MULTICHANNEL SIGNAL CONDITIONER MODEL 482C GENERAL OPERATION MANUAL

 

3

 

indicates a signal level of  +/-10V Peak has been detected and the signal may be in danger of clipping. These fault lights 
are active for all channels, at the same time. 

 

1-4.4.

 

Gain selection / Jumper definition 

 

The 482C15 has internal jumpers that are accessed by opening the top cover.  The top cover is secured by four (4) 
Philips head screws on the bottom of the unit.  Place the unit on a clean, flat surface (Electrical Static Discharge 
protection recommended) and remove the four screws.  Carefully rotate the unit and place it on its feet, then gently 
remove the cover.   
 
The various jumpers are all labeled on the PC board. Each channel can be set to your selected gain and other options. 
The selection of all these features with jumpers allows the 482C15 to have the functionality of many other higher priced 
units at an economical cost. See the table that explains the function and number of each jumper. “X” represents channel 
number.  Once jumpers have been placed as needed, replace cover – taking special care to align the front and back 
panels with the slots in the top cover.  Insert the four (4) screws and tighten to 5 in-lbs. 
 

 
 
 
 
 

          

              

          

        

 

 
 
 
 
  
 
    Gain 1 10 100 

JX02 

 

1 jumper per channel 

     Overload 5V/10V 

J1 

 

Controls all channels   

     Voltage Mode 

JX03 

 

2 jumpers per channel 

     Bypass Filter  

JX00 

 

Remove to install filter module 

     477A Filter   

FILX00   

Install filter here per channel 

 

Gain IC 

Optional 477A 

5V or 10V (J1) 

Voltage mode selection (JX03) 

Pin 1 Indicator (FILX00) 

Gain x10  

ICP

 mode 

Gain x100 

JX00 open 477A ENA 

Summary of Contents for 482C15/NP

Page 1: ...Signal Conditioner Installation and Operating Manual For assistance with the operation of this product contact PCB Piezotronics Inc Toll free 800 828 8840 24 hour SensorLine 716 684 0001 Fax 716 684 0...

Page 2: ...alized tests including sensitivity at elevated or cryogenic temperatures phase response extended high or low frequency response extended range leak testing hydrostatic pressure testing and others For...

Page 3: ...tion on particular operating steps The following symbols may be found on the equipment described in this manual This symbol on the unit indicates that high voltage may be present Use standard safety p...

Page 4: ...China RoHS 2 Disclosure Table Pb Hg Cd Cr VI PBB PBDE O O O O O O PCB X O O O O O O O O O O O X O O O O O O O O O O O O O O O O O O O O O O O O O X O O O O O O O O O X O O O O O O O O O O O X O O O O...

Page 5: ...O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indicates that sai...

Page 6: ...82C05 482C15 GENERAL OPERATION MANUAL For powering ICP sensors and in line charge converters this signal conditioner provides a cost effective method for powering small numbers of sensor channels The...

Page 7: ...2 Safety Considerations 1 1 3 Equipment Rating 2 1 4 Model 482C15 System Description 2 1 5 Model 482C05 System Description 4 1 6 Installation 4 1 7 Grounding Techniques 4 1 8 Unit Power 5 1 9 Maintena...

Page 8: ...heading used in this manual explains dangers that might result in personal injury or death Always read the associated information very carefully before performing the indicated procedure The CAUTION...

Page 9: ...ICP source current 0 to 20 mA Jumper selectable Voltage input mode Jumper selectable gain of x1 x10 x100 482C15 Only Lighted power indicator on the front panel Rear panel BNC input output connectors...

Page 10: ...he PC board Each channel can be set to your selected gain and other options The selection of all these features with jumpers allows the 482C15 to have the functionality of many other higher priced uni...

Page 11: ...that of most sensors and is close to an open circuit voltage of 26VDC The photo on the cover is shown with the unit on and no sensors attached so the OPEN LED is on The SHORT light indicates a very lo...

Page 12: ...plied with unit Plug the adapter into any 100 240VAC 47 63Hz power source attach the supplied ferrite clamp on the AC cord to ensure CE conformance and press the ON OFF button 1 7 2 Power Optional Aut...

Page 13: ...ctral Noise 10 Hz 0 60 V Hz 0 60 V Hz 4 Spectral Noise 100 Hz 0 22 V Hz 0 22 V Hz 4 Spectral Noise 1 kHz 0 22 V Hz 0 22 V Hz 4 Spectral Noise 10 kHz 0 19 V Hz 0 19 V Hz 4 Broadband Electrical Noise 1...

Page 14: ...003 005 FILLETS AND RADII 0 07 0 13 OUTLINE DRAWING 25495 1 OF 1 5X MODEL 482C SERIES 4 CHANNEL SIGNAL CONDITIONER DECIMALS XX 03 XXX 010 DECIMALS X 0 8 XX 0 25 SMB 12 16 11 ECB 12 16 11 KEN L 12 16...

Reviews: