INSTALLATION
GENERAL INFORMATION
The DAT 100 consists of a motherboard, to which are added the options available, accommodated in
a plastic enclosure for DIN rail 35mm.
The DAT 100 should not be immersed in water, subjected to jets of water and cleaned or
washed with solvents.
Do not expose to heat or direct sunlight.
OVERALL DIMENSIONS
ELECTRIC INSTALLATION
DAT 100 is equipped with removable screw, pitch 5.08 mm.
The load cell cable must be shielded and channeled away from power cables to prevent
electromagnetic interferences.
INSTRUMENT POWER SUPPLY
The instrument is powered through the terminals 23 and 24. The
power supply cable must be channeled separately from other cables.
The internal circuit is galvanically isolated from the supply voltage.
Power supply voltage: 24 Vdc± 10%, max. 2W
Summary of Contents for DAT 100
Page 2: ...Page II ...