Page 49
FILTER - SETTING FILTER PARAMETERS
d1G.bAn.
WEIGHT FILTER VALUE [1201]
This parameter adjusts not only the refresh rate of the display, but specially the serial and analogue
output. The maximum refresh rate of the display is limited to 25 Hz
High filter values speed up the weight update.
Low values of the filter slow down the weight update.
Factor
(Hz)
Settling
Time
(mS)
ADC Freq
(Hz)
N readings
Monotony
Time (mS)
Oscillation
time (mS)
Oscillation
range (div)
MANUAL [0]
Selectable
Settable
Settable
Settable
Settable
50 [1]
20
250
5
20
4000
10
25 [2]
40
100
5
40
3000
12
10 [3]
100
50
5
80
2500
16
5 [4]
200
50
10
100
2000
20
2 [5]
500
50
25
250
1500
25
1,25 [6]
800
12,5
10
300
1500
25
1 [7]
1000
12,5
12
400
1500
25
0,7 [8]
1500
12,5
19
500
1200
30
0,5 [9]
2000
12,5
25
600
1000
30
Default: 2 Hz
The following parameters are visible and therefore can be set, only if the parameter selection is MANUAL.
C. rAtE.
ADC SPEED [1202]
With this parameter, the frequency of weight acquisition is adjusted. If the parameter changes to values
higher than 12.5 Hz, the weight must be stable. In case of instability the instrument will immediately
indicate an error message.
Selectable Values:
12,5 [0]
50 [1]
100 [2]
250 [3]
1000 [4]