background image

 

- 3 - 

4.0 Installation  

There are many ways to mount the microphone, from simply placing the microphone in a clip to 
using a two-dimensional array stand. A diagram of an array configuration is provided in Figure 
3.  
For installation of this type, remove the microphone from its package, connect the appropriate 
cable between the signal conditioner and the microphone, and connect a second cable between 
the signal conditioner and the data acquisition device. Cabling may consist of standard low-
impedance coax cables.   Desired electrical connectors would be dependent on the 130 series 
microphone used and the input of the signal conditioner and read-out instrument.   PCB

®

 offers 

these cables in a variety of standard or customer specified lengths. 

 

Figure 3 

The 130A24 microphone  features  a removable  grid  cap and replaceable protective cover.  This 
allows  the  130A24  to  withstand  harsh  environments  while  extending  the  lifespan  of  critical 
sensitive components. If the water and dust resistant cover is damaged, having a removable grid 
cap allows the user to replace this protective cover quickly and easily.  

To  replace  a  grid  cap,  first  unthread  the  grid  cap  from  the  housing.  Next,  peel  off  the  existing 
cover from the end of the housing. Sharp edged tools can be used to aid in cover removal, but it 
is important to keep the tool at  a shallow angle. DO NOT make contact  with the microphone’s 
delicate diaphragm (right hand side of Figure 4).  To ensure a proper seal, gently wipe the top of 
the microphone housing to remove any moisture or debris.  Do not use solvent when removing 
debirs  from  the  housing  because  solvents  have  the  potential  to  damage  the  microphone.  Apply 
the replacement water resistant patch to the end of the housing (left hand side of Figure 4) while 
keeping  it  centered.  Finally,  assemble  the  grid  cap  back  onto  the  housing.    Additional  079A46 
five-pack replacement water resistant pads can be ordered separately. 

 

 

 

 

Figure 4

 

5.0 Calibration 

All microphones are supplied with calibration documentation showing the free field frequency 
response and the conditions under which the calibration was performed.  Calibrations are 
performed with reference microphones traceable to national laboratories specializing in acoustic 
measurements (NIST, PTB, or DFM). PCB is ISO 10012-1 and ISO 9001-2008 
certified.  Calibration methods used by PCB for acoustic pressure have been audited for 
proficiency and accredited for compliance with ISO/IEC 17025:2005; ANSI/NCSL Z540-1-1994 
& ANSI/NCSL Z540.3. Sample calibration certificates are provided in Figure 5,      Figure 6, and 
Figure 7 for the 130F series, 130A23, and 130A24, respectively.   

Summary of Contents for 130A25

Page 1: ...EAMPLIFIER 10 32 CONNECTOR TEDS Installation and Operating Manual For assistance with the operation of this product contact the PCB Piezotronics Inc Toll free 716 684 0001 24 hour SensorLine 716 684 0...

Page 2: ...alized tests including sensitivity at elevated or cryogenic temperatures phase response extended high or low frequency response extended range leak testing hydrostatic pressure testing and others For...

Page 3: ...tion on particular operating steps The following symbols may be found on the equipment described in this manual This symbol on the unit indicates that high voltage may be present Use standard safety p...

Page 4: ...China RoHS 2 Disclosure Table Pb Hg Cd Cr VI PBB PBDE O O O O O O PCB X O O O O O O O O O O O X O O O O O O O O O O O O O O O O O O O O O O O O O X O O O O O O O O O X O O O O O O O O O O O X O O O O...

Page 5: ...O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indicates that sai...

Page 6: ...oustic mode analysis near field acoustic holography or sound intensity measurments To ensure the best performance from your 130 series microphone follow the installation and operating procedures in th...

Page 7: ...g distances with negligible signal loss 3 0 Powering All ICP powered microphones require constant current excitation for proper operation For this reason use only PCB constant current signal condition...

Page 8: ...he end of the housing Sharp edged tools can be used to aid in cover removal but it is important to keep the tool at a shallow angle DO NOT make contact with the microphone s delicate diaphragm right h...

Page 9: ...ce with ISO 9001 ISO 10012 1 and ANSI NCSL Z540 3 4 Due to state of the art limitations 4 1 calibration ratios are not possible on pressure measurement standards microphones and acoustic calibrators C...

Page 10: ...e with ISO 9001 ISO 10012 1 and ANSI NCSL Z540 3 4 Due to state of the art limitations 4 1 calibration ratios are not possible on pressure measurement standards microphones and acoustic calibrators Ca...

Page 11: ...ibration is performed in compliance with ISO 9001 ISO 10012 1 and ANSI NCSL Z540 3 4 Due to state of the art limitations 4 1 calibration ratios are not possible on pressure measurement standards micro...

Page 12: ...ver the microphone output may be affected by harsh environments such as moisture dirt mechanical shock or other unusual conditions This may manifest itself in a number of ways ranging from a loss in f...

Page 13: ...828 8840 ISO 9001 CERTIFIED AS9100 CERTIFIED In the interest of constant product improvement specifications are subject to change without notice Copyright PCB Group Inc 2016 PCB and ICP are registered...

Page 14: ...7 dB 1 Electrical Excitation Voltage 18 to30 VDC 18 to30 VDC Constant Current Excitation 2 to20 mA 2 to20 mA Output Bias Voltage 5 5 to14 VDC 5 5 to14 VDC Output Impedance 150 Ohm 150 Ohm Physical Ho...

Page 15: ...003 005 FILLETS AND RADII 0 07 0 13 OUTLINE DRAWING 72402 1 OF 1 5X MODEL 130A25 ICP MICROPHONE DECIMALS XX 03 XXX 010 DECIMALS X 0 8 XX 0 25 JJF 3 23 20 JJF 3 23 20 MJN 3 23 20 72402 PCB Piezotronic...

Reviews: