High Frequency Response

(-5 %)(x1)

>100 kHz

>100 kHz

High Frequency Response

(-5 %)(x10)

>40 kHz

>40 kHz

High Frequency Response

(-3 dB)(x1,x10)

>100 kHz

>100 kHz

Phase Response(at 1 kHz)

± 1 °

± 1 °

Cross Talk(maximum)

-72 dB

-72 dB

Fault/Bias Monitor/Meter(LED)

Open/Short/Overload

Open/Short/Overload

Environmental

Temperature Range(Operating)

+32 to +120 °F

0 to +50 °C

Electrical

Power Required(for supplied AC power adaptor)

AC Power

AC Power

Power Required

(direct input to unit)

DC Power

DC Power

AC Power(47 to 63 Hz)

100 to 240 VAC

100 to 240 VAC

AC Power

 0.7 Amps

 0.7 Amps

DC Power

<0.25 Amps

<0.25 Amps

DC Power

+32 to 38 VDC

+32 to 38 VDC

Excitation Voltage(± 1 VDC)(To Sensor)

+26 VDC

+26 VDC

DC Offset

<20 mV

<20 mV

Constant Current Excitation(To Sensor)

0 to 20 mA

0 to 20 mA

[1]

Overload Threshold(± 1.0 Vpk)

± 10/5 Vpk

± 10/5 Vpk

[2]

Discharge Time Constant(0 to 50 %)

>7 sec

>7 sec

[3]

Broadband Electrical Noise(1 to 10,000 Hz)(Gain x1)

5.6 µV rms

5.6 µV rms

[4]

Spectral Noise(1 Hz)

0.67 µV/

Hz

0.67 µV/

Hz

[4]

Spectral Noise

(10 Hz)

0.10 µV/

Hz

0.10 µV/

Hz

[4]

Spectral Noise

(100 Hz)

0.06 µV/

Hz

0.06 µV/

Hz

[4]

Spectral Noise

(1 kHz)

0.06 µV/

Hz

0.06 µV/

Hz

[4]

Spectral Noise

(10 kHz)

0.05 µV/

Hz

0.05 µV/

Hz

[4]

Broadband Electrical Noise(1 to 10,000 Hz)(Gain x10)

21 µV/rms

21 µV/rms

[4]

Spectral Noise(1 Hz)

5.10 µV/

Hz

5.10 µV/

Hz

[4]

Spectral Noise

(10 Hz)

0.60 µV/

Hz

0.60 µV/

Hz

[4]

Spectral Noise

(100 Hz)

0.22 µV/

Hz

0.22 µV/

Hz

[4]

Spectral Noise

(1 kHz)

0.22 µV/

Hz

0.22 µV/

Hz

[4]

Spectral Noise

(10 kHz)

0.19 µV/

Hz

0.19 µV/

Hz

[4]

Broadband Electrical Noise(1 to 10,000 Hz)(Gain x100)

165 µV/rms

165 µV/rms

[4]

Spectral Noise(1 Hz)

57 µV/

Hz

57 µV/

Hz

[4]

Spectral Noise

(10 Hz)

5.2 µV/

Hz

5.2 µV/

Hz

[4]

Spectral Noise

(100 Hz)

1.7 µV/

Hz

1.7 µV/

Hz

[4]

Spectral Noise

(1 kHz)

1.8 µV/

Hz

1.8 µV/

Hz

[4]

Spectral Noise

(10 kHz)

1.4 µV/

Hz

1.4 µV/

Hz

[4]

Physical

Electrical Connector(ICP® Sensor Input)

BNC Jack

BNC Jack

Electrical Connector

(Output)

BNC Jack

BNC Jack

Electrical Connector

(DC Power Input)

5-socket DIN (female)

5-socket DIN (female)

Size (Height x Width x Depth)

3.2 in x 8.0 in x 5.9 in

8.1 cm x 20 cm x 15 cm

Weight

1.25 lb

567 gm

All specifications are at room temperature unless otherwise specified.
In the interest of constant product improvement, we reserve the right to change specifications without notice.

ICP

®

is a registered trademark of PCB Group, Inc. 

OPTIONAL VERSIONS

Optional versions have identical specifications and accessories as listed for the standard model 

except where noted below. More than one option may be used.

NOTES:

[1] User adjustable, factory set at 4 mA (± 0.5 mA). One control adjusts all channels.
[2] Jumper selectable on internal circuit board.
[3] Un-buffered output, read out device input impedance affects discharge time constant and low 

frequency response of unit.

[4] Typical.
[5] The low frequency tolerance is accurate within ±25% of the specified frequency.
[6] See PCB Declaration of Conformance PS024 for details.

SUPPLIED ACCESSORIES: 

Model 017AXX Power Cord (1)
Model 488B04/NC Power Convertor (1)

Entered: LK

Engineer: CPH

Sales: ML

Approved: ECB

Spec Number:

Date: 7/18/2017

Date: 7/18/2017

Date: 7/18/2017

Date: 7/18/2017

38208 

3425 Walden Avenue, Depew, NY 14043

Phone: 716-684-0001
Fax: 716-684-0987
E-Mail: [email protected]

Summary of Contents for 482C15/NP

Page 1: ...Signal Conditioner Installation and Operating Manual For assistance with the operation of this product contact PCB Piezotronics Inc Toll free 800 828 8840 24 hour SensorLine 716 684 0001 Fax 716 684 0...

Page 2: ...alized tests including sensitivity at elevated or cryogenic temperatures phase response extended high or low frequency response extended range leak testing hydrostatic pressure testing and others For...

Page 3: ...tion on particular operating steps The following symbols may be found on the equipment described in this manual This symbol on the unit indicates that high voltage may be present Use standard safety p...

Page 4: ...China RoHS 2 Disclosure Table Pb Hg Cd Cr VI PBB PBDE O O O O O O PCB X O O O O O O O O O O O X O O O O O O O O O O O O O O O O O O O O O O O O O X O O O O O O O O O X O O O O O O O O O O O X O O O O...

Page 5: ...O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indicates that sai...

Page 6: ...82C05 482C15 GENERAL OPERATION MANUAL For powering ICP sensors and in line charge converters this signal conditioner provides a cost effective method for powering small numbers of sensor channels The...

Page 7: ...2 Safety Considerations 1 1 3 Equipment Rating 2 1 4 Model 482C15 System Description 2 1 5 Model 482C05 System Description 4 1 6 Installation 4 1 7 Grounding Techniques 4 1 8 Unit Power 5 1 9 Maintena...

Page 8: ...heading used in this manual explains dangers that might result in personal injury or death Always read the associated information very carefully before performing the indicated procedure The CAUTION...

Page 9: ...ICP source current 0 to 20 mA Jumper selectable Voltage input mode Jumper selectable gain of x1 x10 x100 482C15 Only Lighted power indicator on the front panel Rear panel BNC input output connectors...

Page 10: ...he PC board Each channel can be set to your selected gain and other options The selection of all these features with jumpers allows the 482C15 to have the functionality of many other higher priced uni...

Page 11: ...that of most sensors and is close to an open circuit voltage of 26VDC The photo on the cover is shown with the unit on and no sensors attached so the OPEN LED is on The SHORT light indicates a very lo...

Page 12: ...plied with unit Plug the adapter into any 100 240VAC 47 63Hz power source attach the supplied ferrite clamp on the AC cord to ensure CE conformance and press the ON OFF button 1 7 2 Power Optional Aut...

Page 13: ...ctral Noise 10 Hz 0 60 V Hz 0 60 V Hz 4 Spectral Noise 100 Hz 0 22 V Hz 0 22 V Hz 4 Spectral Noise 1 kHz 0 22 V Hz 0 22 V Hz 4 Spectral Noise 10 kHz 0 19 V Hz 0 19 V Hz 4 Broadband Electrical Noise 1...

Page 14: ...003 005 FILLETS AND RADII 0 07 0 13 OUTLINE DRAWING 25495 1 OF 1 5X MODEL 482C SERIES 4 CHANNEL SIGNAL CONDITIONER DECIMALS XX 03 XXX 010 DECIMALS X 0 8 XX 0 25 SMB 12 16 11 ECB 12 16 11 KEN L 12 16...