6

Figure 11.

Typical Test Accelerometer Response

7.3 PCB CALIBRATION PROCEDURE

Numerous precautions are taken at PCB to insure
accurate and repeatable results. This section provides a
brief overview of the primary areas of concern.

Since the Back-to-Back Calibration technique relies on
each sensor experiencing an identical acceleration level,
proper mounting of the test sensor to the reference
standard is imperative. Sensors with mounting holes are
attached directly to the reference standard with a stud
tightened to the recommended mounting torque. A
shouldered mounting stud is typically used to prevent
the stud from “bottoming out” in the hole.

Both mounting surfaces are precision-machined and
lapped to provide a smooth, flat interface according to
the manufacturer’s specification.

A thin layer of

silicone grease is placed between the mating surfaces to
fill any imperfections and increase the mounting
stiffness. The cables are stress-relieved by first routing
them to the shaker head, securing them with tape or
cable ties, then routing them to a nearby stationary
location. This reduces cable motion, which is especially
important when testing charge output sensors and helps
to prevent extraneous noise or stresses from being
imparted into the system. A typical set-up is shown in
Figure 12.

Figure 12.

Typical Calibration Set-Up

Adhesively mounted sensors use similar practices.
However, in this case, a small portion of quick-bonding
gel or similar temporary adhesive is used to attach the
test sensor to a reference standard designed with a
smooth, flat mounting surface.

In addition to mounting, the selection of the proper
equipment is critical.

Some of the more important

considerations include: 1) the reference standard must
be

specified

and

previously calibrated

over

the

frequency and/or amplitude range of interest; 2) the
shaker should be selected to provide minimal transverse
(lateral) motion and minimal distortion; and 3) the
quality of the meters, signal generator, and other devices
should be selected so as to operate within the limits of
permissible error.

7.4 COMMON MISTAKES

Most

calibration

errors

are

caused

by

simply

overlooking some of the fundamental principals of
dynamics. This section attempts to address some of the
more common concerns.

For stud-mount sensors, always mount the accelerometer
directly to the reference standard.

Ensure that the

mounting surfaces are smooth, flat, and free of any
burrs.

Always use a coupling fluid, such as silicone

grease, in the mounting interface to maintain a high
mounting stiffness. Mount the sensor according to the
manufacturer’s recommended mounting torque.

DO

NOT use any intermediate mounting adaptors, as the
mounted resonant frequency may be reduced and
thereby compromise the high-frequency performance. If
necessary, use adaptor studs.

Summary of Contents for 357A19/NC

Page 1: ...ccelerometer Installation and Operating Manual For assistance with the operation of this product contact PCB Piezotronics Inc Toll free 800 828 8840 24 hour SensorLine 716 684 0001 Fax 716 684 0987 E mail info pcb com Web www pcb com ...

Page 2: ...oid the factory warranty Calibration Routine calibration of sensors and associated instrumentation is recommended as this helps build confidence in measurement accuracy and acquired data Equipment calibration cycles are typically established by the users own quality regimen When in doubt about a calibration cycle a good rule of thumb is to recalibrate on an annual basis It is also good practice to...

Page 3: ... warranty against defective material and workmanship for a period of one year from date of original purchase Contact PCB for a complete statement of our warranty Expendable items such as batteries and mounting hardware are not covered by warranty Mechanical damage to equipment due to improper use is not covered by warranty Electronic circuitry failure caused by the introduction of unregulated or i...

Page 4: ... for most models These accelerometers are ideal for structural testing machine monitoring and vehicular shock and vibration measurement tasks where high temperatures preclude the use of accelerometers with built in microelectronics Enclosed is a Specification Sheet which lists the complete performance characteristics of the particular accelerometer 3 0 OPTIONAL FEATURES In addition to these standa...

Page 5: ...surface consider adhesive mounting as a possible alternative Inspect the area checking that there are no burrs or other foreign particles interfering with the contact surface STEP 2 Wipe clean the mounting surface and spread on a light film of grease oil or similar coupling fluid prior to installation Figure 3 Mounting Surface Lubrication Adding a coupling fluid improves vibration transmissibility...

Page 6: ... for most applications It is generally used only for machinery monitoring and other portable trending applications Both the accuracy and repeatability at low 5 Hz and high frequency 1 kHz ranges are questionable 5 0 CABLING Care and attention to installation is essential as the reliability and accuracy of your system is no better than that of the output cable Charge mode sensors require the use of...

Page 7: ...to a precise sensitivity such as 10 0 or 100 0 mV g for ease of data analysis This is accomplished with most PCB laboratory charge amplifiers and some miniature in line units as well For fixed sensitivity in line charge converters such as the PCB Series 422 the system sensitivity mV g is determined as the product of the charge amplifier sensitivity mV pC and the charge sensitivity of the accelerom...

Page 8: ...edure themselves While this may result in both a savings of time and money it has also been attributed to incorrect readings and costly errors Therefore in an effort to prevent the common mistakes associated with customer performed calibration this document includes a broad overview of the Back to Back Calibration technique This technique provides a quick and easy method for determining the sensit...

Page 9: ...2 Figure 12 Typical Calibration Set Up Adhesively mounted sensors use similar practices However in this case a small portion of quick bonding gel or similar temporary adhesive is used to attach the test sensor to a reference standard designed with a smooth flat mounting surface In addition to mounting the selection of the proper equipment is critical Some of the more important considerations inclu...

Page 10: ...rs Incorrect Understand Back to Back Calibration limitations Do not expect the uncertainty of calibration to be any better than 2 In fact the uncertainty may be as high as 3 or 4 for frequencies 10 Hz or 2 kHz Since large sensors may affect high frequency accuracy verify that the test sensor does not mass load the reference standard Validate your calibration system with another accelerometer prior...

Page 11: ......

Page 12: ......

Page 13: ......

Reviews: