6. OPERATIONS  

 

The “SAS” air sampler should be positioned in vertical position in a quiet place on a strong surface. Avoid current of air 
around the instrument. 
2. Insert a Contact plate (RODAC plate) into the aspirating chamber of the air sampler and screw the aspirating head. 
3. Program 1000 litres of air on the “SAS” air sampler and switch it on. 
4. Position the “Pyramid” system (c) on top of the SAS air sampler. 
5. Switch on the “Pyramid” system by briefly pressing the switch button of the flow sensor. The software version will 
appear on the LCD display first, after which the instrument moves into the mode for measuring the current air flow 
(litres per minute or cubic feet per minute). The measuring time is approximately 30 seconds to obtain a stable data 
reading. 
6. The LCD display switches to the converse version (L/minute or cubic meter per minute) by repeated pressing of the 
push button. 
7. Hold the push button pressed down until the unit switches off. 
8. Record the lower and higher values on the record paper. 
9. Repeat the operation from 3 to 6 ten times. Record the values of each cycle on the record paper and calculate the 
average.   

 

7. MAINTENANCE 

 

In order to maintain its precision, it is recommended that the system is carefully treated.  
The system doesn’t require special care and maintenance procedure. Clean the outside surface with a normal cloth. 
Do not subject the system to sudden shock.   
Battery replacement 
It is suggested that the lithium battery 3V type CR2032 is replaced by servicing people or watch dealer. 
(a). Unscrew the six screws of  the metal back. 
(b). Unscrew the screw of  the battery cover. 
(c). Remove the cover and extract the lithium battery. 
(d). Put in the new battery observing the polarity (+ upwards). Screw back the battery cover. Make sure that the tiny 
tongue in the centre of the top is raised so that it touches the metal back. 
(e).  Before refitting the metal back cover, make sure that the seal is clean and securely positioned in its slot.  
(f) Refit the back and screw its six screws.  
 
 

WARRANTY 

 

WARRANTY CONDITIONS 

International pbi warrants this product to be free of defects in materials and functionality. 
The repairs will be effected by International pbi or by authorized dealers as follows: 
 

Faulty components free replacement 

 

The labour is always at customer's expense 

The equipment will be sent at buyer's expenses and risk and the return will be ex-works 

This warranty does not apply to damages resulting from tamperings,abnormal use and installation and does not 
cover transport damages 

International pbi shall not be liable for any commercial damages, whether incidental, consequential or otherwise 
derived from the use of instruments 

 
 

  
 
 

Summary of Contents for 86666

Page 1: ...PYRAMID Flowcheck System Code n 86666 I IN NS ST TR RU UC CT TI IO ON N M MA AN NU UA AL L International PBI S p A Via Novara 89 20153 Milano Italy Tel 39 02 48779 1 Fax 39 02 40090010 International...

Page 2: ...y grounded three prong wall outlet Warning Fire or electric shock hazard do not expose this product to rain or moisture This unit contains dangerous voltages NEVER remove cover unless qualified to do...

Page 3: ...al 29 08 05 Rev 1 3 INDEX 1 Introduction pg 4 2 Technical specification pg 4 3 Component of the Pyramid system pg 5 4 Preliminary inspection pg 5 5 Installation pg 6 6 Operations pg 7 7 Maintenance pg...

Page 4: ...xact air flow that is passing through the fan The Pyramid flowcheck system can display the airflow either in litres per minutes L or cubic feet per minute CF The precision of the measured value is for...

Page 5: ...ded that this scientific system is treated with care and attention in order to maintain its precision 4 PRELIMINARY INSPECTION 4 The apparatus is subject to specific working tests before shipping and...

Page 6: ...ong surface Avoid current of air around the instrument 5 Program 1000 litres of air on the SAS air sampler and switch it on 6 Position the Pyramid system c on top of the SAS air sampler 7 Switch on th...

Page 7: ...ded that the system is carefully treated The system doesn t require special care and maintenance procedure Clean the outside surface with a normal cloth Do not subject the system to sudden shock Batte...

Page 8: ...EN KUNDEN F r Benutzer unserer Apparate sind alle ersatzteilen die sich aus mechanischen Fehlern in normalen Gerbrauch ergeben f r zen Monate garantiert Damit wir Ihnen einem erstklassigen Service die...

Page 9: ...TROL DATE INSTRUMENT MODEL DATE OF LAST CONTROL AVERAGE VALUE OF THE LAST CONTROL 55mm CONTACT PLATE RODAC 84mm CONTACT PLATE 90mm PETRI DISH TEST N 1 2 3 4 5 6 7 8 9 10 LTS MN TEMPERATURE LOCATION AV...

Page 10: ...KED TO THE EXPORT DPT Company Address City Zip Code State Phone Fax Email Name to contact Lab Division return request of technical service Equipment Model Serial Number for calibration repair other sp...

Page 11: ...International pbi S p A Instruction Manual 29 08 05 Rev 1 11...

Page 12: ...153 Milano HEREBY DECLARES under its sole responsibility that the following product PYRAMID Flowcheck System Art 86666 conforms with the technical requirements of the following directives 89 336 CEE 9...

Reviews: